Armenia Medical Device Recalls: Corrective Actions and Field Safety Actions under EAEU Rules

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Union rules distinguish between corrective action and field safety corrective action for medical devices: the former targets the cause of an identified nonconformity or adverse event, while the latter targets the risk of death or serious deterioration in health. The GRZAN conclusion: the trigger criterion for a field safety corrective action is the severity of the risk consequence (death or serious damage to health), not the number of nonconformities; the measures may include returning the medical device to the manufacturer or its authorized representative. Because the two concepts run on two separate internal processes, misclassification at the intake stage is one of the most common audit findings.

2. Legal basis

RegulationReference numberPublication date (entry into force)Provision
Rules for monitoring the safety, quality and effectiveness of medical devicesDecision No. 174 of the Board of the Eurasian Economic Commission, 2015-12-2230 calendar days after the entry into force of the AgreementArticle 4(c); Article 5 (definitions of corrective action and field safety corrective action)
Requirements for the implementation, maintenance and assessment of a quality management system for medical devices based on potential use riskDecision No. 106 of the Council of the Eurasian Economic Commission, 2017-11-10Amended by Council Decision No. 43 of 2026-04-27Definitions of corrective action / correction / critical supplier / nonconforming medical device
Law of the Republic of Armenia on Medical Care and Services for the PopulationNo. ЗР-42 of 1996-03-04Amended by No. ЗР-504 of 2024-12-17Article 45

3. Expert insights

Insight 1: The dividing line between the two concepts is severity of consequence. Article 5 of Decision No. 174 defines corrective action as action taken by the medical device manufacturer to eliminate the cause of an identified nonconformity or adverse event; field safety corrective action means action taken by the manufacturer to reduce the risk of death or serious deterioration in health associated with the use of the medical device for users or third parties, and such action may include returning the medical device to the manufacturer or its authorized representative. GRZAN note: the two concepts correspond to two internal processes — the former is a routine CAPA, the latter is a post-market field safety action including recall; the trigger thresholds and external communication requirements are entirely different.

Insight 2: Decision No. 106 provides the supporting QMS definitions. That decision defines corrective action as action taken by the manufacturer to eliminate the cause of an identified nonconformity or adverse event; correction as action taken to eliminate an identified nonconformity; and nonconforming medical device as a device that does not meet the applicable requirements. GRZAN note: in the QMS context, correction and corrective action are two different actions — a correction addresses the symptom (such as rework or repair), while a corrective action addresses the cause (such as changing a process or a document). During manufacturing inspections this distinction is verified.

Insight 3: Corrective action is one of the three bases of surveillance. Article 4(c) of Decision No. 174 provides that surveillance is based on the manufacturer’s system for collecting and analyzing safety and effectiveness data in the post-market phase and on corrective actions implemented under the QMS requirements. GRZAN note: this means the execution records of corrective actions are a component of the surveillance system, will be consulted during oversight, and are an object of assessment in manufacturing inspections.

Insight 4: Decision No. 106 also defines the critical supplier. A critical supplier is defined as a supplier whose products or services have a direct impact on the safety and/or effectiveness of the medical device. GRZAN note: if a corrective action involves a critical supplier (for example outsourced sterilization or key components), the action must extend to the supplier level, and disposition records on the supplier side must be retained.

Insight 5: Execution workflow, as applied by GRZAN.

  1. Event identification and grading: determine whether a field safety corrective action is triggered (risk of death or serious damage to health) based on the severity of consequences.
  2. Cause analysis: distinguish correction (handling the nonconformity that has occurred) from corrective action (eliminating the cause).
  3. Measure determination: decide the type of measure, including whether devices must be returned to the manufacturer or its authorized representative.
  4. Scope definition: define the affected range by batch or serial number, and include critical suppliers where involved.
  5. External communication: report to the competent authority through the reporting route of Decision No. 174, and notify users and the authorized representative.
  6. Execution and verification: implement the measures, verify their effectiveness, and retain all records.
  7. Integration into the surveillance system: add the disposition records to the post-market data system for use by the unified information database and manufacturing inspections.

4. Practical recommendations

  1. Establish a two-level decision standard: state in the CAPA procedure the trigger threshold for a field safety corrective action (risk of death or serious damage to health), so that an action is not left unstarted when it should begin.
  2. Separate correction from corrective action: use two distinct fields in CAPA forms so the distinction can be shown directly during manufacturing inspections.
  3. Extend measures to critical suppliers: for measures involving critical suppliers, obtain disposition records and remediation evidence from the supplier side.
  4. Design the return channel in advance: state in the authorization letter the authority to accept returned medical devices and the logistics arrangements, avoiding ad hoc coordination.
  5. Route records into the surveillance system: bring all disposition records into the post-market data system for consultation during surveillance and inspections.

5. Verification checklist

  1. □ The trigger threshold for a field safety corrective action (death or serious damage to health) is stated in the CAPA procedure.
  2. □ Correction and corrective action are distinguished and given separate fields in the forms.
  3. □ It has been determined whether the measures include returning devices to the manufacturer or its authorized representative.
  4. □ The affected scope is defined by batch or serial number.
  5. □ For measures involving critical suppliers, disposition records from the supplier side have been obtained.
  6. □ The competent authority has been notified, users and the authorized representative informed, and submission records retained.
  7. □ Disposition records have been added to the post-market data system and are available for surveillance and manufacturing inspections.

Regulatory information verified as of October 30, 2026.

About GRZAN: GRZAN supports medical device manufacturers entering Armenia and the Eurasian Economic Union with regulatory registration, quality management systems, post-market surveillance and import clearance compliance. For inquiries, contact info@grzanglobal.com.

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail