Kazakhstan Medical Device Import: Customs Gates Beyond National Registration
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Obtaining a Kazakhstan national registration certificate resolves only one gate in the customs clearance chain. Medical equipment with electrical and measurement functions must pass two more gates: the metrology track, with type approval and state registration under the law on ensuring the uniformity of measurements, and the technical regulation track, covering TR CU 004/2011 on low-voltage equipment safety and TR CU 020/2011 on electromagnetic compatibility, plus, depending on the product, TR CU 010/2011 on machinery safety, TR CU 012/2011 on explosion safety and TR EAEU 037/2016 on restriction of hazardous substances. The three tracks do not substitute for one another, they are administered by different bodies, and they must be advanced in parallel. As a practical estimate, sequencing the tracks one after another is the single most common cause of delayed market entry, because each track has its own lead time and its own renewal logic.
2. Legal basis
| Regulation | Number | Date | Provisions |
|---|---|---|---|
| Law of the Republic of Kazakhstan on Ensuring the Uniformity of Measurements | — | current version | Type approval and state registration of measuring instruments |
| Union technical regulation on safety of low-voltage equipment | TR CU 004/2011 | current | Electrical equipment from 50 to 1000 V AC and from 75 to 1500 V DC |
| Union technical regulation on electromagnetic compatibility of technical equipment | TR CU 020/2011 | current | Electromagnetic compatibility requirements |
| Union technical regulation on safety of machinery and equipment | TR CU 010/2011 | current | Machinery safety |
| Union technical regulation on safety of equipment for potentially explosive environments | TR CU 012/2011 | current | Explosion safety |
| Union technical regulation on restriction of hazardous substances in electrical and electronic equipment | TR EAEU 037/2016 | current | Restriction of hazardous substances |
| Order of the Minister of Healthcare approving the Rules for State Registration, Re-registration and Changes to Registration Dossiers of Medicines and Medical Devices | No. ҚР ДСМ-16 | February 9, 2021 (Ministry of Justice registration No. 22175) | Registration of medical devices |
3. Expert insights
Insight one: three tracks in parallel, and none can be missing. The registration certificate proves that the product may circulate as a medical device; type approval proves that measurement data are trustworthy; EAC conformity documents prove that electrical and mechanical safety requirements are met. The objects of assessment and the legal bases are entirely different, and any missing item can stop the product at customs clearance or during market surveillance. The three answers answer three different questions asked by three different authorities, which is why one certificate can never serve as evidence for all of them.
Insight two: the metrology track is judged by use and covers seven categories. National metrological control in Kazakhstan covers seven categories of measuring equipment, including measuring instruments for medical diagnosis. The test is the measurement purpose of the device in medical diagnosis, not the product name. Blood pressure monitors, glucose meters, medical analyzers and thermometers are typical products falling within scope. The assessment should be made per measurement function, since a multi-parameter device may bring several measurement functions under the metrology track at once.
Insight three: the technical regulation track is judged by electrical and mechanical characteristics. TR CU 004/2011 applies to electrical equipment powered at 50 to 1000 V AC or 75 to 1500 V DC; TR CU 020/2011 applies to devices with electronic control, displays, data transmission or wireless communication modules. Most medical electrical devices fall within both at once, and machinery safety under 010, explosion safety under 012 and hazardous substance restrictions under 037 may also apply according to construction. The determination should be documented per regulation with the triggering parameter cited, so the scope decision can be defended if the authority or the customer questions it.
Insight four: technical documentation can be reused, but reports cannot be swapped. Circuit diagrams, risk analyses and test plans can be shared across the three tracks, but conformity assessment reports must be issued separately for each. Submitting an EAC certificate in place of the technical test report for registration is a common cause of rework. Planning the documentation architecture at the start, with a shared core and track-specific annexes, converts this risk into a routine filing exercise.
Insight five: metrology requirements continue into the post-market phase. After certification, measuring instruments remain subject to periodic verification, which runs in parallel with post-market monitoring of the medical device. A failed metrological verification during annual hospital checks feeds back into the compliance status of the device, so the two sets of records should be cross-checked regularly. A failed verification found during a hospital audit is far more expensive to fix than a scheduled verification, which is why the practical estimate for after-market support must include the verification calendar from the first day of sales.
4. Practical recommendations
- At project initiation, check each device against a three-track list: whether registration is needed, whether it is a measuring instrument, and which technical regulations apply.
- Advance the three tracks in parallel, sharing the same technical documentation and samples, but issue reports separately.
- Record the metrology use assessment in the product definition statement, consistent with the intended use wording.
- Judge the applicability of technical regulations item by item on electrical parameters and construction features, never by default.
- After market entry, maintain a verification log and cross-check it regularly with the adverse event monitoring log.
5. Verification checklist
- □ Each device has completed the three-track check covering registration, metrology and technical regulation
- □ Type approval and state registration of measuring instruments have been initiated
- □ Applicable Union technical regulations have been assessed item by item, covering 004, 020, 010, 012 and 037
- □ Reports for the three tracks have been issued separately, without substituting EAC documents for the registration technical test report
- □ The metrology use wording is consistent with the intended use statement
- □ A post-market verification log has been established and is compared with the monitoring log
Regulatory information verified as of October 29, 2026.
GRZAN provides registration support in Kazakhstan for medical device manufacturers, importers and authorized representatives, including multi-track compliance planning across registration, metrology and technical regulation. Track applicability is documented per device before any application is filed. For inquiries, contact info@grzanglobal.com.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us