Four Risk Classes and the Highest Risk Classification Rule
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Costa Rica classifies EMB by risk into four classes, Clase 1 to Clase 4, with Clase 1 the lowest risk and Clase 4 the highest, determined under the classification rules of Annex A. Article 8.1 establishes the highest-risk rule: “Si un EMB puede ser clasificado en más de una categoría, se clasificará en la clase de mayor riesgo” — where a product can be placed in more than one class, it is classified in the class of higher risk. Article 8.2 adds a combination rule for IVDs: where analysers, reagents and software are used together with another IVD, both are classified in the higher risk class.
- Four risk classes, Clase 1 to 4, determined under the Annex A rules (Article 8.1).
- Where more than one class applies, the higher risk class governs (Article 8.1).
- IVD combinations, including analysers, reagents and software, are both classified in the higher class (Article 8.2).
- Products not covered by the rules are classified by administrative resolution of the Ministry of Health (Article 8.3).
- A change that alters the risk classification is treated as a new product requiring re-registration (Articles 10.5 and 10.9).
2. Legal basis
The provisions below set out the four-class system, the Annex A rule structure, and the articles that attach registration consequences to the classification outcome.
| Level | Instrument | Provision | Key point |
| Technical regulation | RTCR 505:2022 | Article 8.1 | Four-class system (Clase 1–4) under the Annex A rules; where several classes apply, the higher risk governs |
| Technical regulation | RTCR 505:2022 | Article 8.2 | IVD combinations (including analysers, reagents and software) are both classified in the higher class |
| Technical regulation | RTCR 505:2022 | Article 8.3 | Products not covered by the rules are classified by administrative resolution of the Ministry of Health, which must be incorporated into the regulation |
| Technical regulation | RTCR 505:2022 | Annex A.1 | General EMB rules 1–16 (invasive, non-invasive, active and special groups) |
| Technical regulation | RTCR 505:2022 | Annex A.2 | IVD rules 1–9 (transmissible-agent uses, other uses, and special rules) |
| Technical regulation | RTCR 505:2022 | Article 9 | The corresponding sanitary registration must be obtained before marketing |
| Technical regulation | RTCR 505:2022 | Article 10.5 | A change that alters the risk classification is treated as a new product requiring re-registration |
| Technical regulation | RTCR 505:2022 | Articles 11 / 12 / 13 | Clase 2 / 3 / 4 correspond to three escalating tiers of dossier requirements |
3. Expert insights
Insight one: the rule structure of Annex A is group plus number, with general devices and IVDs forming two separate sets. General EMB (A.1) are divided into four groups totalling 16 rules: A.1.1 invasive devices (rules 1–3), A.1.2 non-invasive devices (rules 4–7), A.1.3 active devices (rules 8–12), and A.1.4 special rules (rules 13–16). IVDs (A.2) are divided into three groups totalling 9 rules: A.2.1 uses related to transmissible agents (rules 1–3), A.2.2 other uses (rules 4–5), and A.2.3 special rules (rules 6–9). The order of determination is: first establish whether the product is a general EMB or an IVD, then apply the rules within the group one by one, reaching Clase 1 only through the residual rules (rules 7, 12 and 9).
Insight two: the highest-risk rule and the IVD combination rule stack, making it easy for IVD systems to jump classes. Article 8.1 handles the case of a single product matching several rules; Article 8.2 handles the case of several components of an IVD system used together — both are classified in the higher risk class. By way of example: an analyser classified as Clase 2 under the rules, with a companion reagent classified as Clase 3, places both at Clase 3. The consequence is that dossier requirements jump from Article 11 (Clase 2) to Article 12 (Clase 3, adding six clinical evidence items). When filing IVD systems, companies should determine the class for the system as a whole first, then configure the dossier, rather than filing component by component.
Insight three: classification is not final; the regulation provides two later adjustment channels. Article 8.3 addresses new products not covered by the rules: the Ministry of Health determines their class by administrative resolution, which must be incorporated into the regulation. Article 22.2 addresses products already marketed as Clase 1: where the Ministry of Health has reasonable grounds, it may require a redetermination of the true classification (see articles 07 and 42 of this volume). The two channels run in opposite directions but serve the same purpose — classification authority ultimately rests with the Ministry of Health, and a company’s own determination is only the initial step.
Insight four: a change of classification triggers re-registration as a new product, which is the most expensive classification error. Article 10.5 provides that where a product changes and the change alters its risk classification, it is treated as a new product and a new registration must be filed under Articles 11, 12 or 13; Article 10.9 further provides that a change of intended use (indication) that alters the risk classification likewise requires re-registration. The cost of a classification error is therefore not one supplementary document but starting over. By comparison, changes that do not alter the classification proceed through the 19 categories of post-registration changes in Article 18 at far lower cost.
4. Practical recommendations
- Maintain a classification determination worksheet: first determine general EMB or IVD, then apply the rule numbers one by one, recording the applicable rule, the reasons for excluding other rules and the conclusion, and flag whether the highest-risk rules of Articles 8.1 and 8.2 are triggered.
- Determine IVD systems as a whole: classify analysers, reagents and software in the highest risk class and configure the dossier once.
- For new products not covered by the rules, apply to the Ministry of Health for an administrative resolution under Article 8.3; do not improvise by analogy with nearby rules.
- Before any design change or intended-use extension, perform a classification impact assessment; where the classification would change, plan a new registration under Articles 10.5 and 10.9 rather than the change channel.
- Place the classification conclusion and its reasons at the front of the dossier as the comparison baseline for later changes and renewals.
5. Verification checklist
- □ Determined whether the product is a general EMB or an IVD (deciding whether A.1 or A.2 applies)
- □ Rules applied one by one with rule numbers and conclusions recorded
- □ Checked whether the highest-risk rule of Article 8.1 is triggered
- □ IVD systems checked against the combination rule of Article 8.2
- □ Items not covered by the rules submitted for an administrative resolution under Article 8.3
- □ Changes and intended-use extensions assessed for classification impact (Articles 10.5 and 10.9)
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Costa Rican market. Our team manages Ministry of Health (Ministerio de Salud) sanitary registration applications, prepares Spanish-language dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us