Kyrgyzstan Medical Device Risk Classification: The EEC Decision No. 173 Rules

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Kyrgyzstan has no classification rules of its own for medical devices. Paragraph 1 of Article 2 of Cabinet Resolution No. 513 expressly provides that the classification of medical devices by potential risk of use follows the Rules for the Classification of Medical Devices by Potential Risk of Use approved by Decision No. 173 of the Executive Committee of the Eurasian Economic Commission dated December 22, 2015. The classes are 1, 2a, 2b and 3. The practical meaning of this arrangement is that even on the national track, risk is measured with the Union ruler: the classification outcome determines the scope of testing, the inspection requirements and the tier of official fees.

2. Legal basis

  1. Cabinet Resolution No. 513 of August 27, 2024, paragraph 1 of Article 2: classification follows Decision No. 173 of the Executive Committee of the Eurasian Economic Commission of December 22, 2015.
  2. Paragraph 2 of Article 2 of Cabinet Resolution No. 513: safety and effectiveness requirements follow Council Decision No. 27 of the Eurasian Economic Commission of February 12, 2016, with the exception of item 63.
  3. Paragraph 3 of Article 2 of Cabinet Resolution No. 513: biological effects evaluation studies follow Council Decision No. 38 of May 16, 2016.
  4. Paragraph 4 of Article 2 of Cabinet Resolution No. 513: technical testing follows Council Decision No. 28 of February 12, 2016.
  5. Paragraph 5 of Article 2 of Cabinet Resolution No. 513: clinical and clinical laboratory trials follow Council Decision No. 29 of February 12, 2016.
  6. Article 14 of Law No. 216: registration requires a positive expert evaluation of safety, quality and effectiveness, taking into account the potential risk level of use of the device.

3. Expert insights

Insight 1: a national procedure combined with Union technical rules is the structural feature of this market.

Article 2 of Resolution No. 513 transfers five Union decisions in one breath, covering classification, general requirements, biological effects, technical testing and clinical trials. This is not legislative inertia but deliberate design: the procedure stays domestic to preserve control, while the technical requirements are aligned with the Union to preserve portability. The most direct consequence for companies is that most of the technical evidence prepared under the national track can be reused when the dossier migrates to the Union track after 2028. This is precisely why taking the national track now remains a rational choice during the transition period.

Insight 2: Decision No. 27 is applied with the exception of item 63, and this carve-out deserves a separate flag.

Paragraph 2 of Article 2, when transferring the general safety and effectiveness requirements, expressly excludes item 63. A carve-out inside a transfer provision is usually the legislator’s signal that one specific requirement is reserved on its local applicability. When building the conformity evidence table, verify Decision No. 27 item by item and flag item 63 separately as not applicable — this avoids being asked during review for evidence that is not needed, and also avoids evidence gaps caused by overlooking the exclusion.

Insight 3: the classification outcome is the common upstream of three cost lines.

The risk class is not a one-off label; it is the starting point of three branches:

  • Scope of testing: the class determines the necessity of technical testing, biological effects evaluation and clinical trials;
  • Inspection requirements: devices in class 2a (sterile), 2b and 3 trigger production site inspections;
  • Official fee tiers: state procedure fees are set in four tiers for classes 1, 2a, 2b and 3 (amounts should be confirmed against the payment notice issued by the competent authority; see Article 34 of this series).

If the classification is wrong, all three branches must be redone. Classification is the highest-return judgment in the entire project; it should be done first, and it should be done as a written argumentation.

Insight 4: classification disputes have a formal resolution channel.

As introduced in Article 02 of this series, one of the functions of the Expert Community is to handle disputes over classification by potential risk level and by type and to refer matters on which no consensus is reached to the consultative committee at the Union level. If a company and the receiving department disagree on classification, this mechanism can be used rather than passively accepting the outcome.

4. Practical recommendations

First, turn the classification argumentation into a reusable standalone document. We recommend preparing a classification justification sheet for every product: compare the criteria of Decision No. 173 one by one (intended purpose, duration of use, invasiveness, mode of contact with the body, whether vital organs are affected) and state the conclusion and its basis. This document is repeatedly used in three stages: testing plan design, inspection application, and change assessment.

Second, for borderline products, seek confirmation before submission. The classification of combination products, drug-device combinations, products with both diagnostic and therapeutic functions, and software products often leaves room for interpretation. We recommend seeking written confirmation from the receiving department before submission and archiving the reply.

Third, build the evidence table item by item against Decision No. 27, and mark item 63 as not applicable. Do not map entries directly from domestic habits or from Annex I of the EU MDR — different entry systems produce confusion in cross-referencing.

Fourth, build a change trigger mechanism around the classification outcome. Any change in intended purpose, duration of use or invasiveness may change the risk class. The change management procedure should include a preliminary check on whether a change affects the classification (see Article 35 of this series).

5. Verification checklist

  • □ Obtain the current text of Executive Committee Decision No. 173 of the Eurasian Economic Commission (December 22, 2015) and compare the criteria item by item;
  • □ Prepare a written classification justification for every product to be registered, stating the class among 1 / 2a / 2b / 3;
  • □ Confirm that the classification outcome is consistent with the testing plan, the inspection application and the official fee tier;
  • □ Build the conformity evidence table against the entries of Decision No. 27 (February 12, 2016), with item 63 separately marked as not applicable;
  • □ For borderline products, written confirmation has been obtained from the receiving department and archived;
  • □ The change management procedure includes a preliminary check on whether a change affects the risk class.

Regulatory information verified as of October 29, 2026.

Our services cover pre-market pathway selection and feasibility assessment, registration dossier preparation and submission, authorized representative arrangements, local testing and clinical trial organization, production site inspection coordination, quality management system conformity support, post-market surveillance and device vigilance system setup, import and distribution compliance, official fee calculation and project budget management.

The GRZAN regulatory team consists of professionals with many years of hands-on experience in Eurasian Economic Union registrations. We maintain a stable local cooperation network in Bishkek, Almaty, Tashkent and Moscow, work in Chinese, Russian and English, and provide executable solutions for the issues foreign manufacturers most often encounter in CIS markets, including authorized representative arrangements, the notarization and legalization chain for powers of attorney, localization of technical documentation, and inspection readiness.

We adhere to one principle: every regulatory conclusion we publish must rest on verifiable official sources. Figures and timelines that cannot be verified are always marked as practical estimates, and experience-based values are never presented as legal requirements.

GRZAN provides registration support in Kyrgyzstan for medical device manufacturers, importers and authorized representatives. For project consultation on Kyrgyzstan or Eurasian Economic Union market access, contact us at info@grzanglobal.com.

GRZAN | Global Medical Device Regulatory Registration and Compliance Services

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail