Class I Dossier: The Sixteen Registration Requirements
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Acuerdo No. 0631-ARSA-2023, the Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario, lists in Article 43 sixteen dossier items for Class I device registration: 1 the application form (with 18 groups of information); 2 the new-registration declaration form; 3 the free sale certificate (CLV); 4 the manufacturer’s good manufacturing practice certificate (BPM); 5 the quality control specification applicable to the device; 6 the safety report (which must include existing sanitary vigilance records); 7 the vigilance corrective action report validated by the product holder (where applicable); 8 the risk management report; 9 the manufacturing or contract manufacturing agreement when the registering company claims to be the product holder; 10 a copy of the company deed or sole trader registration; 11 the primary, secondary and/or tertiary packaging labels; 12 the complementary label; 13 the ISO conformity supporting document when the label bears the word sterile or a similar wording or symbol; 14 a copy of the legal representative’s professional licence; 15 the power of attorney granted to the authorised attorney (notarised deed or letter of attorney with certification); 16 the proof of payment of fees. In addition, biomedical equipment, IVDs and medical disinfectant and antiseptic products must add six, six and seven items respectively. The GRZAN regulatory team notes: Article 43 is the common base for the other three classes — Articles 44, 45 and 46 all start from the items listed in Article 43 and add to them.
2. Legal basis
The table below maps each dossier item of Article 43 to its provision and key point.
| Level | Instrument | Provision | Key point |
|---|---|---|---|
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 43, item 1 | Application form with 18 groups of information: receiving authority, legal representative, agent, six items each on the registering and distributing companies, risk classification report number, trade name, generic name, product holder, manufacturer, international intermediary distributor, description and indications, commercial presentations, reference code, model or series, international coding, CAS number, place and date, signature and fingerprint |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 43, item 3 | Free sale certificate (Certificado de libre venta) |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 43, item 4 | Manufacturer’s good manufacturing practice certificate (BPM) |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 43, items 5 to 8 | Quality control specification, safety report (including existing vigilance), corrective action report validated by the product holder, risk management report |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 43, item 9 | Manufacturing or contract manufacturing agreement when the registering company claims to be the product holder |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 43, items 12 and 13 | Complementary label; ISO conformity document when the label bears the word sterile or a similar symbol |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 43, item 15 | Power of attorney for the authorised attorney (notarised deed or letter of attorney with its certification) |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 43, Section 1 | Biomedical equipment adds six items: technical nameplate, service manual, user manual, equipment images, descriptive information, and manufacturer catalogue information |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 43, Section 2 | IVDs add six items: technical datasheet, insert, technical document supporting shelf life or stability, storage and transport conditions, safety data sheet, qualitative and quantitative formulation |
| Regulation | Acuerdo No. 0631-ARSA-2023, Reglamento para el Control Sanitario de Dispositivos Médicos y otros Dispositivos de Interés Sanitario | Article 43, Section 3 | Medical disinfectant and antiseptic products add seven items: insert, stability document, storage and transport conditions, qualitative and quantitative formulation, sterilisation certificate, container-closure technical document, product effectiveness validation report |
3. Expert insights
Insight 1: four of the sixteen items are report-type documents and are easily mistaken for formatting material. Item 6, the safety report (which must include existing sanitary vigilance records), item 7, the corrective action report validated by the product holder, and item 8, the risk management report, together with item 5, the quality control specification, are the written implementation of the five basic principles of Article 6. The GRZAN regulatory team recommends that the safety report proactively search for and include the device’s international vigilance records; omitting them is treated as insufficient evidence under Article 87, items 1 and 2.
Insight 2: item 9 turns the question of who may claim to be the product holder into a documentary question. The provision states that when the registering company claims to be the product holder, a manufacturing or contract manufacturing agreement must be submitted. This means that a contract manufacturing company must obtain contractual authorisation from the manufacturer in advance; otherwise it can only file as the registering company, and the product holder named in the dossier should still be the actual product holder. This point is frequently overlooked, leaving the application form inconsistent with the agreement.
Insight 3: item 13 is the evidentiary switch for sterile claims. The provision requires that when the label bears the word estéril or any other similar characteristic or symbol, a document supporting conformity with ISO specifications must be submitted. In other words, whether this document is required depends on what the label says. When localising labels, bind the sterile wording to its supporting document so that a label change does not leave the document behind.
Insight 4: the 18 groups of information in the application form must cover four parties at once. Consistent with Article 32, Article 43, item 1, letters d) to k), requires the registering company, the domestic distributing company, the product holder, the manufacturer and — where the purchase is not direct — the international intermediary distributor. The GRZAN regulatory team notes: the four-party information in the risk classification report and the registration application must match word for word, especially addresses and email addresses; inconsistency is the most common source of non-conformity findings.
Insight 5: the three supplementary lists are triggered by product type, not by risk class. The biomedical equipment, IVD and disinfectant and antiseptic supplements in Article 43, Sections 1 to 3, apply to all classes (Articles 44, 45 and 46 all require Article 43 to be satisfied first). A Class III IVD must therefore submit both the four Class III tests and the six IVD supplements of Article 43, Section 2.
4. Practical recommendations
- Write the three reports of items 6, 7 and 8 point by point against the five principles of Article 6, not as generic templates.
- Proactively search for and include the device’s existing international sanitary vigilance records in the safety report.
- Contract manufacturers should obtain the manufacturing or contract manufacturing agreement in advance and confirm the contracting parties match the application form.
- Where the label bears the word sterile or a similar symbol, prepare the ISO conformity supporting document at the same time.
- Compare the four-party information in the risk classification report and the registration application word for word, especially addresses and emails.
- Check the supplementary lists for biomedical equipment, IVDs and disinfectant and antiseptic products separately by product type, not by class.
5. Verification checklist
- □ The sixteen dossier items have been counted one by one against Article 43
- □ The safety report includes existing sanitary vigilance records found by proactive search
- □ Contract manufacturers hold a manufacturing or contract manufacturing agreement with matching parties
- □ ISO conformity documents are ready for labels bearing sterile claims
- □ Four-party information in the classification report and the application has been compared word for word
- □ Product-type supplementary lists have been checked separately by type
Related reading: [Article 22 — What Class IIA adds](/honduras-medical-device-registration/class-iia-iso-13485-and-software/) | [Article 26 — Primary and complementary labels](/honduras-medical-device-registration/labelling-primary-and-complementary/) | [Article 28 — Additional IVD materials](/honduras-medical-device-registration/ivd-specific-requirements/)
Regulatory information verified as of October 29, 2026.
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Honduran market. Our team manages Secretariat of Health (SESAL) registration applications, prepares Spanish-language dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us