Professional Responsible and Authorised Representative: Nine Duties Under Article 6.2.4.2

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Article 6.2.4.2 of RTS 11.03.02:21 assigns 9 duties (a to i) to the professional responsible for medical devices (Profesional responsable de dispositivos médicos): a) handling the sanitary registration of devices intended for marketing, together with its changes, renewal and annual fee; b) handling import authorisations for registered devices; c) ensuring that products sold conform to the registration; d) responding to enquiries from the authority; e) coordinating with the vigilance liaison on vigilance notifications and market withdrawals; f) submitting a resignation to the authority when leaving the appointment; g) reporting any absence exceeding 15 days in writing, naming the substitute and the period, signed and sealed by two professionals and submitted at least 5 days in advance; h) ensuring the quality, safety and efficacy of devices at the applicable regulatory stage; and i) where the role is combined with that of vigilance liaison, meeting the qualifications and functions set by that regulation.

The list is short but unusually dense, because it gathers the entire product lifecycle into a single named individual. Registration, changes, renewal, the annual fee, import authorisation, conformity of what is actually sold, answers to the authority and vigilance coordination all sit with the same person. In practice this makes the professional responsible the natural counterpart to the authority for everything that concerns a product, as distinct from everything that concerns a building.

2. Legal basis

The nine duties, the per-registration binding rule and the fee for changing the appointment are set out in the following provisions.

LevelInstrumentReferencePublishedProvisionKey point
Technical regulationRTS 11.03.02:21Acuerdo No. 14-20222022-08-23Article 6.2.4.2 a)Handling registration, changes, renewal and the annual fee
Technical regulationRTS 11.03.02:21Acuerdo No. 14-20222022-08-23Article 6.2.4.2 b)Handling import authorisations for registered devices
Technical regulationRTS 11.03.02:21Acuerdo No. 14-20222022-08-23Article 6.2.4.2 c) and d)Ensure that products sold conform to the registration; answer enquiries from the authority
Technical regulationRTS 11.03.02:21Acuerdo No. 14-20222022-08-23Article 6.2.4.2 e)Coordinate vigilance notifications and market withdrawals with the vigilance liaison
Technical regulationRTS 11.03.02:21Acuerdo No. 14-20222022-08-23Article 6.2.4.2 f) and g)Resignation submitted to the authority; absence over 15 days reported in writing 5 days in advance
Technical regulationRTS 11.03.02:21Acuerdo No. 14-20222022-08-23Article 6.2.1 e)Each registration must be bound to a professional responsible and an authorised representative; replacement is handled registration by registration
Fee scheduleTarifario de la SRS (2025 edition)SRS Council resolution2025-10Section 3.2.2Change of professional responsible or agent: basic fee USD 48.00 plus USD 0.60 per affected registration

3. Expert insights

Insight 1: The professional responsible is the product-side role, while the technical director is the establishment-side role. Article 6.2.4.2 a) assigns registration, changes, renewal and the annual fee to the professional responsible, and b) assigns import authorisation to the same person, whereas Article 6.2.4.1 gives the technical director the establishment activities and the inspection interface. The two sets of duties do not overlap and cannot be substituted one for the other. A company must therefore either keep two licensed professionals available or, where one individual holds both posts, be able to demonstrate that both sets of duties are actually performed, because an inspection will test the duties and not the organisational chart.

Insight 2: The professional responsible is bound to each individual registration, so replacement must be handled registration by registration. Article 6.2.1 e) requires the registration holder to appoint a professional responsible and an authorised representative for each registration, and provides that resignation or withdrawal of the appointment must be processed for each sanitary registration with notice to the authority. The arithmetic consequence is direct: replacing one professional responsible across a portfolio of 20 registrations means 20 separate change filings. Section 3.2.2 of the tariff reflects exactly this structure, charging a basic fee of USD 48.00 plus USD 0.60 for each affected registration.

Insight 3: Even a temporary absence requires prior written approval. The final paragraph of Article 6.2.1 e) provides that a temporary absence of the professional responsible requires prior written application to the authority for approval, together with the designation of a substitute supported by the corresponding authorisation documents. Article 6.2.4.2 g) then elaborates the case of absence exceeding 15 days into a written report submitted 5 days in advance and signed and sealed by two professionals. The combined effect is that the professional responsible cannot leave first and report afterwards; the approval must precede the absence, which means cover has to be arranged before travel is booked.

Insight 4: Item c) on conformity, item h) on continuing assurance and item e) on vigilance coordination place the professional responsible at the front line of post-market action. Item c) requires that products sold in the country match the content of the registration under that person’s responsibility, item h) requires quality, safety and efficacy to be assured at the applicable regulatory stage, and item e) requires coordination of vigilance notifications and market withdrawals. Taken together, they mean that when a vigilance event occurs, the authority’s first enquiry lands on the professional responsible rather than on the establishment, and the response will be measured against the registration content the person is accountable for.

4. Practical recommendations

  • Where a company holds many registrations, calculate the number of change filings and the total fee before replacing the professional responsible, using USD 48.00 plus USD 0.60 per registration.
  • Draw up a comparison table of the professional responsible and technical director duties so that neither role can deflect a task to the other.
  • Prepare a written absence report template, file it 5 days in advance and obtain the signatures and seals of two professionals.
  • Run a quarterly self-check that registered products match products actually sold, to prevent specification or manufacturer drift.
  • Designate the professional responsible as the first contact for vigilance events and set an internal response deadline in advance.
  • Maintain the authorised representative (Apoderado) alongside the professional responsible, since changes to either appointment must be filed.

5. Verification checklist

  • □ Duty boundaries between professional responsible and technical director tabulated
  • □ Number of change filings and total fee calculated before any replacement
  • □ Written absence report template prepared
  • □ Quarterly conformity self-check between registered and sold products scheduled
  • □ First contact for vigilance events designated
  • □ Authorised representative registration status maintained in parallel

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Salvadoran market. Our team manages Ministry of Health (MINSAL) sanitary registration applications, prepares Spanish-language dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail