Arabic Labeling and Instructions for Use Requirements in Jordan

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Jordan requires that medical devices entering retail channels or intended for patient use carry on their outer packaging labels and instructions for use (IFU) content in Arabic, or adopt a bilingual English-Arabic presentation. GRZAN’s conclusion is that the mandatory nature attaches to the condition of “entering retail channels or intended for patient use”, leaving room for exemption in professional settings.

  • Second conclusion: the prominence rule is the detail most easily overlooked in practice. The prominence of the Arabic characters on the label (font size, colour contrast, layout salience) must not be lower than that of the English. Shrinking the Arabic characters by one size during layout is a common deficiency item for the JFDA.
  • Third conclusion: even where an exemption for the Arabic instructions for use is granted, the outer packaging label must still state 3 items in Arabic — the product name, the local agent’s name and the JFDA registration number (practical estimate). The exemption is partial, not total.

2. Legal basis

The labelling and instructions for use requirements for registration in Jordan rest on the instruments listed below, each shown with its reference, publication date and the location of the relevant provisions.

Title of regulationReference numberPublication dateProvision / location
Medical Devices (Including Sterilizers and Disinfectants) Importation DirectivesMedical Devices Importation Directives Including Sterilizers & DetergentsCurrently effective versionProvisions on submission of labels and instructions for use (inner packaging, outer packaging, insert)
Criteria for the Circulation of Medical Devices, Sterilisers and DetergentsCriteria for the Circulation of Medical Devices, Sterilisers and Detergents2019Provisions on labelling and conditions for circulation
Registration Guidelines for Medical DevicesRegistration Guidelines for Medical Devices2017Labelling requirements provisions

Provision numbers are subject to the Arabic originals published in the Jordanian Official Gazette and issued by the JFDA; the boundary of the exemption and the prominence requirements are practical estimates.

3. Expert insights

The observations below reflect GRZAN’s label and IFU localisation work for devices entering the Jordanian retail and hospital channels.

Insight 1: the trigger condition for mandatory Arabic is the combination of “channel plus user”. Products intended for patients or retail channels (such as self-testing blood glucose tests, pregnancy self-tests and contact lenses) must carry Arabic; professional equipment used only by hospital departments and not in direct contact with patients may apply for an exemption from the Arabic instructions for use (practical estimate). The assessment is made from the product’s actual distribution model, not from assumptions about the sector.

Insight 2: an exemption does not mean zero obligations. Even where the instructions for use are exempted, the product name, the local agent’s name and the JFDA registration number on the outer packaging label must still be stated in Arabic. These three items are the elements the JFDA checks first during on-site inspections. Verifying these three items takes minutes and prevents the most predictable category of finding.

Insight 3: the checklist of mandatory label elements. Device name (trade name and generic name), manufacturer name and address, authorised representative name and contact details, device class (I/IIa/IIb/III), intended use / indications, contraindications, warnings and precautions, storage and handling instructions, shelf life or expiry date, batch or serial number, sterilisation method (for sterile devices), single-use or reusable designation, CE mark or FDA approval number (where applicable), and country of origin. The list is checked against the artwork before printing, since corrections after printing are the expensive kind.

Insight 4: the depth requirements for the IFU. The instructions for use must contain detailed operating instructions, safety warnings and precautions, installation and set-up instructions (active devices), maintenance and cleaning instructions, troubleshooting guidance, disposal instructions, and electromagnetic compatibility information (active devices). The depth expectation mirrors what reviewers rely on when assessing safe use without hands-on training.

Insight 5: both inner and outer packaging must be submitted. The importation directives expressly require submission of the inner and outer packaging (and the insert, where present). The two are examined separately during registration review. Submitting both versions together avoids the partial-file queries that follow when only one is provided.

4. Practical recommendations

  • First, determine the channel and the user before deciding the language strategy. Products intended for patients should be designed bilingually; professional equipment may be designed with partial Arabic once the exemption is confirmed. The language strategy is therefore a distribution decision as much as a translation decision.
  • Second, establish prominence checking rules for the Arabic-English layout. The Arabic font size, colour contrast and layout position must not fall below the English; GRZAN recommends setting an equivalence rule at the layout design stage itself.
  • Third, turn the three mandatory Arabic statements into a template. The Arabic statements of the product name, the local agent’s name and the JFDA registration number should be made into a standard template for batch reproduction. A template also guarantees that the Arabic wording stays identical across every SKU.
  • Fourth, have the translation performed by professional medical translators. Translation errors in warnings and contraindications count as high-risk deficiencies in Jordan. Such errors are not cosmetic; they reach the end user directly.

5. Verification checklist

  • □ The product’s channel and users have been assessed and the mandatory scope of Arabic determined
  • □ The prominence of the Arabic-English bilingual layout meets the equivalence rule (font, colour, layout)
  • □ The outer packaging label carries the Arabic product name, local agent’s name and JFDA registration number
  • □ The checklist of mandatory label elements has been verified item by item
  • □ The IFU contains operation, warning, installation, maintenance, troubleshooting, disposal and EMC information
  • □ Both inner and outer packaging have been submitted as dossier annexes
  • □ The professionalism of the medical translation and the consistency of terminology have been re-checked
  • □ The cited provision numbers have been re-verified against the Arabic text of the Official Gazette

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Jordanian market. Our team prepares JFDA-compliant dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail