Who Must Notify: Sponsor Status and Exempt Devices

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

The WAND notification duty in New Zealand turns on two conditions: the notifier must be a Sponsor (the New Zealand sponsor) and the device must not be an exempt device.

The subject condition appears in reg 3(a) of the 2003 Regulations (SR 2003/325): a person in New Zealand who exports or arranges exportation of the device, who imports or arranges importation, or who manufactures or arranges manufacture for supply. Reg 3(b) excludes persons acting on behalf of a New Zealand resident or a person carrying on business in New Zealand. Those that must notify are importers, exporters and local manufacturers; those that need not include retailers, downstream users such as hospitals, pure agents and carriers.

The object condition sits in reg 5(1): the database information targets each device that is not an exempt medical device. The exempt list sits in Schedule 1, 9 classes from (a) to (i). The notification acts are reg 6 (within 30 working days) and reg 7 (declaration).

Lawfully supplied must be read on two levels: Medsafe’s homepage states, For medical devices to be legally supplied in New Zealand they must be notified to the WAND database.; but SR 2003/325 sets no prohibition on sale without notification, enforcement resting on the reg 6 and 8 time-limit duties, breaches under reg 14 punishable by a fine up to NZ$500. On Medsafe’s reading, notification occurs after first arrival in New Zealand, not as a premarket precondition; the nine Schedule 1 classes, including all IVDs, may be lawfully supplied without notification. Unnotified does not mean illegally supplied.

Only medical devices enter the system. The sole source of the definition is s 3A of the Medicines Act 1981, with the s 4 therapeutic purpose provision; reg 3 of the 2003 Regulations likewise provides that medical device has the same meaning as it has in section 3A of the Act.

2. Legal basis

The table maps each obligation element onto its verified source.

Sponsor (duty subject) definitionMedicines (Database of Medical Devices) Regulations 2003SR 2003/325; made 17 November 2003, in force 1 January 2004regs 3(a)(i)(ii)(iii) and 3(b)
Entry limited to non-exempt devicesMedicines (Database of Medical Devices) Regulations 2003SR 2003/325reg 5(1); reg 5(2) (UPI for Class III and AIMD); reg 5(3) (same-manufacturer Class I/IIa/IIb devices may be grouped as a kind by GMDN and classification)
Notification time limit and declarationMedicines (Database of Medical Devices) Regulations 2003SR 2003/325reg 6 (30 working days), reg 7(a)(b)(c)
9 exempt device classesMedicines (Database of Medical Devices) Regulations 2003SR 2003/325Schedule 1(a)–(i)
PenaltiesMedicines (Database of Medical Devices) Regulations 2003SR 2003/325reg 14 (Sponsor breach of regs 6, 8, 9, 10; fine on conviction up to NZ$500)
Medical device definitionMedicines Act 19811981 No 118; in force 1 August 1984s 3A (inserted by 2013 No 141 s 6, in force 1 July 2014)
Therapeutic purposeMedicines Act 19811981 No 118s 4 (replaced by 2013 No 141 s 7, in force 1 July 2014)
Device safety (not lifted by exemption)Medicines Act 1981Medicines Act 1981s 38
Labelling and advertising (not lifted by exemption)Medicines Regulations 1984SR 1984/143; in force 1 August 1984reg 12(4), regs 10, 11, 25
Medsafe official positionMedsafe WAND pagehttps://www.medsafe.govt.nz/regulatory/DevicesNew/3WAND.asp, revised 1 July 2014Notification mandatory for importers, exporters and local manufacturers; free, no ongoing fees

The nine Schedule 1 exempt classes (verified class by class): (a) devices that are also medicines within s 3(1) of the Medicines Act 1981 (except where rule 13(2) of Schedule 2 applies); (b) devices made at the request of a registered health professional to design features specified by that professional, for the sole use of a particular patient; (c) devices supplied to a medical or dental practitioner for the particular patient; (d) devices imported and under New Zealand Customs control pending export to another country; (e) devices commonly known as in vitro diagnostic devices, for use only within a particular laboratory; (f) diagnostic devices manufactured in a laboratory for use in another laboratory (LDTs); (g) devices imported by a person solely for their own use; (h) devices imported for a clinical trial; (i) devices of a class declared exempt by the Director-General by notice in the Gazette.

3. Expert insights

Insight 1: Subject and object are two independent judgement lines. The common error merges them. An importer of laboratory-developed test products is a Sponsor (reg 3(a)(ii)), but the products fall within Schedule 1(f), so no duty arises; conversely, a device outside the nine exemptions imported by an individual solely for own use (Schedule 1(g)) also generates no duty. GRZAN recommends a two-column table per consignment, entering the workflow only when both answer yes.

Insight 2: Exemption lifts the notification duty, not all duties. Schedule 1’s wording is exempt medical device; the reg 5(1) entry and reg 6 notification duties therefore do not apply to the device. The safety requirements of s 38 of the Medicines Act 1981, labelling under reg 12(4) of the 1984 Regulations, and advertising and information duties under regs 10, 11 and 25 do not condition on exemption status. Clinical trial devices (Schedule 1(h)) need no notification, but their instructions for use, safety responsibilities and post-market handling remain subject to current rules.

Insight 3: Exporters are inside the subject class too. Reg 3(a)(i) places a person in New Zealand who exports or arranges exportation within the Sponsor definition, so a business manufacturing entirely for export equally bears the reg 6 duty. This is often misread as no notification needed for export-only products.

Insight 4: Exempt class (i) is alive, and IVDs come from it. Schedule 1(i) authorises the Director-General to declare exempt classes by Gazette notice, so item (i) is an open entrance, not a static list. The IVD exemption arises this way: the Director-General made the Exemption Medical Devices (In Vitro Diagnostic Devices) Notice (No. 1) 2003 on 12 December 2003, in force 1 January 2004. Medsafe recognises no IVD risk grading model, using uniformly the IVD code with 5-digit GMDN codes.

4. Practical recommendations

  • Characterise first, identify the person second, test exemption last: s 3A and s 4 determine whether the product is a medical device; regs 3(a) and 3(b) determine who the Sponsor is; Schedule 1(a)–(i) determines whether it is exempt. Notification starts only when all three pass.
  • Build an exemption determination trail for every SKU: record each SKU’s exemption class letter and reasoning in the product register, so the company can state which class applies and why.
  • Schedule notification on the strictest reading. Regulation 6 states 30 working days from becoming Sponsor; Medsafe’s own pages are inconsistent, the WAND homepage stating 30 calendar days and the regulatory requirements page 30 working days from first arrival, the statutory reading being reg 6. GRZAN recommends internal deadlines on the strictest reading, 30 calendar days from first arrival (a practical estimate).
  • Keep exempt devices inside post-market management; recall, adverse event reporting and distribution record duties do not disappear with exemption; initial death or serious injury reports are due as soon as possible and within 10 calendar days via devices@health.govt.nz.
  • Review exempt class (i) periodically. Gazette notices may adjust exempt classes; GRZAN recommends adding legislation.govt.nz records and Medsafe’s exempt device page to the review list (a practical estimate).

5. Verification checklist

  • □ Product characterisation is complete under s 3A and s 4, basis recorded
  • □ Reg 3(a) confirms the company as Sponsor, excluding the reg 3(b) agent case
  • □ Schedule 1(a)–(i) has been compared class by class, exemptions recorded by letter
  • □ Exemption is confirmed to lift only the notification duty; s 38, labelling and advertising duties still apply
  • □ IVD products use the IVD code and 5-digit GMDN codes; export-only businesses have confirmed reg 3(a)(i) likewise creates Sponsor status
  • □ The notification time limit is set on the strictest reading, with a clear start point
  • □ The 10 calendar day deadline for death or serious injury reports and the devices@health.govt.nz channel are in post-market processes
  • □ External material describes notification as neither registration nor approval

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the New Zealand market. Our team manages Medsafe sponsor obligations and WAND entry requirements, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail