IVD Classification in Turkey: Madde 49 and Classes A to D
Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
In vitro diagnostic devices follow a classification track of their own in Turkey, separate from the device rules in both structure and terminology. The framework is anchored in Madde 49 of the IVD Regulation, with the conformity assessment routes set out in Madde 50, and it carries an additional normative layer for the highest-risk class. The points below set out the anchors that any IVD registration plan for Turkey must be built on.
- Turkey classifies IVDs by risk into the four classes A, B, C and D. The basis is Madde 49, paragraph 1, of the In Vitro Diagnostic Medical Device Regulation, and the method follows the Ek VIII of that Regulation.
- The conformity assessment routes are written in Madde 50: before placing a product on the market, the manufacturer must complete the relevant procedure under Ek IX to Ek XI.
- Class D IVDs carry additional common specifications: TİTCK issued the Common Specifications for Certain Class D In Vitro Diagnostic Medical Devices in the form of a communiqué (Tebliğ), released in step with the third amendment of the device Regulation.
- IVDs use the system of performance studies (performans çalışması) and performance evaluation, not the device-side system of clinical trials and clinical evaluation; the two terminologies cannot be interchanged in Turkey.
2. Legal basis
The table below consolidates the instruments cited in this article, pairing each with its original Turkish title, its official reference and publication date, and the provisions on which the analysis rests. The communiqué on common specifications deserves particular attention because it operates as an independent compliance layer on top of the Regulation itself.
| Name of regulation | Original Turkish title | Reference / number | Publication date | Provision |
| In Vitro Diagnostic Medical Device Regulation | İn Vitro Tanı Amaçlı Tıbbi Cihaz Yönetmeliği | Resmî Gazete Sayı 31499 (1. Mükerrer) | 2021-06-02 | Madde 49, Madde 50, Ek VIII, Ek IX–XI |
| Communiqué on Common Specifications for Certain Class D IVDs | Belirli Sınıf D İn Vitro Tanı Amaçlı Tıbbi Cihazlara Yönelik Ortak Spesifikasyonlar Hakkında Tebliğ | Resmî Gazete Sayı 32635 | 2024-08-17 | Full text |
| Current text on the Legislative Information System | Mevzuat Bilgi Sistemi | MevzuatNo 38658 | Verification date 2026-10-29 | Full text and Ek annexes |
3. Expert insights
Insight one: Class A is treated in many markets as nearly barrier-free, but in Turkey it does not mean no action. Class A products still carry obligations of manufacturer registration, UDI (for Class A from 2027-05-26), post-market surveillance and vigilance. Reading Class A as exempt from oversight loses points directly at the ÜTS and surveillance stage. The correct reading is that the class determines the conformity assessment route, not the breadth of the general obligations, and the general obligations follow the product wherever it is classified. Teams that plan Class A products without a surveillance budget or a reporting owner therefore inherit the same post-market workload as any other class, only with less preparation behind it, and the gap becomes visible precisely at the moments when evidence is requested.
Insight two: the common specifications communiqué is an independent compliance layer. This is a normative document issued in the form of a Tebliğ (communiqué), sitting at a different hierarchical level from the Regulation itself, yet in practice it binds Class D products directly in its technical requirements. Its release alongside 2024-08-17 Sayı 32635 shows that Turkey has followed the European pace in regulating the high-risk tier of IVDs. Manufacturers of Class D products therefore need to run two checklists in parallel: one against the Regulation and its annexes, and one against the common specifications, because a product can satisfy the Regulation and still fail the finer-grained specification requirements.
Insight three: two Ek VIII annexes, nominally isomorphic but substantively different. The device Regulation has an Ek VIII and the IVD Regulation also has an Ek VIII. They share a name and a broadly similar chapter structure, but the objects they assess are entirely different. Internal documents must state clearly which Regulation’s Ek VIII is being cited; otherwise the reference points to the wrong rule spectrum. GRZAN has seen filing packages where the citation was correct in form and wrong in substance precisely because the annex name alone does not discriminate between the two regimes, and the error surfaced only when a reviewer traced the rule numbers against the wrong annex text. The remedy is mechanical and cheap: every internal citation should carry the full regulation name alongside the annex reference, and a single style rule in the document control procedure is enough to enforce it across the whole IVD portfolio.
4. Practical recommendations
- Build a two-column ledger for IVD classification: product, class, applicable rule provision, and assessment route (which of Ek IX / X / XI applies).
- For Class D projects, maintain a separate common-specifications conformance checklist, because the granularity of the common specifications is usually finer than the body text of the Regulation.
- Unify terminology internally: IVD projects must use performance study and performance evaluation throughout, and must not use clinical trial or clinical evaluation, so that submission materials are not returned or challenged.
5. Verification checklist
- □ The IVD class has been confirmed as one of A / B / C / D.
- □ The Ek VIII of the IVD Regulation (the annex file of MevzuatNo 38658) has been downloaded, not the Ek VIII of the device Regulation.
- □ Class D products have been checked item by item against the common specifications communiqué of Sayı 32635.
- □ The assessment route has been pinned down to the specific procedure in Ek IX / Ek X / Ek XI.
- □ Internal terminology has been unified to performance study and performance evaluation.
Regulatory information verified to 29 October 2026. The provisions and issue numbers listed in this article were checked against the Official Gazette index of Turkey and the original texts on the legislative information system (MevzuatNo 38658).
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Turkish market. Our team manages TITCK approval applications and UTS registrations, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us