Panama and Guatemala Compared: Central America Six-Country Device Rules
Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Panama and Guatemala are both in Central America, but device regulation there consists of two separate systems that cannot be substituted for one another. In Panama, the authority is DNDM within MINSA under Article 3 of Ley 90; risk is graded A, B, C and D under Article 30 of DE 490 following the latest GHTF and IMDRF rules, with the higher class applying where a product meets two; the working credential is the technical standard certificate (CCT), valid five years, free and mandatory for public procurement under Article 21 of DE 490, while the start of sanitary registration remains with the health authority under Article 18 of Ley 92; devices are exempt from import duty under Article 28 of Ley 90; and vigilance includes a 72-hour serious adverse event report under Article 56 of DE 490.
In Guatemala, the authority is DRCPFA under MSPAS and above it DGRVCS; classification has four categories, I to IV, built on the axis of use combined with sterility under Annex 1 of NT-37; the credential is the Inscripción Sanitaria, valid five years, after which a fresh application under Article 5 of NT-37 is required rather than a renewal; the registrant must be a domestic establishment holding a Licencia Sanitaria; and the current technical standards set no device vigilance reporting deadline.
Both countries are in transition: Panama through WTO notification PAN/135 of 8 December 2025, Guatemala through GTM/106, notified on 11 February 2026, proposed to take effect on 1 December 2026 and to repeal Government Agreement 712-99. Costa Rica, Nicaragua, Honduras and El Salvador await verification in their own volumes and are not inferred here.
2. Legal basis
The comparison below is organised by function rather than by instrument, so that equivalent provisions can be read side by side. Each row carries the source for both jurisdictions, because the two systems reach comparable outcomes by different legal techniques.
| Dimension | Panama | Panama source | Guatemala | Guatemala source |
| Authority | DNDM, National Directorate of Medical Devices | Ley 90, Articles 2 and 3 | DRCPFA, under DGRVCS | Acuerdo Gubernativo 115-99, Article 35; NT-37 and NT-40 |
| Superior law | Ley No. 90 (2017-12-26), amended by Ley No. 92 (2019-09-12) | Constitution Article 109, then Ley 90, then DE 490 | Health Code Decreto 90-97, Article 162 et seq.; Government Agreement 712-99 | Constitution Article 96, then Health Code 90-97, then 712-99, then NT |
| Risk classification | A, B, C, D, risk based | DE 490, Article 30 (latest GHTF and IMDRF rules; higher class applies) | I, II, III, IV, on the axes of use and sterility | NT-37, Annex 1 (USO EXTERNO or INTERNO by NO ESTERIL or ESTERIL) |
| Reagent classification | Same A to D system as devices | DE 490, Article 30 | Categoría I, II and III sanitary risk | NT-40, Article 3 |
| Main credential | CCT, Certificado de Criterio Técnico | DE 490, Article 21 | Inscripción Sanitaria | NT-37, Article 5 |
| Credential validity | Five years, free, mandatory for public procurement | DE 490, Article 21 | Five years; late renewal requires a new application | NT-37, Article 9 |
| Operating licence | Five categories (Fábrica, Acondicionadora, Distribuidora, Minorista, Almacén), five years, renewal 60 working days before expiry | DE 490, Articles 13 and 14 | Licencia Sanitaria for the establishment | NT-37, Article 5.1.13 |
| Holder | Titular del producto is the manufacturer; titular del registro sanitario is the certificate signatory; the proveedor needs an accredited link | DE 490, Article 2 definitions 90 and 91; Ley 90, Article 6 definition 37 | Registrant must be a domestic establishment with a Licencia Sanitaria; foreign manufacturers cannot hold | NT-37, Article 5 and Volume 08 quick table |
| Import taxation | Devices and biomedical equipment exempt from import duty; CSS and MINSA also exempt from ITBMS | Ley 90, Article 28 | No equivalent exemption identified; no inference made | — |
| Vigilance deadline | Serious adverse event reported urgently within 72 hours | DE 490, Article 56 | Current NT sets no device vigilance reporting deadline | Guatemala volume, article 43 |
| Inspection and measures | Unannounced inspections, temporary closure, seizure or sealing; investigation of 90 working days or less | Ley 90, Articles 30, 31, 32, 41 and 43 | Market surveillance under NT-36 v6-2023, in force 2023-07-21; test failure cancels the registration, with withdrawal within 15 days | NT-36 v06-2023; NT-37 Annex 2 note |
| Recent change | WTO notification PAN/135 (2025-12-08; adoption pending) | WTO TBT | WTO notification GTM/106 (2026-02-11; proposed adoption 2026-06-01; effect 2026-12-01, repealing 712-99) | WTO TBT |
3. Expert insights
Insight 1: the classification axes differ, so class numbers cannot be converted. Article 30 of the Panamanian decree is a risk-based A to D system following the latest GHTF and IMDRF rules, with the higher class prevailing. Annex 1 of NT-37 builds four categories from the axis of use and sterility. A product classified as B in Panama may carry a different class in Guatemala, with no conversion formula. Copying one class into the other’s form is the most common technical error in multi-country projects.
Insight 2: the credentials differ in legal character. The CCT is a transitional credential valid five years, free and mandatory for public procurement, while the start of the final sanitary registration remains with the health authority under Article 18 of Ley 92. The Inscripción Sanitaria is a final registration valid five years, but Article 9 of NT-37 provides that expiry does not lead to renewal and that a fresh application under Article 5 is needed. The two five-year periods differ in meaning, and lapse is heavier in Guatemala.
Insight 3: the vigilance gap is the widest, and neither model can be transplanted. Article 56 of DE 490 sets a 72-hour deadline for serious adverse events, with the six criteria in Article 49 as the gate. Guatemalan standards set no device reporting deadline and place vigilance on the pharmaceutical side under NT 19-2009. Following the Guatemalan habit in Panama would commit a minor offence under Article 35.2 of Ley 90, while applying the Panamanian 72 hours in Guatemala would create an obligation that does not exist.
Insight 4: both countries are in transition, but convergence is not a six-country registration. PAN/135 and GTM/106 entered the WTO process in late 2025 and early 2026, showing that the main markets are rebuilding device rules simultaneously. Each text is notified, approved and brought into force separately, with no mutual recognition. Companies cannot defer an application in either market on that basis, nor assume that approval in one carries into the other.
4. Practical recommendations
- Manage the two classification outcomes separately. Give Panama (A to D) and Guatemala (I to IV) independent fields in the product master data and prohibit reuse across markets.
- Plan renewals against each credential cycle. Panama follows the five-year CCT and licence renewal 60 working days before expiry; Guatemala follows the five-year registration, which must be completed before expiry because a lapse means a full new application.
- Build the vigilance system to the stricter market. Use the Panamanian 72 hours as the baseline for classification and escalation procedures, and apply current Guatemalan requirements there while tracking the implementation of GTM/106.
- Do not suspend applications because a new generation of rules is coming. The proposed adoption date of PAN/135 is pending and GTM/106 is proposed to take effect on 2026-12-01; until an approved text appears, the current rules apply in both countries.
- Leave the other four countries to their own volumes. The authorities, classifications and credentials of Costa Rica, Nicaragua, Honduras and El Salvador must be verified in their own volumes rather than inferred here, which avoids contaminating the comparison.
5. Verification checklist
- □ Panama classification assigned as A, B, C or D under Article 30 of DE 490, with the higher-class rule recorded
- □ Guatemala classification assigned independently under Annex 1 of NT-37, with no reuse of Panamanian class numbers
- □ CCT and Inscripción Sanitaria managed separately, with names not interchanged
- □ Five-year CCT and licence renewal 60 working days before expiry scheduled for Panama
- □ Guatemalan renewal scheduled before the five-year expiry, with the re-application consequence flagged
- □ Vigilance procedure built on the Panamanian 72-hour baseline and not applied to Guatemala
- □ Remaining four countries marked as awaiting verification, with no inference made
- □ Status of PAN/135 and GTM/106 recorded, with current applications still running
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Panamanian market. Our team manages Ministry of Health (MINSA) sanitary registration applications, prepares Spanish-language dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us