Four Risk Classes: I, IIA, IIB and III

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Honduras classifies medical devices into 4 risk classes (Article 29 of Acuerdo No. 0631-ARSA-2023, the Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario): Clase I, low risk, subject to general controls, not used to protect or sustain life and not of special importance in the prevention of damage to health, and posing no potential risk of illness or injury; Clase IIA, moderate risk, requiring special controls at the production stage to demonstrate effectiveness; Clase IIB, high risk, requiring special controls at the design and production stages to demonstrate safety and effectiveness; Clase III, very high risk, requiring special controls, used to protect or sustain life or of substantial importance in the prevention of damage to health, or whose use poses a potential risk of illness or injury. Classification takes the intended use (finalidad prevista) declared by the product holder as the pivot, together with the anatomical site of application, the duration of contact with the body, the degree of invasiveness, and whether the device releases medicines or energy. Article 30 empowers ARSA to compile and update device lists.

  • Intended use is the only subjective input; the other four criteria are objective determinants, so the same hardware can fall into different classes depending on the declared intended use.
  • Dossier requirements accumulate in sequence: a Clase III file equals the full Clase I set plus the IIA, IIB and III additions.
  • The classification report is valid for 6 months and can be extended once, for a further 6 months, under Article 31.

2. Legal basis

The provisions on classification and its documentary consequences are the following.

RegulationAcuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés SanitarioArticle 29Four classes: Clase I low risk; Clase IIA moderate risk; Clase IIB high risk; Clase III very high risk.
RegulationAcuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés SanitarioArticle 29, criteriaIntended use declared by the product holder as the pivot, with reference to anatomical site, duration of contact, degree of invasiveness, and release of medicines or energy.
RegulationAcuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés SanitarioArticle 30ARSA compiles and updates device lists, enabling dynamic regulation.
RegulationAcuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés SanitarioArticle 31The classification report precedes registration; valid 6 months, extendable once by 6 months.
RegulationAcuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés SanitarioArticle 4316 items for the Clase I dossier (the base for the other classes).
RegulationAcuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés SanitarioArticle 44Clase IIA adds ISO 13485; IVD adds a verification report; software adds 7 items.
RegulationAcuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés SanitarioArticle 45Clase IIB adds device-specific ISO standards (13485, 10993-1, 14155 NTC, 14155, 6872, among others).
RegulationAcuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés SanitarioArticle 46Clase III adds a summary of manufacture for devices of human origin, four tests, the certificate of analysis of the finished product, and 5 categories of equivalent documents.

3. Expert insights

Insight 1: classification accumulates in sequence rather than running in parallel tracks. Article 44 requires Clase IIA applicants to submit the documents of Article 43 plus ISO 13485; Article 45 requires Clase IIB applicants to submit those of Articles 43 and 44 plus the specific standards; Article 46 requires Clase III applicants to submit those of Articles 43, 44 and 45 plus the four tests and equivalent documents. In other words, a Clase III dossier equals the full Clase I set plus the IIA additions, the IIB additions and the III additions. Budgets and timelines must be estimated by cumulating up to the highest class, not by the increment between adjacent classes.

Insight 2: the definitions of Clase I and Clase III are mirror images. The Clase I definition contains three negative elements: not used to protect or sustain life, not of special importance in the prevention of damage to health, and posing no potential risk of illness or injury. The Clase III definition restates nearly the same elements affirmatively. This drafting means that borderline products turn on the wording of the intended use: the same monitor declared for life support and declared for general monitoring will land at opposite ends.

Insight 3: the list of Article 30 is the operational carrier of classification. The provision empowers ARSA to compile and update device lists, while Article 28 offers a classification certificate route for products not on the list. The practical advice for applicants is to search the list first and apply for a classification certificate only if the product is not found; do not skip the list and file a self-assessed class, because Article 31 expressly requires the classification report from ARSA first, and a self-assessment has no procedural effect.

Insight 4: the classification outcome feeds back into establishment duties. Article 9 lists Clase I devices as one of the three exemptions from the technical manager requirement, while Clase IIA and above add ISO and testing documents. A single classification outcome therefore determines staffing, document volume and cost at once. The classification report (CDM) should thus be the first milestone of the project schedule, not an annex to the registration documents.

Insight 5: the classification report has a clock, and higher classes run out of time more easily. Article 31 gives the classification report 6 months of validity with a single 6-month extension, while Clase III requires four tests and a certificate of analysis of the finished product, whose preparation often exceeds 6 months. The GRZAN regulatory team advises that Clase III projects complete testing before the classification report is issued; otherwise the project depends on the extension or even a new classification application.

4. Practical recommendations

  • Estimate budget and timeline on a cumulative-to-highest-class basis: Clase III equals the full Clase I set plus the IIA, IIB and III additions.
  • Have regulatory and marketing departments co-sign the intended use wording of borderline products, avoiding two classifications for one piece of hardware.
  • Search the ARSA list first and apply for a classification certificate under Article 28 only where the product is not found; never skip the list in favour of self-assessment.
  • Set the classification report (CDM) as the first project milestone, starting before document preparation.
  • For Clase III projects, schedule biocompatibility, stability, cytotoxicity and electrical safety testing before the classification report is issued.
  • After the classification conclusion is issued, recheck technical manager staffing and the document list to catch the additions.

5. Verification checklist

  • □ The risk class of the product has been confirmed in a formal ARSA classification report
  • □ Budget and timeline have been estimated on the cumulative-to-highest-class basis
  • □ The intended use wording of borderline products has been co-signed by regulatory and marketing
  • □ The ARSA list was searched first, and products not found have a pending or obtained Article 28 certificate
  • □ The four tests of Clase III projects are scheduled before issue of the classification report
  • □ Staffing and the document list were rechecked once the classification conclusion was issued

Regulatory information verified as of 2026-10-29.

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Honduran market. Our team manages Secretariat of Health (SESAL) registration applications through the regulatory directorate, prepares Spanish-language dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail