Four Risk Levels I to IV
Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Nicaragua classifies medical devices into four risk classes, from lowest to highest I, II, III and IV. The semantic basis is definition item 33 of Chapter 4 of NTON 24001:2023, and the decision basis is the 26 classification rules of Annex B of the same standard, which is a normative annex (normativo). The risk class is determined by ANRS itself under Annex B (Chapter 5 Article 3), and it also determines whether the product falls within controlled-technology devices under definition item 10. Two textual issues must be noted: item 33 uses the word Clase while Annex B uses Nivel de Riesgo throughout; and Rule 8 d) of Annex B refers to a Nivel de Riesgo VI, a class that does not exist in the standard.
2. Legal basis
The table below lists the classification provisions of the standard and the pending amendment that touches Chapter 4.
| Technical standard | NTON 24001:2023, Tecnología de la salud. Dispositivos médicos. Requisitos para el Registro Sanitario (La Gaceta No. 192, 2024-10-21) | Chapter 4 definition item 33 | Nivel de riesgo: divided into Clase I, II, III and IV with increasing risk; I is very low risk, II low to moderate, III moderate to high, and IV the highest risk |
| Technical standard | NTON 24001:2023 | Annex B (normativo), Rules 1 to 26 | Classification decision rules; Rules 1 to 10 are general rules, Rule 11 covers software, Rule 19 covers nanomaterials, and Rules 20 to 25 cover in vitro diagnostics and RUO and IUO products |
| Technical standard | NTON 24001:2023 | Chapter 5 Article 3 | ANRS determines the risk class of the device to be registered under Annex B, independently of the classification registered in another country |
| Technical standard | NTON 24001:2023 | Chapter 4 definition item 10 | Controlled-technology devices include devices placed in Classes III and IV under this standard, prototypes, and second-hand or refurbished equipment, all subject to special controls |
| Technical standard | NTON 24001:2023 | Chapter 16 | This standard enters into force 90 days after publication in the official gazette La Gaceta |
| Draft amendment | Adenda Número 2026-05 (NTON 24001:2023 + Ad.01, adopted by the Comité Técnico de Reglamentación on 2026-05-25; WTO notification G/TBT/N/NIC/178/Add.1, distributed 2026-06-01) | Chapter 4 | The first visible change is a new definition of SaMD (Software as a Medical Device) with 7 notes; whether the classification rules are adjusted at the same time is a matter for the officially published final text |
3. Expert insights
Point 1: The terminology is dual-track within the standard itself. Item 33 reads Se clasifican en Clase I, II, III y IV, while the rules of Annex B consistently read Nivel de Riesgo I to IV. The two formulations coexist in the same standard and mean the same thing. This volume uses risk class I, II, III and IV as the uniform expression and keeps the original wording when quoting specific provisions, so that the text can be set against the official version line by line.
Point 2: Item 33 gives risk semantics, not decision rules. It describes roughly what degree of risk each class corresponds to: I very low, II low to moderate, III moderate to high, and IV the most critical in terms of risk. It provides no decision elements at all. A company cannot claim Class I by asserting that its product looks low-risk; classification must land on a specific rule number of Annex B.
Point 3: Rule 8 d) of Annex B contains the phrase se clasifican en Nivel de Riesgo VI, a class that does not exist in the standard. Subparagraphs a) to c) and e) to j) of the same rule all take values between II and IV, and item 33 also lists only four classes. This volume reads that subparagraph as Class IV, and recommends written confirmation of the class with the DDM for implantable or long-term invasive surgical devices involving chemical change in the body. This is one of the numbering errors found in the official text during the preparation of this volume.
Point 4: Rule 5 of Annex B contains two subparagraphs lettered b, and the provisions that follow never receive the letter c. This is the second numbering problem in the same annex. Companies citing Rule 5 for classification must cite the specific subparagraph, to avoid ambiguity in later renewals or changes.
Point 5: Annex B Rule 11 is the only classification landing point for software in Nicaragua, and it is three-tier. Software used for therapeutic or diagnostic decisions is Class II; where the decision could cause death or irreversible harm to health it is Class IV, and where it could cause serious harm to health or require surgical intervention it is Class III. Software used to observe physiological processes is Class II, and Class III where the vital parameters observed could, if they varied, immediately endanger the patient. All remaining software is Class I. This rule must be read together with the SaMD definition that Adenda 2026-05 proposes to add.
Point 6: The class does more than determine the document list. Item 10 lists devices placed in Classes III and IV under the standard as one of the constituent elements of controlled-technology devices. The class therefore triggers special controls as well, and it is the starting point of the chain of obligations, not its end.
4. Practical recommendations
- Classify along the path from rule to class: compare the product against the 26 rules of Annex B one by one, form a written conclusion and note the rule numbers relied on, rather than fixing the class first and looking for a rule afterwards.
- For implantable or long-term invasive surgical devices falling under Rule 8 d) of Annex B, meaning chemical change in the body, prepare the dossier at Class IV and confirm the class with the DDM in writing.
- Classify software products through the three tiers of Rule 11: first separate decision-type from observation-type software, then assess the tier of the harm consequence.
- Do not use the classification registered in another country as the basis for classification in Nicaragua (Chapter 5 Article 3); FDA or CE classifications may be kept in internal files as background only.
- For borderline products, where it is uncertain whether the item is a device or needs registration, apply for an official classification ruling under Annex D, with the DDM issuing the conclusion under Annex B.
- Archive the classification conclusion together with the rule numbers relied on, for reuse at renewal, post-registration changes and family grouping expansions, so that the same product does not end up with two classes at different stages.
5. Verification checklist
- □ The product has been compared rule by rule against Annex B, with a written classification conclusion noting the rule numbers relied on
- □ The classification conclusion does not cite any foreign classification as its basis (Chapter 5 Article 3)
- □ It has been checked whether the product meets the constituent elements of controlled-technology devices in item 10
- □ Software products have been classified through the three tiers of Rule 11 of Annex B
- □ Products under Rule 8 d) have been prepared at Class IV and confirmed with the DDM in writing
- □ The consistency between the classification conclusion and the Annex D classification certificate has been cross-checked
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Nicaraguan market. Our team manages Ministry of Health (MINSA) sanitary registration applications through the regulatory directorate, prepares Spanish-language dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us