Measuring Function Definition for Class A Devices in India

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

India applies a dual test of intended use plus unit display to determine whether a medical device is a device with a measuring function. A device is considered to have a measuring function where the manufacturer intends it to be used for quantitative measurement of a physiological or anatomical parameter, or of a quantifiable characteristic of energy or substances delivered into or removed from the body, and the measurement result is displayed in legal units or other accepted units. The presence of graduations alone does not make a device a measuring device; the decisive question is the intended use. This distinction decides which filing route is open, and it is one of the most frequent points of misjudgement in Class A exemption planning.

  • The test combines intended use, the quantity measured and the unit display.
  • Graduations on a product are not, by themselves, a measuring function.
  • The official definition is given in CDSCO FAQ question 72.
  • Measuring function is one of the three disqualifiers for the Class A exemption.
  • The assessment is made on the English-language materials actually marketed.

2. Legal basis

The operative definition is administrative rather than statutory, given in the CDSCO FAQ, and it interacts with the classification provisions of the Rules and with the Class A exemption. The three sources below should always be read together, because the same product may sit under all three at once: it must be classified, its measuring status must be judged, and the exemption depends on the outcome of both.

CDSCO Frequently Asked Questions on the Medical Devices Rules, 2017, question 72:

“Any medical device intended by the manufacturer for the quantitative measurement of a physiological or anatomical parameter, or of the quantity or a quantifiable characteristic of energy or substances (including drugs) delivered into or removed from the human body, and where the result of the measurement is displayed by the device in legal units or other acceptable metrological units, is deemed to be a medical device with a measuring function.”

SubjectLegal source
Definition of measuring functionCDSCO FAQ question 72
Where classification authority sitsMDR 2017 Rule 3; CDSCO FAQ question 77
Application of the exemptionCDSCO FAQ question 69

3. Expert insights

Insight one: the sequence of the assessment is use, then quantity, then units. Step one asks whether the intended use declared in the IFU and on the labels includes quantitative measurement. Step two asks whether what is measured is a physiological or anatomical parameter, or a quantity of energy or substances delivered into or removed from the body. Step three asks whether the result is displayed in units. Only when all three steps are satisfied is the device a measuring device. Working the steps in order matters, because a device can display numbers for convenience without its intended use being measurement, and treating every displayed number as a measuring function would sweep in ordinary products that the definition was never meant to capture.

Insight two: a scale is not a measurement. An ordinary container is not a measuring device; the same container, if graduated and intended for precise dosing, may be found to be one. The dividing line is whether the manufacturer holds out quantity as a function of the device in its claims. This is a documentation question before it is an engineering question: the answer lives in the words the company chooses for its labelling and instructions, not in the moulded lines on the product.

Insight three: the eligibility risk of the exemption route sits exactly here. Measuring function is one of the three disqualifying conditions of the Class A exemption. Registering a Class A product that does have a measuring function through the exemption route has a severe consequence: the registration does not confer manufacturing or import authorisation — the product is in substance unlicensed, falling directly under s.27(b) of the Drugs and Cosmetics Act. The saving in fees and time achieved by the wrong route is trivial against the exposure it creates.

Insight four: the judgement is made on the English text of the product labelling and IFU. CDSCO looks at the materials actually placed in the Indian market. If the Chinese instructions say the product is for holding only, but the English IFU says it is graduated for accurate dosing, the English governs. Companies marketing in multiple languages must therefore keep their claims consistent across languages, because the strictest reading of any market-facing text will be the one applied.

4. Practical recommendations

  • GRZAN does not judge the measuring function from the product name; it prepares a three-column comparison: the claim in the home-market materials, the English IFU text and the label artwork.
  • Whenever the English materials contain wording conveying quantitative readings together with units, the product is treated strictly as having a measuring function and routed through the regular MD-14/MD-15 channel.
  • Claims are aligned across all language versions of the IFU and labelling before filing, so that the Indian-market English text cannot contradict the source materials.
  • The determination is recorded in the classification rationale file together with its reasoning, so that the same conclusion can be defended consistently if queried.
  • The trade-off is stated openly to the client: the cost of being strict is fees and time; the cost of being lenient is criminal risk — and that calculation is not left to chance.

5. Verification checklist

  • □ The English IFU has been checked for quantitative measurement wording.
  • □ The label artwork has been checked for graduations and units.
  • □ The object of measurement falls within physiological or anatomical parameters, or quantities delivered into or removed from the body.
  • □ The home-language claims and the English materials are consistent.
  • □ The determination has been written into the classification rationale file.

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support across South Asia, including India. Our team prepares dossiers to CDSCO expectations, manages SUGAM submissions and deficiency responses, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Regulatory information verified as of 27 October 2026.

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail