Equivalence Recognition: The Eight Reduced Documents

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Article 17 of the main regulation establishes a conditional reduction mechanism: for EMB that have been manufactured, registered and already sold in a country whose registration system has been recognised as equivalent, the Ministry of Health requires only eight documents (17.2.1 to 17.2.8), namely the application form, technical and medical specifications, the CLV, the radiation source authorisation where applicable, the label, payment of fees, the third-party manufacturing contract, and the power of attorney. What is waived are the heavy documents such as the BPM certificate and clinical evidence. However, Article 17.3, as amended in 2024, retains the right of the Ministry of Health to demand missing documents at any time, and it identifies three break situations in which the missing documents must be completed at renewal or change.

2. Legal basis

The table below lists the recognition basis, the eight-item reduced list item by item, the 2024 amendment to Article 17.3, and the Annex B procedure on which recognition rests.

LevelInstrumentProvisionKey point
Technical regulationRTCR 505:2022Article 17.1Pursuant to Article 6.1 of the WTO Agreement on Technical Barriers to Trade, the Ministry of Health may accept conformity assessment results from other registration systems
Technical regulationRTCR 505:202217.1.1 to 17.1.3Recognition basis: experience and trust, objective demonstration of an appropriate level of protection, and effective sanitary control systems
Technical regulationRTCR 505:2022Article 17.2Products from recognised countries submit only eight documents
Technical regulationRTCR 505:202217.2.1Application form (digitally signed)
Technical regulationRTCR 505:202217.2.2Technical and medical specifications (per Article 11.4)
Technical regulationRTCR 505:202217.2.3CLV (per Article 11.3)
Technical regulationRTCR 505:202217.2.4Radiation source authorisation (per Article 11.6, where applicable)
Technical regulationRTCR 505:202217.2.5Label (per Article 11.7)
Technical regulationRTCR 505:202217.2.6Payment according to the ministerial tariff
Technical regulationRTCR 505:202217.2.7Third-party manufacturing contract (per Article 11.8)
Technical regulationRTCR 505:202217.2.8Power of attorney (per Article 11.9)
Technical regulationRTCR 505:2022Article 17.3 (2024 amendment)The Ministry of Health may require any waived document on technical and scientific grounds; three situations require completion at renewal or change
Technical regulationRTCR 505:2022Annex BProcedure for recognising equivalence of registration systems, including revocation after recognition

3. Expert insights

Insight 1: Article 17.1 writes the recognition basis as three items, and all three are standards that can be argued about rather than mechanically proven. Item 17.1.1 refers to experience, knowledge and existing trust in the relevant measures; 17.1.2 requires that the norms of the recognised country objectively demonstrate that they achieve an appropriate level of protection; 17.1.3 requires that the sanitary control system of the recognised country enforces these requirements effectively and efficiently. All three carry room for judgement, so recognition does not happen automatically and must go through the Annex B procedure. This also explains why the regulation calls recognition a reconocimiento, an acknowledgement, rather than an automatic equivalence.

Insight 2: The eight-item list retains the CLV, the label and the power of attorney, and these three are precisely the carriers of the source-country identity. Compared with the eleven categories of Article 11, what is waived are the BPM certificate under Article 11.5 and the clinical evidence of Clase 3 and 4, while the radiation source authorisation is retained at 17.2.4. The CLV at 17.2.3 and the label at 17.2.5 are expressly retained. This design of reducing the burden without reducing identity verification is exactly why the 2024 amendment wrote the CLV issuing country into the break situations of Article 17.3: the CLV is the direct credential for judging whether the product still enjoys the protection of mutual recognition.

Insight 3: The precondition for applying Article 17.2 is the three parallel conditions of having been manufactured, registered and sold in the recognised country. The source text reads EMB fabricados, registrados y de venta en los países. None of the three states can be missing: a product registered in the recognised country but not sold there, or sold there but manufactured in a third country, does not automatically qualify for the eight-item list. This parallel condition is the first sieve a company should apply when self-checking whether the reduced list is available.

Insight 4: The hidden cost of the reduction is the uncertainty of demands for completion at any time. Article 17.3 as amended states expressly that the Ministry of Health may, during the registration process or after registration, require the submission of any waived document on technical and scientific grounds. The meaning is that the documentary obligations of a dossier filed on the eight-item list have not disappeared; they have only been deferred. Companies should budget and schedule projects on the assumption that the BPM certificate and clinical evidence may be demanded at any time, rather than treating the reduction as a permanent exemption.

4. Practical recommendations

  • Self-check eligibility for the reduced list against the three parallel conditions of Article 17.2: manufactured, registered and sold in the recognised country.
  • File on the eight-item list (17.2.1 to 17.2.8) while keeping the BPM certificate and clinical evidence ready in parallel, to answer a demand under Article 17.3.
  • Bring the CLV issuing country into change management: before switching the issuing country, assess whether the completion obligation of Article 17.3 is triggered, as detailed in article 19 of this series.
  • Track the status of recognition resolutions under Annex B, since revocation of a recognition triggers the completion obligation.
  • Allocate the completion obligation to the licence holder in the commercial arrangements, and agree the deadlines and cost bearing for document provision in the contract.

5. Verification checklist

  • □ Self-check completed against the three conditions of manufactured, registered and sold
  • □ Source country confirmed to be on the current equivalence recognition list
  • □ Dossier organised on the eight-item list, 17.2.1 to 17.2.8
  • □ BPM certificate and clinical evidence ready for a demand under Article 17.3
  • □ CLV issuing country recorded and brought under change monitoring
  • □ Revocation status of Annex B recognition resolutions tracked

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Costa Rican market. Our team manages Ministry of Health (Ministerio de Salud) sanitary registration applications, prepares Spanish-language dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail