Kazakhstan Medical Device Classification: Dual-Source Rules and the 2025 Amendment
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Kazakhstan’s medical device classification rules have two sources. The national route applies Order No. ҚР ДСМ-281/2020 of the Ministry of Health (2020, effective 10 calendar days after first official publication); the union route applies Decision No. 173 of the Board of the Eurasian Economic Commission (December 22, 2015, as amended by Decision No. 65 of May 23, 2023 and Decision No. 18 of February 18, 2025). Both rule sets use the same four-class structure (1 / 2а / 2б / 3), but the detailed rules differ, and the 2025 amendment in Decision No. 18 specifically adjusted the treatment of software and in vitro diagnostic devices. A classification conclusion must state the version of the rules on which it relies; without that statement it cannot pass review. Because classification determines testing scope, inspection triggers and post-market obligations, an error here propagates through the entire compliance chain rather than staying contained in one document. In practice, the classification memo is one of the first documents reviewers open, because every downstream document cites its conclusion.
2. Legal basis
| Instrument | Reference number | Date | Provisions |
|---|---|---|---|
| Order of the Minister of Health of Kazakhstan approving the classification rules for medical devices | No. ҚР ДСМ-281/2020 | 2020 | full classification rules; effective 10 calendar days after first official publication |
| Decision of the Council of the Eurasian Economic Commission on the classification rules for medical devices depending on the potential risk of application | Decision No. 173 (as amended by Decision No. 65 and Decision No. 18) | December 22, 2015; amended May 23, 2023 and February 18, 2025 | union four-class rules; updated treatment of software and IVDs |
| Order of the Minister of Health of Kazakhstan approving the Rules for National Registration, Re-registration and Changes to Registration Dossiers of Medicines and Medical Devices | No. ҚР ДСМ-16 | February 9, 2021 (Ministry of Justice registration No. 22175) | the place of classification conclusions in the dossier |
| Decision of the Board of the Eurasian Economic Commission approving the Rules for Registration and Expert Evaluation of the Safety, Quality and Efficacy of Medical Devices | Decision No. 46 | February 12, 2016 | application of classification on the union route |
3. Expert insights
Insight one: fix the pathway first and then the classification; the order cannot be reversed. The choice of classification rules follows from the registration route, not the other way around. Companies sometimes classify under the rules they know best and only then select the route, which produces a mismatch between the classification basis and the filing procedure that must be redone before submission. The correction is cheap at the planning stage and expensive after the dossier has been assembled, so the route decision should be captured in writing before any classification work begins. Recording the route decision and the named classification rules in the same memo keeps both decisions auditable together.
Insight two: Decision No. 18 of February 18, 2025 is the most recent substantive adjustment. The union-side rules were amended once in 2023 and once in 2025, and the 2025 amendment targets software and in vitro diagnostic devices. Every classification conclusion formed before 2025, especially for companion diagnostics, stand-alone software and self-test IVDs, should be re-checked under the new treatment before it is cited in a fresh submission, because reviewers will apply the current text and not the version the company used. The re-check should also cover accessories and combination products, whose classification can change under the amended software rules.
Insight three: a class conclusion must state basis, clause and version. Writing only that the product is class 2а is not sufficient. The classification rationale should state whether Order No. ҚР ДСМ-281/2020 or Decision No. 173 is relied on, which specific rule applies, and the date of the version used. Reviewers check exactly these three items, and a missing version date is one of the most easily avoided remedy triggers in an otherwise complete dossier. Where both routes are planned for the same product, prepare two classification memos rather than one compromise text that serves neither well.
Insight four: the overlap between the two rule sets is the foundation for reusing national and union dossiers. The two systems reach highly consistent outcomes for conventional active and non-active devices; the differences concentrate on software, IVDs and drug-device combinations at the frontier categories. GRZAN therefore plans evidence packages on a strategy of reuse for conventional categories and dual assessment under both rule sets for frontier categories, which keeps the dossier work honest without doubling the effort where it is not needed.
Insight five: a classification conclusion shapes the whole compliance chain, not only the depth of review. The class determines the scope of technical testing and toxicological studies, whether a manufacturing site inspection is triggered and how wide its coverage is, and the intensity of post-market surveillance. The cost of a classification error is global rather than local: it resurfaces in every downstream document built on top of it.
4. Practical recommendations
- State three elements in the classification rationale: the name of the rules relied on, the applicable clause, and the version date.
- After Decision No. 18 of February 18, 2025, re-check existing conclusions for software and IVD categories once.
- Handle conventional categories with national and union reuse, and frontier categories with dual assessment under both rule sets.
- Cite the classification conclusion consistently in the test plan, the inspection preparation and the surveillance plan.
- Fix the classification basis at the evidence-package planning stage to avoid redoing work after a mid-course route change.
5. Verification checklist
- □ The registration route has been determined and the classification rules selected accordingly
- □ The classification rationale contains three elements: rules name, applicable clause, version date
- □ Software and IVD conclusions have been re-checked under Decision No. 18 of 2025
- □ The classification conclusion is cited consistently in the test plan, inspection preparation and surveillance plan
- □ The reuse or dual-assessment strategy for conventional and frontier categories is defined
- □ The dossier adjustment path in case of a classification change is predefined
GRZAN provides registration support in Kazakhstan for medical device manufacturers, importers and authorized representatives, covering dual-source classification, dossier preparation, expert review follow-up and post-grant registration maintenance. For assistance with this topic, contact info@grzanglobal.com.
Regulatory information verified as of October 29, 2026.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us