EMB Definition and the Treatment of Software and Medical Apps
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Costa Rica does not use the term “medical device” (dispositivo médico); the statutory concept is biomedical equipment and materials (Equipo y Material Biomédico, EMB). Article 2.1 of the main regulation expressly extends its scope to EMB accessories, software and medical mobile applications (apps), and definition 4.2 further defines a mobile application as two kinds: an application that itself meets the criteria for classification as EMB, and an application used as an accessory to an already regulated EMB or that converts a mobile platform into a regulated EMB. Standalone software (SaMD) and mobile apps therefore fall within EMB jurisdiction in Costa Rica; they are not outside the law.
- EMB, not dispositivo médico, is the statutory term in Costa Rica.
- Article 2.1 covers EMB accessories, software and medical mobile apps.
- Definition 4.1 requires an accessory to be specifically designed by the manufacturer and not to achieve the intended function on its own.
- Software is classified under the active-device rules of Annex A (rules 8–11), by function rather than by carrier form.
- Software validation study is a listed dossier item for Clase 3 (Article 12.2.4), not an optional extra.
2. Legal basis
The provisions below set out how EMB, accessories, software and mobile applications are brought into scope, defined and classified under RTCR 505:2022.
| Level | Instrument | Provision | Key point |
| Technical regulation | RTCR 505:2022 | Article 2.1 | Applies to EMB, their accessories, software and medical mobile applications (apps de uso médico) |
| Technical regulation | RTCR 505:2022 | Definition 4.2 | Aplicación para dispositivos móviles: a mobile application meeting the criteria for EMB; or used as an accessory to a regulated EMB, or converting a mobile platform into a regulated EMB |
| Technical regulation | RTCR 505:2022 | Definition 4.1 | Accesorio: an item specifically designed by the manufacturer for use together with an EMB to support, supplement or enhance its performance, not achieving the intended function on its own |
| Technical regulation | RTCR 505:2022 | Definition 4.59 | Relevant definitions for software as a medical device (Article 4 of the main regulation contains 62 definitions in total) |
| Technical regulation | RTCR 505:2022 | Article 11.4.1 a) | Technical specifications must include a detailed description of the EMB and its identification code |
| Technical regulation | RTCR 505:2022 | Article 12.2.4 | Clase 3 submissions must include a software validation study (estudio de validación del software), where applicable |
| Technical regulation | RTCR 505:2022 | Article 8.2 | Analysers, reagents and software used with an IVD are classified, together with the companion IVD, in the higher risk class |
| Technical regulation | RTCR 505:2022 | Annex A rules 8, 9, 10 and 11 | Each classification rule for active devices expressly states “incluido cualquier software” (including any software) |
3. Expert insights
Insight one: the main regulation uses EMB rather than dispositivo médico, and the terminology difference directly affects searching and filing. Chinese-language materials often render Costa Rica’s EMB simply as medical device, but when preparing Spanish-language documents, searching Ministry of Health forms, or checking Regístrelo portal fields, the statutory term EMB must be used. Article 1 of the main regulation opens with “en adelante denominados EMB” and uses only the abbreviation thereafter. In multi-country registration programmes run in parallel, companies should take particular care not to mix Costa Rican terminology with that of neighbouring countries that use dispositivo médico.
Insight two: the definition of accesorio contains three qualifying conditions, all of which must be met. Definition 4.1 requires that an accessory be specifically designed by the manufacturer, used together with an EMB, and intended to support, supplement or enhance that EMB’s performance, while not achieving the intended function on its own. The third condition is the key exclusion: if an item can achieve a medical function independently, it is not an accessory but an EMB in its own right and requires its own registration. Article 7.6 of the main regulation (covered in detail in article 21 of this volume) builds on this by allowing accessories to be registered separately.
Insight three: software classification does not follow a standalone rule; it is determined by the host function or by the active-device rules. The main regulation does not create a dedicated software rule in Annex A; instead it writes software explicitly into four rules for active devices: rule 8 (ionizing radiation, including software that controls or monitors such equipment, Clase 3), rule 9 (active therapeutic devices, Clase 2 upwards), rule 10 (active diagnostic devices, Clase 2 upwards), and rule 11 (devices for infusion or removal of substances, Clase 2 upwards). The implication is that the class of software is determined by the function it controls or performs, not by its carrier form. Software for mammography (rule 8 b)) falls in Clase 3, while general radiography software from the same manufacturer falls in Clase 2.
Insight four: the software validation study is an expressly listed dossier item for Clase 3, not an option. Article 12.2.4 writes “estudio de validación del software, si aplica” into the clinical evidence list for Clase 3. For products whose core is an algorithm, this item often determines whether the review can be passed; the Clase 2 dossier list (Article 11) does not list it separately. When planning timelines, companies should prepare software validation materials to the Clase 3 standard so as to avoid being asked for supplements mid-review.
4. Practical recommendations
- At the product-characterization stage, first determine whether the item is a standalone EMB or an accessory: items that achieve a medical function independently proceed as standalone EMB through registration; items that only support another device are handled as accessories, with the host device identified in the technical specifications.
- Classify software and apps by function, applying Annex A rules 8–11, and note the different baseline classes of the four rules for radiation, therapy, diagnosis and infusion.
- For Clase 3 and above, prepare the software validation study early (Article 12.2.4) and state the version and the algorithm’s purpose in the technical specifications.
- Draft the technical specifications item by item under Article 11.4.1 a) to l); the “detailed description and identification code” of software must include the version number and deployment form.
- Use the EMB term uniformly in all Spanish-language documents; do not carry over the dispositivo médico wording used in neighbouring countries.
5. Verification checklist
- □ Confirm the product is an EMB (including accessories, software and medical apps) and not an excluded item
- □ Determine whether the item is a standalone EMB or an accessory (definitions 4.1 and 4.2)
- □ Classify software by function under Annex A rules 8–11
- □ Clase 3 projects have the software validation study prepared (Article 12.2.4)
- □ Technical specifications include version, identification code and deployment form (Article 11.4.1 a))
- □ Spanish-language documents use the EMB term uniformly
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Costa Rican market. Our team manages Ministry of Health (Ministerio de Salud) sanitary registration applications, prepares Spanish-language dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us