SFDA Authorized Representative: Mandatory Requirement for Foreign Manufacturers

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Wherever the manufacturer has no legal entity established in Saudi Arabia, an AR (Authorized Representative) must be appointed before an MDMA application can be completed. Article 1 of M/54 defines the AR as a legal entity located within Saudi Arabia, holding written authorisation from the foreign manufacturer, acting on its behalf to implement the law and the regulation. The AR is not a nominal agent: it is the statutory contact point between the Manufacturer and SFDA, with duties including adverse event reporting, traceability cooperation and designation of the responsible SFDA liaison. Chinese manufacturers most often underestimate the processing time of the AR licence and the notarisation and legalisation chain — both frequently finish later than the technical documentation and decide whether the GHAD submission can be filed at all. The publicly cited AR licence fee is SAR 2,600 per contract per year (practical estimate).

  • Article 1 of M/54 confines the AR to a legal entity in Saudi Arabia with written authorisation from the foreign manufacturer.
  • The AR itself must first obtain an SFDA licence; signing an agreement alone is not enough.
  • The publicly cited AR licence fee is SAR 2,600 per contract per year, plus 15% VAT (practical estimate).
  • The notarisation and legalisation chain runs home-country notarisation, Foreign Ministry attestation, then Saudi embassy legalisation; a missing step means GHAD will not accept the file.
  • MDS-REQ 11 V2.0 obliges the manufacturer and the AR to report adverse events and complaints to the NCMDR and to designate responsible SFDA liaison persons.

2. Legal basis

The table traces the AR obligations back to their sources, from the definition in the Law to the reporting duties in the post-market requirements.

RegulationReference numberPublication dateArticles and key points
Law of Medical Devices and SuppliesRoyal Decree No. M/54; Hijri 1442/7/618 February 2021Article 1: Authorized Representative defined — a legal entity within Saudi Arabia holding written authorisation from the foreign manufacturer, implementing the law and regulation on its behalf
Requirements for Medical Devices Marketing AuthorizationMDS-REQ1 V6.0; MDS-REQ-001-V6/21121919 December 2021Article 4 (AR-related submission requirements; item numbers not captured one by one) (practical estimate)
Requirements for Licensing of Medical Devices EstablishmentsMDS-REQ9 V1.031 July 2022Establishment licensing conditions (AR licence article numbers not captured) (practical estimate)
Requirements for Post-Market Surveillance of Medical DevicesMDS-REQ 11 V2.0; MDS-REQ-011-V2/23032323 March 2023The manufacturer, the AR and healthcare institutions must report adverse events and complaints to the NCMDR; the manufacturer, AR, importers and distributors must designate responsible SFDA liaison persons
SFDA Reliance Approach for Medical DeviceMDS-G030 V1.0; MDS-G-030-V1/2609099 September 2026Step 1: create an account on the SFDA E-Services portal and obtain an Establishment Number
AR licence feeNo official fee schedule referenceDate of the fee figure not obtained (practical estimate)SAR 2,600 per contract per year, plus 15% VAT

3. Expert insights

Why Saudi Arabia confines AR eligibility to a domestic legal entity

The M/54 definition of the AR carries three qualifiers: located within Saudi Arabia, a legal entity, and written authorisation from the foreign manufacturer. All three are indispensable and reveal the regulatory logic — SFDA needs a subject that can be served locally, held accountable and inspected. This is stricter than the EU MDR’s EC-REP: MDR Article 11 only requires the authorised representative to be established in the EU, and a natural or legal person is theoretically permissible. The US FDA requires a U.S. Agent for foreign firms, limited to communication and receipt of information, not extending to post-market reporting duties. China’s NMPA requires foreign registrants to appoint a Chinese enterprise legal person as responsible party, the closest parallel in scope to the Saudi AR, but Saudi Arabia adds an independent AR licence. The consequence is direct: the AR does not take office on a signed agreement; it must first obtain its own SFDA licence.

The practical impact on Chinese manufacturers: the AR decides whether you can file, not merely who files

GRZAN’s scheduling experience is that three AR workstreams must start in parallel with the technical documentation. First, vetting the AR entity’s qualification — whether it holds a valid Establishment licence covering the product category. Second, signing the AR agreement and running the notarisation and legalisation chain: home-country notarisation, Ministry of Foreign Affairs attestation, Saudi embassy legalisation. Third, completing the AR fields in the MDMA application and obtaining the Establishment Number (MDS-G030 Step 1 expressly requires opening an account on the SFDA E-Services portal). Run in series, these three items usually close later than the technical file. Another frequently ignored knock-on effect is data ownership: MDS-REQ 7 V4.0 requires UDI data to be submitted and maintained in Saudi-DI by the manufacturer or the AR; if the AR lacks data maintenance capability, post-market model changes and UDI-DI updates will break the chain.

The AR duty list: from adverse events to final destruction

The AR’s obligations are scattered across several documents and group into four. First, report any adverse event and complaint to the NCMDR, follow up the investigation and provide all documents, within deadlines of 2 days for a serious public health threat, 10 days where death or serious deterioration is possible, and 30 days otherwise. Second, trace devices at SFDA’s request and maintain a tracking system recording supply and distribution information. Third, designate a responsible SFDA liaison person, on the same footing as manufacturers, importers and distributors. Fourth, cooperate in the post-market stages of resale, lending and donation of second-hand devices, after-sales maintenance and destruction of discarded devices. By comparison, the EC-REP’s core duties are to keep a copy of the technical documentation and cooperate with the competent authority, without directly carrying vigilance reporting; the U.S. Agent bears no such duties at all. Choosing an AR in Saudi Arabia is therefore in substance choosing a local compliance partner able to carry vigilance business, not a mailing address.

The notarisation and legalisation chain of the AR agreement is described at https://www.grzan.cn/regulations/saudi/ar-agreement-legalization-chain-32 , and the AR licence duty list at https://www.grzan.cn/regulations/saudi/authorized-representative-licensing-and-duties-31 .

4. Practical recommendations

  • Start AR screening alongside the technical documentation, vetting whether the Establishment licence scope covers the product category; do not wait until just before filing.
  • The AR agreement must fully embody the three elements — a domestic legal entity, written authorisation, and implementing the law and regulation on the manufacturer’s behalf — with clauses on vigilance cooperation, document retention and traceability.
  • Allow time for the notarisation and legalisation chain — home-country notarisation, Ministry of Foreign Affairs attestation, Saudi embassy legalisation; a missing step means GHAD will not accept the submission.
  • Open the SFDA E-Services portal account early to obtain the Establishment Number (MDS-G030 Step 1), so it does not become the final bottleneck.
  • Agree in writing with the AR on the division of Saudi-DI data maintenance: name the persons responsible for model names, GMDN codes and UDI-DI.
  • Define vigilance reporting responsibilities: the AR must complete NCMDR reports within the 2-day, 10-day and 30-day deadlines and answer NCMDR follow-up within 5 days.
  • Budget the AR licence fee at SAR 2,600 per contract per year (Saudi Riyal, SAR) plus 15% VAT (practical estimate); in practice the GHAD invoice governs.
  • The AR-related item numbers of MDS-REQ1 Article 4 and MDS-REQ9 were not obtained one by one in this review; download the current versions and verify before submission (practical estimate).

5. Verification checklist

  • □ Confirmed the company has no Saudi legal entity and must appoint an AR
  • □ AR is a legal entity located in Saudi Arabia holding a valid Establishment licence
  • □ AR agreement contains the elements: domestic legal entity, written authorisation, and implementation of the law and regulation on the manufacturer’s behalf
  • □ All three steps of the notarisation and legalisation chain completed (notarisation, Foreign Ministry, Saudi embassy)
  • □ SFDA E-Services account created and Establishment Number obtained
  • □ Saudi-DI data maintenance responsibilities agreed in writing with the AR
  • □ Vigilance reporting mechanism rehearsed, the three deadlines (2, 10 and 30 days) written into the SOP
  • □ AR licence fee budgeted at SAR 2,600 per year plus 15% VAT (practical estimate)
  • □ AR-related item numbers of MDS-REQ1 Article 4 and MDS-REQ9 verified against current downloaded versions

Regulatory information verified as of 29 October 2026.

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Saudi Arabian market. Our team manages SFDA medical device marketing authorisations and Saudi Authorized Representative arrangements, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.