Tecnovigilancia under Acuerdo No. 0185-ARSA-2025
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Honduran tecnovigilancia has moved from principle to clock, and the marker is Acuerdo No. 0185-ARSA-2025, the Reglamento de Tecnovigilancia (La Gaceta No. 36,954, published 2025-09-29, containing 29 articles, with Article 29 providing that it enters into force 20 days after publication). It sets four reporting clocks for companies: serious adverse events must be reported as soon as possible and no later than 3 calendar days (72 hours) from becoming aware (Article 18); non-serious events with a rising incidence rate, after analysis and verification by the liaison officer, within 30 calendar days (Article 19); discovery of substandard or counterfeit devices within 5 calendar days (Article 25); and the semiannual report submitted in the first 10 business days of January and July (Article 20). As for institutional structure, Article 4 establishes the National Tecnovigilancia System (SNTV) and the National Tecnovigilancia Centre (CNTV), and Article 6 places the CNTV under the medical devices directorate (DDM) of ARSA.
2. Legal basis
The table below consolidates the reporting clocks, the event classification and the institutional provisions of Acuerdo No. 0185-ARSA-2025.
| Level | Instrument | Provision | Key point |
|---|---|---|---|
| Regulation | Acuerdo No. 0185-ARSA-2025 (La Gaceta No. 36,954, 2025-09-29) | Article 18 | Post-market serious adverse events: as soon as possible and no later than 3 calendar days (72 hours) |
| Regulation | Acuerdo No. 0185-ARSA-2025 | Article 19 | Non-serious events with rising incidence: 30 calendar days after analysis and verification by the liaison officer |
| Regulation | Acuerdo No. 0185-ARSA-2025 | Article 25 | Discovery of substandard or counterfeit devices: 5 calendar days |
| Regulation | Acuerdo No. 0185-ARSA-2025 | Article 20 | Semiannual report: first 10 business days of January and July; a zero report is required even when no events occur |
| Regulation | Acuerdo No. 0185-ARSA-2025 | Article 13 | Four-level classification: serious and non-serious adverse events, and incidents associated with each |
| Regulation | Acuerdo No. 0185-ARSA-2025 | Article 14 | Eight criteria for serious adverse events (including IVD misdiagnosis and defects in labelling, IFU and promotional material) |
| Regulation | Acuerdo No. 0185-ARSA-2025 | Articles 4 and 6 | SNTV and CNTV; the CNTV sits under the medical devices directorate (DDM) of ARSA |
| Regulation | Acuerdo No. 0185-ARSA-2025 | Article 29 | Entry into force 20 days after publication |
3. Expert insights
Insight 1: the four clocks are independent and must not be read as one. The 72-hour clock of Article 18 applies to serious events and starts from becoming aware (desde que se ha tenido conocimiento); the 30-day clock of Article 19 applies to non-serious events and starts after analysis and verification by the liaison officer; the 5-day clock of Article 25 applies to the discovery of substandard or counterfeit devices; and Article 20 is a periodic 10-business-day window. Their starting points, subject matter and submission methods all differ, so companies must run the clocks separately according to the nature of the event.
Insight 2: two of the eight criteria for serious events point directly at documentation and diagnosis. Article 14, item f) includes among serious events indirect harm caused by an erroneous diagnosis or test result from an IVD used according to the manufacturer’s instructions; item h) includes serious deterioration of a patient’s health caused by omissions or defects in the labelling, instructions for use and promotional material. In other words, the correctness of technical documentation is itself an object of vigilance, and a labelling error by an IVD company can directly trigger a 72-hour report.
Insight 3: reporting is not limited to companies. Article 18 makes clear that users or patients, health professionals, and tecnovigilancia liaison officers of the device industry at home and abroad may all report to the CNTV. This means a company’s 72-hour clock may be started externally before the company itself is even aware; an internal mechanism for recovering information from distributors and hospitals is therefore indispensable.
Insight 4: vigilance obligations cover every actor in the chain. Article 8 brings within scope the registration holder, the device product owner, manufacturers, refurbishers, repackagers, importers, exporters, warehousing parties and distributors, as well as entities engaged in rental, lending, transport, sale, research and maintenance, requiring them to establish a vigilance system covering the entire life cycle of the device and to keep records traceable. Article 9 separately imposes four obligations on healthcare providers and independent health professionals.
Insight 5: events during clinical trials are received centrally by the CNTV. Article 21 requires investigational devices to comply with the national clinical research rules; Article 22 assigns the reception, identification, registration, analysis, evaluation and follow-up of adverse events during trials to the CNTV. Together with Article 86 of 0631-ARSA-2023 (ethical evaluation by CEIB of UNAH, approval by ARSA and issuance of the trial-site permit), this creates two parallel regulatory lines during the trial phase.
4. Practical recommendations
- Build an internal decision tree around the four-level classification of Article 13: classify first, start the clock second, so that a serious event is never processed as a non-serious one.
- Engineer the 72 hours as an end-to-end process: awareness at the distributor or hospital end, verification by the liaison officer, use of the ARSA form (Article 12, item 3), submission to the CNTV — with no manual hand-offs in between.
- IVD companies should watch labelling and instructions especially closely: under Article 14, items f) and h), misdiagnosis and documentation defects can both constitute serious events.
- Time substandard and counterfeit discoveries separately: the 5 calendar days of Article 25 are tighter than 30, and the report must contain at least the four items of Article 26 — product name, place of discovery, batch number and product owner or manufacturer, and the grounds for the judgement.
- Bring the external reporting route of Article 18 into the monitoring scope: agree information return deadlines with local distributors and healthcare institutions so that the company’s awareness date is never later than that of the CNTV.
- For clinical trial projects, interface with both the CNTV and CEIB: trial-phase events go to the CNTV under Article 22, while protocols pass ethical evaluation and ARSA approval under Article 86.
5. Verification checklist
- □ The decision tree for the four-level classification of Article 13 has been written into the vigilance SOP
- □ The 72-hour reporting path is open end to end (awareness, verification, ARSA form, submission)
- □ A responsible person has been designated for the analysis and verification step of 30-day non-serious event reports
- □ The four-content template for 5-day substandard or counterfeit reports is ready (Article 26)
- □ The January and July windows (first 10 business days) for the semiannual report are in the calendar
- □ Dual contacts for the CNTV and CEIB have been named for clinical trial projects
- □ The vigilance record system meets the requirements of integrity, accuracy, reliability, confidentiality and traceability (Article 12, item 2)
Related reading: [Article 46 — Semiannual report and vigilance liaison](/honduras-medical-device-registration/semiannual-report-and-vigilance-liaison/) | [Article 49 — Regulatory timeline 2020–2026 and the 180-day window](/honduras-medical-device-registration/honduras-regulatory-timeline-and-180-days/) | [Article 44 — Five cancellation grounds and eight sale prohibitions](/honduras-medical-device-registration/cancellation-grounds-article-88/)
Regulatory information verified as of October 29, 2026.
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Honduran market. Our team manages Secretariat of Health (SESAL) registration applications, prepares Spanish-language dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us