JMDN Generic Name Coding and Route Determination
Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
The smallest administrative unit of Japanese pharmaceutical regulation is the generic name, not the product name chosen by the company. Once the generic name is fixed, four things follow automatically: the risk class, the existence or absence of a certification standard, the filing route, and the body that conducts the quality system investigation. GRZAN lists the determination of the generic name as the first gate of a Japan project because the cost of an error here far exceeds the cost of a translation error: a translation error can be corrected, whereas a coding error forces the project to be restarted from the classification stage onward. The restart is not only a documentation exercise; the route decision, the entity decision, and the evidence plan all inherit the code and must be revisited with it.
- The generic name is the carrier of classification, and every downstream step inherits it.
- Whether certification is available depends on the generic name, not on the maturity of the technology.
- Multi-function combination products converge to one primary generic name, with remaining functions declared as accompanying features.
2. Legal basis
The coding system and its legal anchors are set out below. The designation, the essential requirements, and the certification standards are all expressed at the level of designated items, which is what makes the generic name the operative unit of the entire regime.
- PMD Act Article 2, Paragraphs 5 through 7: the designation of general, controlled, and highly controlled medical devices is conducted by item.
- PMD Act Article 41, Paragraph 3: the standards established by the Minister of Health, Labour and Welfare for medical devices (essential requirements).
- PMD Act Article 42, Paragraph 2: certification standards, being the standards established by the Minister of Health, Labour and Welfare for designated medical devices.
- PMD Act Article 23-2-13: certification for designated items; Article 23-2-5, Paragraph 1: approval for the remainder.
- Enforcement Regulations Article 114-8: the designation of the processes subject to the manufacturing business registration, which differs according to the item classification.
- The medical device generic name search published by PMDA and the certification item listings published by the Ministry of Health, Labour and Welfare are the official search entry points for generic names and the scope of certification.
3. Expert insights
Insight one: the generic name is the carrier of classification, not a translation of the product name.
Each entry of the JMDN, the Japanese medical device nomenclature, contains a generic name, a definition, a code, and a risk classification. When a company translates its Chinese product name literally into Japanese and tries to match it against the entries, it frequently lands on irrelevant entries. The correct method is to search from the intended purpose and the structural principle, and only then to check whether the definition of the candidate entry matches the product. The order matters because the wording of a product name rarely matches the wording of a definition, while the purpose and the structure are what the definition actually describes.
Insight two: the existence of a certification standard is decided by the generic name, not by the company’s technical capability.
Many Chinese companies reason that their products are mature and should therefore qualify for certification. In reality, the feasibility of certification depends entirely on whether the Ministry of Health, Labour and Welfare has established a certification standard for that generic name and has designated it. Technical maturity can only influence the quality of the dossier; it cannot change whether the route is available at all, and treating it as if it could wastes planning cycles. The only question that changes the route is whether the Ministry has spoken on that generic name, and that answer already exists before the company arrives.
Insight three: one physical product may hit several generic names.
Multi-function combination devices often fall within several generic names simultaneously. The Japanese treatment is to determine 1 generic name according to the primary intended purpose, with the remaining functions explained in the dossier as accompanying functions. Where a company files the functions as equals, the reviewers will require convergence at the review stage, and the convergence requested at that late stage is more restrictive than the convergence the company would have chosen itself. Choosing the primary purpose early, with reasons recorded, keeps that choice in the company’s own hands.
4. Practical recommendations
- Adopt the three-step retrieval method of intended purpose, then structural principle, then generic name. Fix the Japanese wording of the intended purpose first, search for entries with similar structural principles, and verify the definition last.
- Have the retrieval result double-checked by two people. Let a regulatory specialist and a Japanese-technical translator search independently; where their conclusions disagree, escalate the matter to a PMDA consultation before submission.
- Write the generic name into a fixed field of the project documents. All later materials, including the application form, the attached documentation, and the quality system materials, should use the same code to avoid inconsistency between documents.
- Confirm certification feasibility on the very day the generic name is fixed. If a standard exists, contact an RCB immediately; if none exists, prepare for approval without waiting in the hope of an exception.
- Re-run the retrieval when the product specification changes. A design change can move the product to a different entry, and the code inherited from the earlier specification then misleads every downstream document.
5. Verification checklist
- □ Was the retrieval sequence led by intended purpose rather than by a literal translation of the product name?
- □ Has the existence of a certification standard under the generic name, and any designation, been confirmed?
- □ Has the combination product been converged to 1 primary generic name?
- □ Is the generic name code recorded as a fixed field in the project documents for later reference?
- □ Has the two-person double check, or a PMDA consultation, been completed?
Regulatory information verified as of October 29, 2026.
How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, in-country representative and local testing support for the Japanese market. Our team manages PMDA submissions across Shonin/Todokede routes, coordinates J-QMS conformity, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us