Brazil Family, System and Kit Grouping under IN 320/2024
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
The direct legal basis for combined filings in Brazil is not a standalone grouping regulation but the sole paragraph of Article 3 of RDC 751/2022: ANVISA also grants notification or registration to 3 types of unit — family (família), system (sistema), and kit (conjunto/kit) — with the specific grouping rules set by a dedicated normative instruction, currently IN nº 320/2024. GRZAN’s judgment: grouping compresses the number of registration numbers, not the burden of proving technical differences. The more models inside a group, the more the technical dossier must explain, item by item, why each difference does not constitute a new risk.
2. Legal basis
- Requirements for the classification, notification and registration of medical devices and for labeling and instructions for use — RDC nº 751/2022, resolution of September 15, 2022 (Board meeting of September 14), published in DOU No. 180 of September 21, 2022 (Seção 1, p.172), effective March 1, 2023 (Article 67): Article 3 sole paragraph, Article 4(XX), Article 4(XXVI), Article 5, Article 6, Article 7, Article 14(V)
- Grouping rules for families, systems and kits — IN nº 320/2024 (its purpose was confirmed on ANVISA’s official device page; the full text was not obtained in this verification, and the specific grouping criteria defer to ANVISA publications)
- Amendments in force to the requirements for the classification, notification and registration of medical devices — RDC nº 777/2023, resolved and effective March 1, 2023 (its existence was confirmed through Brazil’s WTO TBT notification; article-level content is secondhand — a practical estimate; defer to ANVISA publications)
- Unique device identification requirements — RDC nº 591/2021, effective since January 2022: Article 15 (as amended by RDC nº 884/2024) (the deadline table reflects common industry compilation; defer to ANVISA publications)
- Transmission and management of UDI data in the SIUD database — IN nº 426/2026, issued February 13, 2026, published in the DOU on February 18, 2026, effective March 1, 2026 (Article 16): Article 3, Article 5, Article 11
3. Expert insights
Where is the value of the sole paragraph of Article 3? GRZAN’s reading: in Brazil, compliance costs attach to each registration number — the technical dossier, the Portuguese labeling and instructions, the TFVS cost items, and afterwards every change and every technovigilance duty follow the number. Allowing one number per family, system, or kit is essentially letting homogeneous products share a single compliance vehicle and squeezing out duplicated cost.
But the boundary of grouping is drawn by definitions, not by the applicant’s preference. Article 4(XX) defines a family as a set of products similar across 5 dimensions — indications or use, mechanism of action, technology, composition, and precautions. Article 4(XXVI) sets a harder test for kits: the components must come from the same manufacturer or the same manufacturing group, and no interdependence may exist among them. GRZAN’s conclusion: kits and systems are two different kinds of combination — a system permits components to work together, while a kit requires precisely that components be independent — and preparing documentation for both with the same mindset will inevitably go wrong.
The second judgment is that grouping does not compress identification granularity. Article 5 of IN 426/2026 requires each set of data to correspond to one unique UDI-DI and one device model, and Article 3 requires the UDI-DI to be transmitted before the device is placed on the Brazilian market, while the UDI-PI is maintained by the holder within its own quality system to preserve traceability. This runs in the same direction as the labeling deadlines set by risk class in Article 15 of RDC 591/2021: however many models the group contains, that many identification data sets must be prepared — not 1 of them can be skipped.
The third judgment is that the burden of proof is not lightened by combining. Differences among family members in indications, materials, or performance must still be argued item by item; what combination buys is administrative efficiency, not a looser review standard. Stuffing obviously cross-category models into one group costs a deficiency notice for the entire group.
The fourth judgment is that the CBPF is not reduced by grouping. The certificate requirement for Class III (high-risk) and Class IV (maximum-risk) production units still follows Article 14(V); grouping operates on the filing unit, not on the manufacturing floor.
4. Practical recommendations
First, draw the family boundary with a 5-dimension calibration table. The GRZAN regulatory team recommends filling in, per model at project initiation, the indications or use, mechanism of action, technology, composition, and precautions; any non-similar dimension means the model is split out.
Second, for kits check the entity and independence first. Confirm that all components come from the same manufacturer or manufacturing group, and state in writing that no interdependence exists among them; if it fails, file as a system or as individual products.
Third, plan identification data together with grouping. Prepare a UDI-DI per model and schedule the IN 426/2026 transmission before market placement in Brazil, not after the registration number arrives.
Fourth, count the change impact. When any model in the group triggers a change requiring approval, the documentation of the whole group may be affected; inventory drawdown and revised production scheduling should be considered together with the grouping strategy.
Fifth, complete a final verification before filing. The grouping details of IN nº 320/2024 and the article-level amendments of RDC 777/2023 were not obtained from official originals in this review; GRZAN recommends completing a final check against anvisalegis and the DOU before formal submission and filing the verification record in the project archive.
5. Verification checklist
- □ The family boundary has been calibrated model by model against the 5 dimensions of Article 4(XX)
- □ Kit components are confirmed to come from the same manufacturer or manufacturing group, with no interdependence among components
- □ The grouping conclusion is recorded in writing, and cross-dimension models have been decided for splitting
- □ UDI-DIs are prepared per model, with transmission scheduled before Brazilian market placement
- □ The CBPF requirement for Classes III/IV has not been mistaken as reducible by grouping
- □ The grouping details of IN nº 320/2024 have undergone final verification on anvisalegis or the DOU
Regulatory information verified as of October 29, 2026
GRZAN Regulatory Affairs Team
GRZAN supports manufacturers entering the Brazilian market with ANVISA registration pathway and classification strategy services, including family, system, and kit grouping design under RDC 751/2022 and IN 320/2024, per-model UDI planning under IN 426/2026, and grouping strategies that cut registration numbers without inflating review risk. To test whether your portfolio should be grouped or split, contact the GRZAN regulatory team at info@grzanglobal.com.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us