Israel Declaration Route: Registration Within 48 Hours

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Israel’s Declaration Route went live in September 2023, with formal guidelines published in January 2024. It applies to Class I devices (including sterile, measuring, surgical and reusable types) and IVD Class A laboratory equipment, on the condition that the product is already registered and marketed in a reference country. The system issues a technical confirmation email within 4 hours, and the registration confirmation arrives no later than 48 hours. Dossiers are organised and submitted according to 7 folders, and no government fees are charged for new registrations, amendments or renewals. The submission window for the annual renewal declaration runs from 1 October to 31 December of each year.

Taken together, these features make the Declaration Route the fastest and least burdensome entry point into the Israeli market for low-risk products that already hold approvals abroad. Its speed, however, rests entirely on the applicant preparing the dossier exactly as prescribed, because the automated confirmation process leaves little room for informal or incomplete submissions.

2. Legal basis

The route rests on a dedicated AMAR guideline supported by the parent regulations, as summarised below.

ElementContent
Title of instrumentGuidelines for Registering Medical Equipment – Declaration Route (Class I medical devices or low-risk IVD laboratory equipment)
Reference numberAMAR procedure guideline (issued January 2024)
PublicationJanuary 2024 (the Declaration Route was launched in September 2023)
ProvisionsDossier folder-structure clause (7 folders); acceptance and registration time-limit clause (4-hour receipt, 48-hour registration); annual declaration clause (1 October – 31 December)
Parent lawSection 8 of the Regulations 5773-2013 (notice for continued placing on the market of low-risk non-sterile devices), published on 12 June 2013
FeesSince 15 August 2022, no government fees are charged for registration, amendments or renewals

3. Expert insights

The 7-folder structure of the Declaration Route is a hard requirement, and the numbering must be preserved exactly:

  1. Declaration letter (choosing the corresponding declaration template according to “individual or legal entity” and “new registration, renewal or amendment”)
  2. Regulatory approvals (with four sub-folders: FDA 510(k) or registration listing, CE certificate, Declaration of Conformity, and approvals from other countries)
  3. Importer documents (ISO 9001 or GMP certificate of the IRH or the importer)
  4. Manufacturer documents (QMS certificate such as ISO 13485)
  5. Technical documentation (not mandatory)
  6. Catalogue (catalogue, instructions for use, labels)
  7. Cover letter (not mandatory)

Item 5 — technical documentation being non-mandatory — is the clause most easily misread. GRZAN’s interpretation is that not submitting the file does not mean the file need not exist. Although Class I manufacturers are not required to provide an ISO 13485 certificate, they must submit a declaration of conformity with the requirements of that standard; and although the technical documentation is not filed with the application, the registration holder must still keep the complete supporting documentation ready for post-market spot checks. In effect, the Declaration Route moves the burden of proof from “the moment of submission” to “the moment of inspection”.

The second point of interest is the content of the eight declarations. The declaration letter must state: the truthfulness of the materials and the commitment to import and sell the product in Israel; that the product is identical to the product approved in the reference country and is sold there; that the product and manufacturer comply with all regulatory and quality requirements of the reference country; that the holder will report changes in the materials to the Ministry of Health; that the holder possesses all supporting documentation; that materials will be transferred on request; that product changes will be reported; and that adverse events and recalls in Israel or worldwide will be reported. These eight items constitute a continuing obligations checklist for the IRH, not a one-time signature.

The practical weight of this list should not be missed. Because the declarations survive beyond the 48-hour confirmation, the holder’s obligations continue for as long as the registration does. A company that treats the declaration as a formality at submission, and then fails to report a later change or recall, has not merely filed an incomplete document once — it is in continuing breach of the very instrument on which its registration rests. Governance inside the company should therefore treat these eight commitments as inputs to the post-market compliance system, with named owners and reporting procedures behind each one.

4. Practical recommendations

  • Organise the ZIP file strictly by folder number: do not submit extra folders and do not alter the numbering — files that do not match the prescribed format are rejected outright by the system.
  • Build a standby technical documentation package: even though it is not submitted, prepare and retain it at the depth of a Standard Route dossier so it can be produced within 48 hours if a spot check occurs.
  • Batch the annual declarations: consolidate the renewals of all Class I devices and Class A IVDs and submit them in one action inside the 1 October window.
  • Run renewals and amendments through the same route: renewal materials resemble those for a new registration; for amendments only the relevant documents need to be filed, but this applies to non-substantial changes only.

5. Verification checklist

  • □ Classification confirmed as Class I or IVD Class A, and the product is registered and marketed in a reference country
  • □ ZIP file organised according to the 7-folder structure, with numbering matching the guideline
  • □ Declaration letter uses the correct template for “individual or legal entity” and “new registration, renewal or amendment”
  • □ Reference-country approvals, ISO 9001 (IRH or importer) and ISO 13485 (manufacturer) are all in place
  • □ Standby technical documentation package established and quickly retrievable
  • □ Annual declaration reminder set for 1 October

Regulatory information verified as of 29 October 2026. The names, numbers, publication dates and article numbers cited in this article were checked against official Israeli legal databases and authoritative public sources on that date; AMAR procedure documents are updated frequently, so please rely on the current version published on the Ministry of Health website before formal submission.

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Israeli market. Our team manages MOH registration applications across risk classes, prepares reference-country dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail