Clinical Evaluation Requirements and Exemptions for Medical Devices
Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Article 14 of Law No. 12/2025 requires medical devices to undergo clinical evaluation; however, Dispatch No. 5/ISAF/2026 on the catalogue of medical devices exempted from clinical evaluation lists 2 categories of exemptions: (1) devices with a clear working mechanism, finalised design and mature manufacturing processes, that do not change conventional uses, and whose similar devices have been in clinical use for many years with no record of serious medical device adverse events; and (2) other situations in which the safety and effectiveness of the medical device can be demonstrated through non-clinical evaluation. Outside the exemptions, clinical evaluation may be conducted using materials such as the clinical literature, clinical experience and clinical trial data of similar devices; a clinical trial is needed only where the existing evidence is insufficient to demonstrate safety and effectiveness.
- The first exemption category rests on 4 cumulative elements, all of which must be established together.
- The absence of a record of serious adverse events requires evidence, not a mere declaration.
- The second category is a catch-all with a demanding threshold: non-clinical evidence must independently support both safety and effectiveness.
- Exemption from clinical evaluation does not exempt the applicant from submitting clinical evaluation documents with an exemption argument.
- The similar-device comparison route should be preferred over a new trial wherever the clinical literature, clinical experience or clinical trial data of similar devices can carry the case.
2. Legal basis
The duty of clinical evaluation, the exemption catalogue and the required documents are governed by the following instruments, verified as of 29 October 2026.
| Regulation | Reference | Publication | Provisions |
| Catalogue of medical devices exempted from clinical evaluation | Dispatch No. 5/ISAF/2026 | Published in 2026; in force on 1 July 2026 | Full text of the dispatch (the 2 categories of exemptions) |
| Specific requirements for clinical evaluation of medical devices | Dispatch No. 2/ISAF/2026 | Published in 2026; in force on 1 July 2026 | Full text of the dispatch |
| Medical Device Supervision System | Law No. 12/2025 (Issue No. 30/2025 of the Official Gazette) | Published on 28 July 2025; in force on 1 July 2026 | Article 14 (clinical evaluation) |
| Implementing Rules of the Medical Device Supervision System | Administrative Regulation No. 11/2026 (Issue No. 20/2026 of the Official Gazette) | Published on 18 May 2026; in force on 1 July 2026 | Article 10(2), item (9) (clinical evaluation documents) |
3. Expert insights
Insight 1: the 4 elements of the first exemption category are cumulative. A clear working mechanism, a finalised design and mature manufacturing processes, no change to conventional uses, and years of clinical use of similar devices without a record of serious adverse events — all 4 must hold at the same time. Companies often argue only the first 2 and overlook the element of no record of serious adverse events, which requires active proof.
Insight 2: the absence of a record of serious adverse events requires evidence, not a single declaration. This element requires the applicant to provide evidence of the safety history of similar devices, which usually comes from the vigilance databases of overseas regulatory authorities, public literature, and the company’s internal complaint and adverse event records. The quality of this evidence directly determines whether the exemption is accepted.
Insight 3: the second exemption category is a catch-all, but its threshold is not low. The phrase other situations in which safety and effectiveness can be demonstrated through non-clinical evaluation means that the non-clinical evidence must independently support both conclusions, of safety and of effectiveness. For most invasive or implantable products, non-clinical evidence can hardly demonstrate clinical effectiveness on its own — such products should proceed directly to planning a clinical evaluation.
Insight 4: exemption from clinical evaluation does not mean exemption from clinical evaluation documents. Article 10(2), item (9) of the Implementing Rules lists clinical evaluation documents as a mandatory part of the registration dossier. Even where the exemption applies, a set of clinical evaluation documents arguing the applicability of the exemption must still be submitted — not a blank, and not a mere citation of the catalogue entry, but a chapter that demonstrates element by element why the device falls within the exemption and what evidence supports each element.
4. Practical recommendations
- Write the exemption argument as a standalone chapter: argue each of the 4 elements item by item with the sources of evidence attached, as the core content of the clinical evaluation documents.
- Run the safety-history search first: complete the vigilance database and literature search for similar devices before drafting, and produce a written search report.
- Plan the clinical trial early where the evidence is insufficient: do not wait until a supplement is requested; the clinical trial has the longest cycle.
- Prefer the similar-device comparison over a new trial: where the case can be made through the clinical literature, clinical experience or clinical trial data of similar devices, a new trial is unnecessary and the comparison evidence should be organised into the clinical evaluation documents directly.
5. Verification checklist
- □ Whether an exemption is possible has been assessed against the 2 categories of Dispatch No. 5/ISAF/2026
- □ The 4 elements of the first category have been argued in writing item by item
- □ The safety-history search for similar devices (including serious adverse event records) has been completed
- □ Whether non-clinical evidence can independently support safety and effectiveness has been confirmed (second category)
- □ The clinical evaluation documents (including the exemption argument) have been prepared under Article 10(2), item (9)
- □ Where evidence is insufficient, the clinical trial and prior authorisation have been planned early
Regulatory information verified as of 29 October 2026. The regulations, guidelines and effective dates described in this article are limited to the content published by the official sources listed under Legal basis as of that date; where official documents are subsequently updated, the latest published version prevails.
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for Macao, China. Our team manages Health Bureau (SSM) pre-market registration and notification applications, prepares reference-approval dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us