GHAD Electronic Portal: SFDA E-Services Submission System
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
GHAD is the unified electronic submission system of SFDA for medical devices. MDMA applications, AR accounts and technical file uploads all run through it, with no parallel paper or email channel. Third-party accounts say SFDA launched the unified GHAD electronic system in December 2019, replacing the two former gateways of MDNR (low risk) and MDMA (medium and high risk), and after several extensions fully replaced the legacy system on 27 September 2022 (these dates are not corroborated by the original SFDA website text and are treated as a practical estimate). MDS-G030 V1.0 expresses it in the application steps as the SFDA E-Services portal, so the prudent formulation is the SFDA electronic services platform (GHAD / E-Services). Its institutional meaning has three layers: single gateway, checkpoint of data consistency, and the vehicle through which SFDA maintains continuous supervision of companies and products.
- GHAD is the single electronic gateway; MDMA applications, AR accounts and technical file uploads all run through it.
- GHAD launched in December 2019 and fully replaced the legacy MDNR and MDMA gateways on 27 September 2022 (practical estimate).
- MDS-G030 V1.0 Step 1 requires an SFDA E-Services portal account and an Establishment Number.
- In the reliance pathway, Section (2.1) of the Overview part must be answered Yes to activate Section (2.1.1), and the uploaded Annex (1) file must be named Reliance Form with the country set to Saudi Arabia.
- Model names and GMDN codes must be accurate and identical in both GHAD and Saudi-DI (MDS-REQ 7 V4.0).
2. Legal basis
The table lists the documents and systems behind the GHAD submission flow, including the reliance pathway steps.
| Regulation | Reference number | Publication date | Articles and key points |
| SFDA Reliance Approach for Medical Device | MDS-G030 V1.0; MDS-G-030-V1/260909 | 9 September 2026 | Step 1: create an account on the SFDA E-Services portal and obtain an Establishment Number; Step 2: in the MDMA application Overview answer Yes at Section (2.1) to activate Section (2.1.1); Step 3: upload the Reliance Form of Annex (1) with the country of applicability set to Saudi Arabia |
| Launch of, and switch-over to, the unified GHAD electronic system | No official reference number | Launched December 2019; fully replaced the legacy system on 27 September 2022 (practical estimate) | A single gateway replacing the two former MDNR and MDMA gateways |
| Requirements for Unique Device Identification (UDI) for Medical Devices | MDS-REQ 7 V4.0; MDS-REQ-007-V4/220524 | 24 May 2022 | The manufacturer or the AR must submit and maintain data in Saudi-DI; model names and GMDN codes in GHAD and Saudi-DI must be accurate |
| Requirements for Medical Devices Marketing Authorization | MDS-REQ1 V6.0; MDS-REQ-001-V6/211219 | 19 December 2021 | Technical documentation and marketing authorisation submission requirements |
| Launch of the Saudi-DI database (Ramz) | No official reference number | Launched 1 October 2020, open for voluntary registration of all device types (practical estimate) | The UDI data registration database |
3. Expert insights
A single gateway means data inconsistency will always surface
After the MDNR and MDMA gateways merged, the risk class, classification decision, model name, GMDN code and UDI-DI data of the same device all sit in one data foundation. MDS-REQ 7 V4.0 requires the manufacturer or the AR to submit and maintain the data in Saudi-DI, with the data in GHAD and Saudi-DI — model names, GMDN codes — accurate, and new UDI-DI data in the database when the device is placed on the market. A model name differing by a single space between the registration team and the UDI owner can be caught in the comparison at customs clearance or market sampling. Compared with EUDAMED, whose modules went live on a scattered timetable and some of which are not yet fully mandatory, and the US GUDID, where data submission and 510(k) review are also two systems, Saudi Arabia binds registration and the database much more tightly. The benefit is high regulatory efficiency; the cost is that companies must internally integrate three data sources — registration, UDI and labelling.
The submission details of MDS-G030: two hard gates hidden in three steps
The reliance pathway application steps are written concretely, and two spots are easiest to get wrong. The first is the Establishment Number at Step 1: if the company has not opened an account on the SFDA E-Services portal, none of the subsequent actions can start — which is why GRZAN recommends running this in parallel with AR screening. The second is the linkage of Step 2 and Step 3: the question at Section (2.1) of the Overview part of the MDMA application must be answered Yes before Section (2.1.1) is activated; Country and Title are then completed item by item and each approved Evidence Document uploaded. The Step 3 file, the Reliance Form of Annex (1), must be named exactly Reliance Form, with the country of applicability set to Saudi Arabia. Such hard conventions on field and file names show GHAD runs automated validation no human explanation can bypass. Once through, the standard review timeline is 35 to 60 working days, identical across the four risk classes.
Comparison with China’s NMPA, the EU and the US systems: open the account once, maintain it long term
China’s NMPA electronic submission (eRPS) mainly serves registration filings, with separate systems for production licensing and adverse events; the EU’s EUDAMED covers economic operator registration, UDI, certificates and vigilance in several modules; the US FDA separates FURLS/Device Facility from GUDID. Saudi GHAD concentrates these functions further, with the AR account, entity qualification, technical files and product authorisation sharing one account system. GHAD account permission management must therefore be treated as a compliance asset: who holds submission and edit rights, and how leavers’ access is revoked, need written rules. One mistaken submission creates an official record immediately, and withdrawal and correction cost far more than prior control. GRZAN recommends a three-way verbatim comparison before submission, aligning the product model in GHAD, the UDI-DI in Saudi-DI and the physical label. This benefits EU and US projects too, but in Saudi Arabia it is a hard requirement: the registration system and the device database share the same keys, and once the model diverges, every subsequent change amplifies the mismatch. In addition, SFDA provides manufacturers free technical support (call centre 19999, mailbox md.rs@sfda.gov.sa, meetings with the SFDA technical team and workshops); system-level rule questions can be put directly to the authority, keeping the reply on file.
Account opening and its AR link are discussed at https://www.grzan.cn/regulations/saudi/sfda-authorized-representative-mandatory-05 , and the reliance pathway submission details at https://www.grzan.cn/regulations/saudi/sfda-reliance-pathway-mds-g030-12 .
4. Practical recommendations
- Open the SFDA E-Services portal account in parallel with AR screening and obtain the Establishment Number early.
- Establish a GHAD account permission management system: name the holders of submission, edit and approval rights, revoking access immediately on departure or change.
- Integrate registration, UDI and labelling data sources; keep model names and GMDN codes identical in all three, and confirm new UDI-DI data is in the database before market placement.
- Follow Step 2 strictly: answer Yes at Overview Section (2.1) to activate Section (2.1.1), then complete Country and Title item by item and upload the Evidence Documents.
- When uploading the Annex (1) Reliance Form, name the file exactly Reliance Form and set the country of applicability to Saudi Arabia, checking both verbatim before submission.
- Schedule the review cycle at 35 to 60 working days, with the same figure for Classes A, B, C and D.
- For system-level rule questions, use the free SFDA support channels (call centre 19999, mailbox md.rs@sfda.gov.sa) for official confirmation, retaining the replies.
- Cite the GHAD launch and switch-over dates as a practical estimate; formally defer to current SFDA announcements and portal notices.
5. Verification checklist
- □ SFDA E-Services account created and Establishment Number obtained
- □ GHAD account permission list established, covering submission, edit and approval rights and revocation
- □ Model names and GMDN codes identical across GHAD, Saudi-DI and label
- □ New UDI-DI data retrievable in Saudi-DI before market placement
- □ Reliance application confirmed with Overview Section (2.1) answered Yes and Section (2.1.1) activated
- □ Reliance Form file name and country field checked verbatim
- □ Review cycle scheduled at 35 to 60 working days
- □ Replies from official technical support channels retained
- □ GHAD-related dates flagged as a practical estimate, listed for re-checking
Regulatory information verified as of 29 October 2026.
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Saudi Arabian market. Our team manages SFDA medical device marketing authorisations and Saudi Authorized Representative arrangements, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us