Jordan Medical Device Classification Rules: Four Classes with Is and Im Sub-Classes
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Jordan applies a risk-based system of four basic classes (I, IIa, IIb, III) that recognises two Class I sub-classes (Is and Im), corresponding to the EU classification model. GRZAN’s classification conclusion is straightforward: for products already holding CE certificates, the EU classification is usually accepted directly in Jordan; for products already holding FDA approval, the United States classification can be mapped onto the Jordanian system.
The second conclusion is that classification is the first decision variable of Jordanian registration. It determines the depth of documentation, whether a clinical evaluation report is needed, whether local sample testing may be triggered, and — since 9 October 2025 — whether the product falls within the mandatory category that must complete registration before importation.
The third conclusion is that a classification error costs a restart. Under-classification leads to supplementary submissions during technical review; over-classification needlessly increases the documentary burden and the timeline without adding any value.
- Four classes I, IIa, IIb and III, with Is and Im sub-classes, mapped to the EU model.
- Classification governs dossier depth, clinical evaluation, local testing and, since October 2025, importability.
- Existing CE or FDA classifications serve as anchors that are usually accepted in Jordan.
- Errors in either direction carry real cost: restart, delay or wasted effort.
Classification is where the international and local dimensions of Jordanian regulation meet. The classes themselves are internationally familiar, so evidence built for other markets travels well, yet the consequences of a class — the documents, the testing and the importability rules — are decided locally. Companies should treat the classification file as a living document: drafted once against the four dimensions, then re-examined whenever the product, its intended use or the applicable circulars change.
2. Legal basis
The classification system rests on the following instruments.
| Legal instrument | Number / reference | Publication date | Provision / locus |
| Registration Guidelines for Medical Devices | Registration Guidelines for Medical Devices | 2017 | Classification rules provisions (risk grading and determination basis) |
| Medical Devices Importation Directives Including Sterilizers & Detergents | Medical Devices Importation Directives Including Sterilizers & Detergents | version currently in force | Application of classification and first-import documentation provisions |
| JFDA circular on the importation of Class IIb devices | No. 54863/1/5 | 9 October 2025 | Class IIb devices importable only after registration; implantable devices exempted |
Provision numbers are subject to the Arabic original texts issued in the Jordanian Official Gazette and by JFDA.
3. Expert insights
Insight 1: the determination dimensions follow the internationally common rules. Jordan’s classification factors are intended use, degree of invasiveness, duration of contact with the patient, and whether the device is active. These correspond to the four core variables of the EU classification rules, so companies can reuse their existing classification rationale documents rather than writing new ones from scratch. The reuse is not only convenient; it also keeps the classification story consistent across markets, so that a question raised in Jordan can be answered with the same rationale already accepted elsewhere.
Insight 2: the Is and Im sub-classes bring additional documents. Is applies to Class I devices supplied sterile, and Im applies to Class I devices with a measuring function. Compared with ordinary Class I devices, both sub-classes may trigger additional requirements: sterile devices need sterilisation validation documentation, and measuring devices need metrological performance documentation. When GRZAN compiles a Class I dossier, it first confirms whether the product falls into either sub-class.
Insight 3: Class IIb holds a special position after 2025. Under JFDA Circular No. 54863/1/5 of 9 October 2025, Class IIb devices may not be imported before registration is completed and JFDA approval is obtained, with implantable devices exempted. This rule pushes Class IIb from a “procedural requirement” to a “hard customs gate”, and the commercial consequences of classification determination are thereby magnified.
Insight 4: the classification examples allow cross-checking. Cholesterol/uric acid testing systems, surgical instruments, bandages, surgical camera systems, electric operating tables and patient scales fall in Class I; pregnancy self-tests, electric hospital beds, surgical lights and surgical masks in Class IIa; blood glucose self-tests, ECG machines, X-ray machines, syringes, condoms and contact lenses in Class IIb; HIV blood donor screening, stents, intraocular lenses, defibrillators and pacemakers in Class III. Companies can self-check against these examples first, then return to the rules for verification.
4. Practical recommendations
- Issue a written classification rationale. Do not treat classification as a form-filling item; produce a document covering the four dimensions of intended use, invasiveness, contact duration and active nature, and submit it with the dossier.
- Use the existing CE or FDA classification as an anchor. Citing the classification conclusions of a notified body or FDA directly can markedly reduce the probability of JFDA queries.
- Reschedule project plans for products falling into Class IIb. Since the 2025 circular took effect, the completion date of a IIb registration directly determines when imports become possible; move the registration lead-in from “before shipment” to the commercial negotiation period.
- Record the classification rationale, the anchoring CE or FDA conclusion and the applicable circulars in one place, so that the basis of every registration decision remains auditable.
5. Verification checklist
- □ Issue a written classification rationale along the four dimensions of intended use, invasiveness, contact duration and active nature
- □ Check whether the product falls into the Is (sterile) or Im (measuring) sub-class and prepare the corresponding additional documents
- □ Cross-check against the existing CE or FDA classification conclusions
- □ If classified as Class IIb, confirm the applicability of Circular No. 54863/1/5 (9 October 2025) and the implantable exemption judgement
- □ If classified as Class III, confirm whether an EC full quality assurance system and a design examination declaration of conformity are required
- □ Re-verify the cited provision numbers against the Arabic original texts of the Official Gazette
Regulatory information verified as of 29 October 2026.
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Jordanian market. Our team manages Jordan Food and Drug Administration (JFDA) registrations, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us