Azerbaijan Medical Device Registration 2026: Biosafety Information and When It Can Be Exempted
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Article 4.2.6 requires the biosafety information provided by the manufacturer, while Article 4.3 provides that this item does not apply to devices that do not come into contact with the human body when used. Of the three exemptions in Article 4.3, this is the easiest to judge and also the most frequently misapplied one.
The test is whether the device comes into contact with the human body when used (insanın bədəni ilə təmasda olmayan tibb vasitələri). This test is closely related to Article 3.1.5 of Decision No. 8 (contact or interaction with human tissues or organs), but the two serve different purposes: Article 3.1.5 is used for classification, while Article 4.3 is used for exemption from a dossier item. A device may therefore be exempt from the biosafety documents yet still be classified with reference to the contact dimension.
2. Legal basis
- Resolution No. 345 of the Cabinet of Ministers of the Republic of Azerbaijan of July 18, 2024 on the Rules for the State Registration of Medicinal Products, Drug Substances and Medical Devices, Their Entry in the State Register and the Maintenance of the State Register (effective July 19, 2024; Xalq qəzeti No. 151 of July 20, 2024; Legislative Compilation 2024, No. 7, item 906; repealing Resolution No. 108 of July 13, 2007 as of July 20, 2024): Article 4.2.6 (biosafety information), Article 4.2.7 (additional documents for sterile devices), Article 4.3 (4.2.6 does not apply to devices that do not contact the human body), Article 5.1.4 (inconsistent documents are a ground for rejection).
- Decision No. 8 of the Collegium of the Ministry of Health of the Republic of Azerbaijan of May 21, 2024 on the Classification of Medical Devices by Risk Degree (adopted under the second sentence of Article 6.8 of the Law on Medicinal Products; signed by Minister Teymur Musayev; sent to the Ministry of Justice within 3 days of adoption for entry in the register of legal acts): Article 3.1.5 (contact or interaction with human tissues or organs), Article 5.6 (classification of non-invasive devices contacting injured skin or mucous membranes), Articles 8.7 and 8.8 (products manufactured from human or animal tissues or cells).
- Law of the Republic of Azerbaijan on Medicinal Products (Dərman vasitələri haqqında), No. 208-IIIQ, adopted December 22, 2006, effective February 28, 2007, Legislative Compilation 2007, No. 2, item 70, as amended by Law No. 979-VIQD of July 14, 2023 (adding Articles 1-1 and 9-1 with new wording, Legislative Compilation 2023, No. 8, item 1127): Article 1.0.14 (definition of the quality of a device).
3. Expert insights
Insight 1: What is exempted is the document, not the assessment. Article 4.3 removes the obligation to submit biosafety information, but it does not make biosafety irrelevant. Article 3.1.5 of Decision No. 8 still lists contact or interaction with human tissues or organs as a classification dimension, and Article 5.1.4 still permits rejection where inconsistencies related to safety are found in the documents. The conclusion should first be assessed and recorded in the classification rationale document, and only then should the decision on submitting the information be made.
Insight 2: The judgement of no contact with the human body must be made at concrete steps of use. Many devices do not touch the human body during their principal use but do so during cleaning, calibration or sampling. For example, an in vitro diagnostic instrument itself never enters the human body, but its sampling consumables do; imaging equipment does not touch the patient, but coupling gel or positioning pads do. Judge by the least favourable step of use, not by the main function, and document the reasoning for each step evaluated.
Insight 3: Article 4.2.6 and Article 4.2.7 are two independent lines and cannot replace each other. Item 4.2.6 is biosafety information (exempt where there is no contact with the human body), and 4.2.7 is additional documentation for sterile devices (with no such exemption). A device that is both sterile and body-contacting requires both; a device that is sterile but non-contacting, such as certain sterile-packaged in vitro consumables, may be exempt from 4.2.6 but still requires 4.2.7.
Insight 4: Articles 8.7 and 8.8 add a further step up for materials of biological origin. Devices manufactured using devitalised or inactivated human or animal tissues, cells or their derivatives are classified in the highest degree (8.7), and products of animal origin that only contact intact skin are classified in the high degree (8.8). Such products both contact the human body and are of biological origin, so the requirements under 4.2.6 are the heaviest, and the biological evaluation is usually prepared following the approach of the ISO 10993 series (practical estimate: the Rules do not name a specific standard; follow the requirements of the AEC).
4. Practical recommendations
- Judge contact with the human body by the least favourable step of use, and record the conclusion in the classification rationale document.
- Where the conclusion is no contact, invoke the exemption of Article 4.3 for item 4.2.6 in the covering letter and state the basis of the judgement.
- For body-contacting devices, compile the biosafety documentation and the sterile documentation under 4.2.7 into separate volumes, each marked with its clause number.
- Reassess the risk degree for materials of biological origin under Articles 8.7 and 8.8, and do not treat them as ordinary contact devices.
- Keep the biological statements consistent with the other documents (4.2.3 and 4.2.8) to avoid triggering Article 5.1.4.
- Because the Rules do not name a specific biological evaluation standard, apply the practical estimate of following the AEC dossier requirements and retain written confirmation.
5. Verification checklist
- □ Contact with the human body has been judged by the least favourable step of use.
- □ The conclusion has been recorded in the classification rationale document with the clause basis noted.
- □ The exemption situation has been invoked under Article 4.3 in the covering letter.
- □ The biosafety and sterile documentation to be submitted have been compiled into separate volumes.
- □ Materials of biological origin have been reassessed under Articles 8.7 and 8.8.
- □ The biological statements have been checked for consistency with 4.2.3 and 4.2.8.
Related reading: Additional Documents for Sterile Devices: Validation, Bioburden, Pyrogens and Sterility (https://www.grzanglobal.com/azerbaijan-medical-device-registration/sterile-device-additional-documents/); Notarised Translation of Instructions for Use: Only for Devices with Medicinal Substances (https://www.grzanglobal.com/azerbaijan-medical-device-registration/notarized-translation-of-instructions-for-use/); Declaration of Conformity with Safety and Effectiveness Requirements and the Risk Degree (https://www.grzanglobal.com/azerbaijan-medical-device-registration/safety-effectiveness-declaration-risk-level/).
GRZAN provides regulatory consulting for medical device manufacturers and exporters, covering Azerbaijan state registration dossiers, biological evaluation strategy, risk classification and exemption reasoning. For tailored support on this topic, contact GRZAN at info@grzanglobal.com.
Regulatory information verified as of October 30, 2026.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us