Armenia Medical Device Registration Exemptions 2026: Special Channels and Testing Waivers

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

The EAEU medical device registration rules do not apply to two categories of devices by design: devices needed in a state of emergency, and devices used to diagnose emerging, naturally endemic or especially dangerous infectious diseases — both categories are governed by the legislation of the member states. In addition, there is a list of products exempt from technical testing (control materials, calibrators, cleaning solutions and culture media). The GRZAN judgment: exemption from registration in Armenian projects is a narrow lane and should not be used to design the main pathway; the genuinely routine burden-reduction openings are the exemption from technical testing and the import-document reduction for Class 1 devices. Companies that build their market-entry plan around an emergency or epidemic exemption will find that the exemption depends on an authority determination they cannot obtain commercially.

2. Legal basis

Legal instrumentReference / numberPublication (entry into force)Provisions
Rules for the Registration and Expert Assessment of Medical Devices for Safety, Quality and EfficacyEEC Council Decision No. 46 of 2016-02-12Amended by Council Decision No. 144 of 2021-12-24 (enters into force 180 calendar days after publication)Article 1 (exclusion from scope)
Rules for the Conduct of Technical Testing of Medical DevicesEEC Council Decision No. 28 of 2016-02-12Amended by Council Decision No. 25 of 2022-03-17List of products exempt from technical testing
Procedure for Importing Medical Devices into the Territory of Armenia and the List of Documents Required for Import Expert AssessmentGovernment of the Republic of Armenia Decision No. 429-Ն of 2023-03-30Amended by No. 1619-Ն of 2024-10-17Annex 1, point 10.1
Law of the Republic of Armenia On Medical Assistance and Services to the PopulationNo. ЗР-42 of 1996-03-04Amended by Law No. ЗР-504 of 2024-12-17Chapter 10

3. Expert insights

Insight 1: the exclusion in Article 1 of Decision No. 46 is a single, narrowly framed registration-free channel. The second paragraph of Article 1 of the rules as replaced provides that the rules do not apply to medical devices needed in emergency situations or to devices used for the diagnosis of emerging, naturally endemic or especially dangerous infectious diseases, whose circulation is regulated by the legislation of the member states. GRZAN notes three points: first, the trigger is an emergency situation or diagnosis of the specified infectious diseases, which is a public-health emergency attribute; second, what is excluded is not the registration obligation itself but the application of the Union registration rules, which are replaced by national procedures of the member state; third, a company cannot self-declare that it qualifies — the applicability must be established by the competent authority.

Insight 2: the list of products exempt from technical testing was expanded in 2022. After Council Decision No. 25 of 2022-03-17 amended Decision No. 28, the list of products not subject to technical testing was supplemented with control materials, calibrators, cleaning solutions and culture media, on the grounds that technical testing does not apply to such products. GRZAN notes that this is a substantive burden reduction for IVD companies: control materials and calibrators were previously often required to undergo technical testing, and they can now be removed from that stage of the documentation plan altogether.

Insight 3: the 2022 amendment also lowered the qualification threshold for laboratory personnel. The same amendment package reduced the mandatory work experience of testing laboratory specialists from 3 years to 2 years and standardized the forms of records for technical testing and for biological evaluation testing. GRZAN notes: this change widens the pool of eligible laboratories, and the compliance criteria companies use when selecting a testing site must be updated accordingly, otherwise valid laboratories are rejected on outdated standards.

Insight 4: biological evaluation results can be extended across a group of products. After Council Decision No. 26 of 2022-03-17 amended Decision No. 38, it became explicit that the study results for standard samples may be extended to a group of similar medical devices. GRZAN notes: for product lines with many models, such as dressings, catheters and infusion sets, this rule can significantly reduce testing cost and duration, but a group justification must be completed first (the same legal manufacturer, the same technical documentation, the same constituent materials, and the same scope and conditions of use).

4. Practical recommendations

  1. Do not build the main pathway on the emergency lane: the exclusion for emergency and infectious-disease-diagnosis devices requires an authority determination, and commercial projects should not be planned on that basis.
  2. Make the testing-exemption judgment first for IVD products: before control materials, calibrators, cleaning solutions and culture media enter a testing plan, record a written judgment on whether they fall within the exempt list.
  3. Use the group-extension rule fully: for product lines with many models, organize a representative sample plus a group justification first, and write the extension mechanism of Decision No. 38 into the testing plan.
  4. Update laboratory qualification standards: enter the requirement of 2 years of professional work experience into the supplier qualification checklist.
  5. Claim the Class 1 import-document reduction: Class 1 non-sterile, non-IVD products may substitute batch quality certificates for the declaration of conformity under Annex 1, point 10.1, of Decision No. 429-Ն.

5. Verification checklist

  1. □ It has been confirmed that the product does not fall within the exclusion of Article 1 of Decision No. 46, or an authority determination has been obtained.
  2. □ Each IVD product has been judged item by item against the exempt list (control materials / calibrators / cleaning solutions / culture media).
  3. □ The representative-sample justification for group extension of biological evaluation results has been completed and written into the testing plan.
  4. □ The qualification standards for testing laboratories have been updated (2 years of professional experience required).
  5. □ Whether the Class 1 product qualifies for the import-document reduction under point 10.1 has been confirmed.
  6. □ The amendment content of Decisions No. 25 and No. 26 of 2022 has been merged into internal work standards.
  7. □ Communication records with the competent authority on the applicability of the exclusion have been established.

Regulatory information verified as of October 30, 2026.

GRZAN services. GRZAN International assesses testing exemptions, group-extension strategies and import-document reductions for EAEU and Armenian projects. For inquiries, contact info@grzanglobal.com.

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail