Risk Classes 1, 2a, 2b and 3 for Medical Devices in Russia
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
The Russian potential risk application class (klass potentsialnogo riska primeneniya) has four levels: 1, 2a, 2b, and 3. The class is not a labeling nicety; it determines three concrete things. First, it decides whether the testing scope includes toxicological studies and human clinical trials. Second, it decides whether a manufacturing site inspection is required within the state registration procedure. Third, it sets the tier of the state fee for expert review. In short, the class is the main input to both the timeline and the budget of a Russian registration project, and an error at this stage propagates through every later step.
Under the national pathway, the class is determined according to the catalogue of Order No. 4n of the Ministry of Health. Under the union pathway, it is judged according to EEC Decision No. 173. The two systems correspond at the level of classes, but the details of determination differ, and the same product can legitimately receive different treatment in each framework. Manufacturers operating in both systems should treat classification as a dual exercise, performed twice with two legal references, rather than assuming that one analysis serves both routes.
2. Legal basis
| Legal act | Number | Date | Provisions |
|---|---|---|---|
| Order of the Ministry of Health On the Nomenclature Classification of Medical Devices | Order of the Ministry of Health of Russia No. 4n | June 6, 2012 (as amended) | mapping of type codes to classes |
| Decision of the Board of the Eurasian Economic Commission On Classification Rules Based on Potential Risk | EEC Decision No. 173 | December 22, 2015 (as amended by No. 65 of 2023 and No. 18 of 2025) | classification rules for non-invasive, invasive, implantable, and active devices |
| Government Resolution of the Russian Federation On the State Registration Rules for Medical Devices | Government Resolution No. 1684 | November 30, 2024 | para. 5 (form of conformity assessment determined by class) |
| Tax Code of the Russian Federation | Article 333.32.2 of the Tax Code of the Russian Federation | current version | state fees for expert review, tiered by class |
3. Expert insights
Insight 1: The class is an input to the cost function. The expert review fee is tiered by class. The current practical estimate is 72,000 rubles for class 1, 104,000 rubles for class 2a, 136,000 rubles for class 2b, and 184,000 rubles for class 3. If the same device is assigned different classes in the two databases, the consequence is not only a difference in fees but also a difference in the testing program: a higher class may pull toxicological studies or clinical trials into scope that a lower class would not require. Budget models should therefore treat the class as a variable to be verified in both frameworks, not as a single fixed assumption copied from another market.
Insight 2: Contact time and invasiveness are the core variables. The union classification rules use non-invasive devices, invasive devices of short-term, transient, or long-term duration, and implantable devices as the main axis, with the contact site serving as an escalation trigger, for example the central circulatory system or the central nervous system. Active implantable devices intended to come into direct contact with the heart, the central circulatory system, or the central nervous system fall into class 3. The dwell threshold in the definition of an implantable device is 30 days or more. Reading these rules against the actual anatomy of use, rather than against generic product descriptions, is what produces defensible classifications.
Insight 3: The 2025 amendment changed the determination approach for software and IVD. EEC Amendment Decision No. 18 (February 18, 2025) aligns the treatment of software-based devices and in vitro diagnostic devices with the approaches of the IMDRF documents, and repeals the separate rules on distinguishing device elements and merging multiple models contained in Decision No. 116 of 2018 and Decision No. 123 of 2018. This means that the class of the same software product under the new approach may differ from the previous judgment. Legacy classification conclusions should not be carried forward without review.
4. Practical recommendations
- Support every classification conclusion with a written argument: list each determination basis for the product, including contact site, duration, invasiveness, implantability, and whether the device is active, rather than merely citing the catalogue result.
- When the same product is placed in both Russia and the EAEU, perform the determination twice, once under Order No. 4n and once under Decision No. 173, and compare the outcomes explicitly in the project record.
- Re-verify the class of software and IVD products after the 2025 amendment; conclusions issued under the previous approach should not be reused without a documented review.
- Recalculate the expert review fee tier against the confirmed class at project start and again before filing, so that budget assumptions match the amounts actually charged.
5. Verification checklist
- □ The type code and recommended class have been obtained from the catalogue of Order No. 4n
- □ A cross-determination has been completed under Decision No. 173 (including Amendment No. 18)
- □ The four determination bases of contact site, duration, invasiveness, and active status have been documented in writing
- □ The expert review fee tier has been estimated from the class and the amounts re-checked against the current schedule
- □ The class conclusions for software and IVD products have been re-verified after the 2025 amendment
GRZAN provides risk classification services for the Russian and EAEU frameworks, including written classification arguments aligned with Order No. 4n and EEC Decision No. 173, dual-framework comparison for products placed in both systems, and fee and testing scope planning based on the confirmed class. Our classification records are structured for reuse in later amendments and union conversion projects, saving clients repeated analysis work. For classification and registration support in Russia, contact GRZAN at info@grzanglobal.com.
Regulatory information verified as of October 29, 2026.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us