Reclassification and Referral to DNPF
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
The end point of a modification application is not necessarily an approved modification. Article 48 of Acuerdo No. 0631-ARSA-2023 gives ARSA two powers. When evaluating a modification, ARSA may reclassify the device (reclasificación) according to its level of sanitary risk — the trigger being that three elements, namely specifications, intended purpose or chemical composition, have undergone major changes that may affect sanitary safety — and the consequence may be a corresponding change to the sanitary registration. If the technical evaluation further concludes that the intended purpose of the device now falls within the pharmaceutical sphere, the file is referred to the Dirección Nacional de Productos Farmacéuticos y Otros del Ramo de Interés Sanitario (DNPF); the applicant must then proceed under the pharmaceutical regulation, and the referral must be notified to the applicant. The practical lesson is that Article 48 offers two exits that both hang on the single entrance of modification review, while modifications themselves are constrained by the twelve-category list of Article 52: once a product undergoes a substantive change, what lies ahead is not merely document preparation but potentially a change of regulatory pathway.
2. Legal basis
The table below consolidates the provisions governing reclassification during modification review and the referral mechanism to the pharmaceutical directorate, together with the risk-classification framework that both mechanisms draw upon.
| Level | Instrument | Provision | Key point |
|---|---|---|---|
| Regulation | Acuerdo No. 0631-ARSA-2023 | Article 48 | ARSA may reclassify according to sanitary risk level; triggered by major changes in specifications, intended purpose or chemical composition that may affect sanitary safety; may lead to a change of the registration |
| Regulation | Acuerdo No. 0631-ARSA-2023 | Article 48, second paragraph | Where technical evaluation finds the intended purpose falls within the pharmaceutical sphere, the file is referred to the DNPF, handled under the pharmaceutical regulation, and the applicant is notified |
| Regulation | Acuerdo No. 0631-ARSA-2023 | Article 29 | Four risk classes I / IIA / IIB / III, with intended purpose as the primary axis of determination |
| Regulation | Acuerdo No. 0631-ARSA-2023 | Article 25 | Definition of a device: the principal action is not achieved by pharmacological, immunological or metabolic means |
| Regulation | Acuerdo No. 0631-ARSA-2023 | Article 30 | ARSA compiles and updates the classification list |
| Regulation | Acuerdo No. 0631-ARSA-2023 | Article 31 | The classification report is a prerequisite for registration, valid for six months and extendable once |
| Regulation | Acuerdo No. 0631-ARSA-2023 | Article 52, item 3, e) and h) | The application channels for changes of specifications and of chemical composition |
| Regulation | Acuerdo No. 0631-ARSA-2023 | Article 87, item 7 | No authorisation is granted to products that scientific analysis finds to be substandard or counterfeit |
3. Expert insights
Insight 1: reclassification is not a standalone procedure but the outcome of modification review. Article 48 opens with the words al evaluar las modificaciones — when evaluating modifications. In other words, a company will never receive a reclassification application form; instead, after submitting a modification under Article 52 (especially e) specifications or h) chemical composition), ARSA determines of its own motion whether the risk class must move upward.
Insight 2: the axis of determination matches Article 29 exactly. Article 29 establishes intended purpose as the primary axis, with reference to anatomical site, duration of contact, degree of invasiveness and whether the device releases medicines or energy. The trigger elements of Article 48 are precisely specifications, finalidad (intended purpose) and chemical composition. The two provisions are the same logic applied at initial classification and at subsequent adjustment.
Insight 3: referral to the DNPF is a change of route, not a rejection. The wording of the provision is se remitirá a la DNPF, and the applicant must then satisfy the requirements of the pharmaceutical regulation. For drug-device combinations, drug-coated products or products whose pharmacological action is enhanced by a reformulation, this means switching from a device dossier to a pharmaceutical dossier, with entirely different timelines and materials — project budgets must be prepared along both paths separately.
Insight 4: the referral must be notified, which is a procedural safeguard. The final sentence of Article 48, Dicha remisión será notificada al peticionario, makes notification an obligation. Only after receiving the referral notice does the company become obliged to supplement its file under the pharmaceutical regulation; before that, it need not switch pathways on its own initiative.
Insight 5: the referral mechanism and the device definition of Article 25 mark the same boundary. Article 25 requires that the principal intended action of a device is not achieved by pharmacological, immunological or metabolic means (medicinal, immunological or metabolic action being only ancillary). Once a modified product crosses this line it no longer belongs to the device category, and the referral mechanism of Article 48 is the enforcement device for that boundary.
4. Practical recommendations
- Run a boundary self-assessment before modifying: determine whether the change touches any of the three elements — specifications, intended purpose or chemical composition — and whether it may affect sanitary safety (Article 48).
- Where medicinal ingredients, drug coatings or adjustments to the mechanism of action are involved, prepare contingency plans and budgets for the pharmaceutical path in advance, rather than scheduling on the device path alone.
- When changing specifications (item e)) or composition (item h)), review each of the four reference factors of Article 29 — anatomical site, duration of contact, degree of invasiveness, and release of medicines or energy — and anticipate whether the class will move upward.
- After an upward reclassification, prepare the additional documents for the new class in parallel: ISO 13485 for Class IIA (Article 44), device-specific ISO certification for Class IIB (Article 45), and the four test items and finished-product analysis certificate for Class III (Article 46).
- Upon receiving the DNPF referral notice, switch immediately to the document list of the pharmaceutical regulation and treat the date of the notice as the starting point of the new pathway.
- Retain the initial classification report (Article 31) and all modification approval documents as traceable evidence of how the risk class has evolved.
5. Verification checklist
- □ The intended change has been self-assessed against the three Article 48 elements (specifications, intended purpose, composition)
- □ The four reference factors of Article 29 have been reviewed one by one and the risk of an upward reclassification evaluated
- □ The list of additional documents after an upward reclassification (Articles 44, 45 and 46) has been prepared
- □ Contingency plans for the pharmaceutical path exist for drug-containing or pharmacologically enhanced cases
- □ A responsible person has been designated for receiving and responding to DNPF referral notices
- □ The initial classification report and all modification approvals have been archived
Related reading: [Article 43 — Eight refusal grounds](/honduras-medical-device-registration/refusal-grounds-article-87/) | [Article 41 — Twelve categories of post-registration changes](/honduras-medical-device-registration/post-registration-changes-twelve-types/) | [Article 09 — Four risk classes and the Article 29 axis](/honduras-medical-device-registration/four-risk-classes-i-iia-iib-iii/)
Regulatory information verified as of October 29, 2026.
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Honduran market. Our team manages Secretariat of Health (SESAL) registration applications, prepares Spanish-language dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us