Panama Risk Classification A/B/C/D under DE 490 Article 30

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Panama classifies medical devices into four classes by risk: class A (low risk), class B (low to moderate risk), class C (moderate to high risk) and class D (high risk). The legal basis is Article 30 of Decreto Ejecutivo N° 490, not Article 15, which deals with the prohibition on operating a device business inside a dwelling. The classification rules must apply the most current version of the GHTF and IMDRF risk classification (Grupo de Trabajo de Armonización Global / International Medical Device Regulators Forum) or the documents of a successor body, and the DNDM is the authority for risk classification, with binding decisions (Article 16 of Ley 90). The claim circulating in Chinese-language material that Panama uses classes I, II and III is wrong.

  • Four classes, lettered A to D, defined in Article 30 of DE 490.
  • The classification rules are a living reference to the current GHTF and IMDRF version, not a frozen annex.
  • The manufacturer’s own declaration is the starting point of the class, which places responsibility on the applicant.
  • Class drives document volume, review format and the phase-in deadline, so it is the highest-leverage decision in a project.

2. Legal basis

Article 30 of DE 490 defines the classes and the rules for choosing between them; Article 16 of Ley 90 fixes the authority; Articles 31, 33 and 37 then attach consequences to the class. The consequences are what make the classification decision commercially significant rather than a labelling exercise.

Decreto Ejecutivo N° 490 (4 October 2019, Gaceta 28875-A)Article 30Devices are classified according to the level of risk associated with their use and the vulnerability of the human body, and according to the manufacturer’s declaration, into 1. class A: low risk; 2. class B: low to moderate risk; 3. class C: moderate to high risk; 4. class D: high risk
Decreto Ejecutivo N° 490Article 30, second paragraphDetermining the appropriate class requires applying the most current version of the GHTF and IMDRF risk classification or the rules of a successor body
Decreto Ejecutivo N° 490Article 30, third paragraphWhere a product can fall into two classes at the same time, the higher of the two is taken
Ley N° 90 (26 December 2017, Gaceta 28433-B)Article 16The procedure and requirements for a device sanitary registration are set according to the risk class and the manufacturer’s recommendation; the most current GHTF and IMDRF classification is applied; the DNDM is the authority for classification and its decisions are binding
Decreto Ejecutivo N° 490Article 33The higher the risk, the more complex the assessment and the stricter the requirements
Decreto Ejecutivo N° 490Article 37Special requirements accumulate class by class from A to D
Decreto Ejecutivo N° 490Article 31Classes C and D require the involvement of at least 2 specialist users of the relevant discipline in the assessment; for classes A and B this is optional

3. Expert insights

Insight one: the reference to the most current GHTF and IMDRF version is a living clause. Article 30 of DE 490 and Article 16 of Ley 90 both use the phrase most current version. The direct consequence is that Panamanian classification rules move when the IMDRF documents move, without any amendment to DE 490 itself. A class that was assigned in another country under an older rule set may therefore not hold in Panama. At the classification stage, state the version of the IMDRF document relied on, rather than writing that the product was classified according to GHTF rules in general terms.

Insight two: classification follows the manufacturer’s declaration, which puts the first responsibility on the company. The opening sentence of Article 30 ties the class to the risk associated with use and to the vulnerability of the human body, and to what the manufacturer declares. Combined with Article 33, under which all information submitted rests on a sworn declaration, the company carries responsibility for the truth of the class it declares. A false declaration leads directly to the cancellation grounds in Article 67 of DE 490 and to the very serious infringement in Article 37 of Ley 90. Classification is a legal liability point, not a technical formality.

Insight three: because the DNDM decision is binding, the dispute route is constrained. The closing sentence of Article 16 of Ley 90 makes the DNDM the authority for risk classification and makes its decisions binding. A company that disagrees with a class cannot fall back on industry practice; the argument has to be made inside the DNDM process. The more effective course is to make the classification argument solid before submission rather than to contest it afterwards, because a contested file has already lost the review time that a well-argued one keeps.

Insight four: the class setting moves three cost lines at once. The first is document volume, since Article 37 adds special requirements step by step from A to D. The second is the review format, since Article 31 requires at least two specialist physicians for classes C and D while leaving that optional for A and B, which changes review duration. The third is the transitional period, since Article 23 of Ley 90 sets 18, 24, 30 and 42 months for classes D, C, B and A respectively. Of all the decisions in a Panamanian project, the class has the highest leverage.

4. Practical recommendations

  • In the classification rationale, name the GHTF or IMDRF document version and the specific rule numbers applied. Do not accept a formulation that rests on customary practice.
  • Where two classes are plausible, prepare the file for the higher class immediately rather than submitting at the lower class and being asked to supplement.
  • For class C and D projects, reserve time in the schedule for the specialist assessment and do not count it as running in parallel with document preparation.
  • Keep the class consistent across the intended use, the label and the instructions for use; drift in any one of the three invites a fresh classification.
  • Treat a change of class as a substantial change and discharge the notification duty under Article 43 of DE 490.

5. Verification checklist

  • □ Confirm the classification provision is Article 30 of DE 490, not Article 15
  • □ Confirm the four classes are A (low), B (low to moderate), C (moderate to high) and D (high)
  • □ Record the version of the GHTF or IMDRF document relied on
  • □ Check whether the product can fall into two classes at once (Article 30, third paragraph)
  • □ Confirm whether class C or D triggers the specialist assessment (Article 31)
  • □ Check the special requirements for the class against Article 37

Regulatory information verified to 29 October 2026.

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Panamanian market. Our team manages Ministry of Health (MINSA) sanitary registrations through the DIGEMID-equivalent medical device directorate, prepares Spanish-language dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail