Uzbekistan Medical Device Registration: Sterility and Biological Safety Documentation

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Annex 3a of Resolution No. 738 covers sterility and biological safety through four items: item 2.8 sterilization process information, including process validation, microbiological test results or bioburden levels, bacterial endotoxins, the sterility test method, and packaging validation details, applicable to sterile products; item 2.14 stability study reports, where products with a shelf life must attach a national-language translation of the test results and conclusions; item 2.16 the protocol of biological action evaluation studies (tests); and item 2.17 biological safety information. The four items are independent of one another and none can be omitted: submitting only a sterility test report while lacking sterilization process validation is the most common reason for a request to supplement the dossier. Because each item answers a different question about the same product, they should be planned as one evidence program covering process, shelf life, and patient contact rather than as four disconnected documents.

2. Legal basis

RegulationNumberDateProvision
Cabinet Resolution on approving the Regulation on the procedure of state registration of medicines and medical devices and issuance of registration certificatesNo. 738Adopted November 24, 2025, effective February 26, 2026Item 2.8: sterilization process information, including process validation, microbiological test results (bioburden levels), bacterial endotoxins, the sterility test method, and packaging validation details, applicable to sterile products
Cabinet Resolution on approving the Regulation on the procedure of state registration of medicines and medical devices and issuance of registration certificatesNo. 738Adopted November 24, 2025, effective February 26, 2026Item 2.14: stability study report; products with a shelf life must attach a national-language translation of the test results and conclusions
Cabinet Resolution on approving the Regulation on the procedure of state registration of medicines and medical devices and issuance of registration certificatesNo. 738Adopted November 24, 2025, effective February 26, 2026Item 2.16: protocol of the biological action evaluation studies (tests) performed to demonstrate the device’s conformity with the general biological requirements
Cabinet Resolution on approving the Regulation on the procedure of state registration of medicines and medical devices and issuance of registration certificatesNo. 738Adopted November 24, 2025, effective February 26, 2026Item 2.17: biological safety information (items 2.16 and 2.17 do not apply to software and in vitro diagnostic devices)

3. Expert insights

Insight 1: sterility evidence has two layers, process and result, and only one is usually submitted. The wording of item 2.8 includes both process validation, meaning validation of the sterilization process itself, and results, meaning bioburden levels, bacterial endotoxins, and sterility testing. Process validation demonstrates that the process reliably achieves sterility; results demonstrate that a given batch is sterile. Where only batch release testing is provided and the process validation is missing, the reviewer cannot judge whether conformity will continue across future production, because a single passing batch proves nothing about the stability of the process that produced it. Both layers belong in the dossier for every sterile product.

Insight 2: packaging validation is part of the sterility evidence, not a packaging design file. Item 2.8 places packaging validation details under the same item as sterilization information, because packaging determines whether the sterile barrier is maintained. What packaging validation must demonstrate is that the packaging maintains the sterile state throughout the claimed shelf life, which is why it supports and is supported by the stability studies under item 2.14. Treat the sterile barrier system as a tested component of the product rather than an ancillary design choice, and the two items will reinforce each other in review.

Insight 3: the translation requirement for stability studies is precise down to results and conclusions. Item 2.14 requires products with a shelf life to submit the stability study report together with a national-language translation of the test results and conclusions. This wording deliberately limits the translation workload to the concluding content, while raw data tables may remain in the original language. Manufacturers can use this to control translation costs, but the conclusions must never be omitted, because they carry the reviewer’s decision on the granted shelf life. The shelf life printed on the label must remain within what the translated conclusions support.

Insight 4: the biological evaluation items do not apply to software and IVDs, but do apply to most contacting devices. Item 2.16, the biological action evaluation study protocol, and item 2.17, biological safety information, are explicitly stated not to apply to software and in vitro diagnostic devices. This means patient-contacting devices, implants, and devices containing biological materials all need both items, while standalone software and IVDs do not. Checking this exemption line by line, rather than by analogy, avoids both gaps and unnecessary work; a device with indirect patient contact still generally falls within scope, so the contact pathway should be documented.

4. Practical recommendations

  1. For sterile products, prepare the evidence in three layers: process validation, batch results, and packaging validation, with no omission of the process documentation.
  2. Align packaging validation with the stability studies so the sterile barrier is demonstrated across the claimed shelf life.
  3. Translate only the test results and conclusions of the stability study, retaining raw data in the original language, satisfying compliance while controlling cost.
  4. Prepare biological evaluation evidence under items 2.16 and 2.17 for patient-contacting, implantable, and biological-material devices.
  5. Keep the labeled shelf life consistent with the stability study conclusions, never beyond what the research supports.

5. Verification checklist

  • □ Sterile products include sterilization process validation documentation
  • □ Results are submitted for bioburden levels, bacterial endotoxins, and the sterility test method
  • □ Packaging validation details are submitted and aligned with the stability study conclusions
  • □ Products with a shelf life include the stability report and national-language translation of results and conclusions
  • □ Patient-contacting and implantable devices include the biological evaluation protocol and biological safety information
  • □ The labeled shelf life matches the stability study conclusions

GRZAN supports manufacturers, importers, and authorized representatives with medical device registration in Uzbekistan, including sterility evidence under item 2.8, stability documentation under item 2.14, and biological evaluation files under items 2.16 and 2.17. For assistance, contact info@grzanglobal.com.

Regulatory information verified as of October 29, 2026.

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail