China UDI: Three Implementation Batches, Scope and Database Maintenance

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

The unique device identification (UDI) consists of two parts — the device identifier (UDI-DI) and the production identifier (UDI-PI) — a structure fixed by Article 7 of the Rules for the Unique Identification System for Medical Devices.

  • Upload is a pre-marketing act: Article 15 requires the holder or filer to upload the identifier and related data to the UDI database before the product is placed on the market.
  • Implementation has advanced in three batches: Announcement No. 106 of 2020 covering part of Class III; Announcement No. 114 of 2021 covering the remaining Class III devices, including IVDs; Announcement No. 22 of 2023 covering 15 categories and 103 Class II varieties.
  • The scope is defined at the level of the secondary product category under a sub-catalogue of the Classification Catalogue. Judging scope by catalogue heading or primary category is the commonest technical error.
  • Class II devices selected in centralised procurement but outside the implementing list are not yet mandatory. Selection alone does not bring a product into scope.
  • The issuing agency must meet Article 10; the choice determines how complete the database fields will be.
  • Catalogue adjustments — four batches, 123 entries since 2020 — can change a product’s secondary category and therefore its UDI scope.

2. Legal basis

Administrative regulation. Medical Device Supervision and Administration Regulation (State Council Order No. 739), promulgated 9 February 2021, effective 1 June 2021.

  • Articles 13: Class I filed, Class II and III registered, with lifecycle liability for holder and filer.
  • Article 20, paragraph 1: five duties, including traceability and recall; paragraph 2: the designated Chinese legal entity assists in these duties.

Departmental rules. Measures for the Administration of Medical Device Registration and Filing (SAMR Order No. 47), promulgated 26 August 2021, effective 1 October 2021.

  • Article 14, paragraph 2: the overseas applicant or filer designates a Chinese legal entity as agent, to assist in the Article 20 duties. Article 15: quality system.
  • Article 123: registration and filing of devices from Hong Kong, Macao and Taiwan follows the imported-device route.

Production rules. Measures for the Supervision and Administration of Medical Device Production (SAMR Order No. 53), promulgated 10 March 2022, effective 1 May 2022. Article 4: production is classified, Class II and III requiring a licence, Class I a filing. Article 36: UDI requirements (wording outside the verification scope here; the text in force governs).

Normative documents. Rules for the Unique Identification System for Medical Devices (2019): Article 7 on composition, Article 10 on issuing agencies, Article 15 on upload before marketing. Announcement No. 106 of 2020 (first batch); No. 114 of 2021, 17 September 2021 (second); No. 22 of 2023 (third).

Practical estimate. Scope for all three batches is determined at the secondary product category under a sub-catalogue, so the category must be located before the batch lists are checked. On issuing-agency lists, deadlines for database field changes and catch-up entry of historical data, no verifiable official rule is held here; the requirement in force must be confirmed with the UDI database administration at NMPA.

3. Expert insights

DI and PI divide the maintenance workload. Article 7 splits the identifier into a static part, pointing to model, specification and holder, and a dynamic part pointing to batch, serial, production and expiry dates.

PartNatureChange frequencyUpload pointMain risk
UDI-DIStaticWith certificate particulars and specificationsOnce, before marketing (Article 15)DI data not updated after a certificate change
UDI-PIDynamicWith every production batchWith productionCoding rules not linked to the production system

The workload is lopsided: DI turns on synchronising change, PI on system integration. As a one-off coding project, UDI fails on the PI side: the rule sits in the quality manual while the execution system generates no PI accordingly, so label, database and product disagree.

The three batches follow risk from high to low and categories from narrow to broad.

BatchInstrumentScopeLogic
FirstAnnouncement No. 106 of 2020Part of Class IIIPilot varieties chosen from the highest risk class
SecondAnnouncement No. 114 of 2021Remaining Class III, including IVDsCompleting the Class III field
ThirdAnnouncement No. 22 of 202315 categories, 103 Class II varietiesExtending down to Class II by named list

Bringing IVDs into the second batch shows the line is risk and management necessity, not device versus reagent. The third batch names 103 varieties rather than covering the class: Class II enters variety by variety, so scope must be checked product by product.

MethodReliabilityTypical error
Judging by primary catalogue categoryUnreliableThe heading is listed but the secondary category is not; unnecessary coding cost
Checking each secondary categoryReliableSlower, but the conclusion can be verified

Class II status also shifts with catalogue adjustments: four batches, 123 entries since 2020 (2025 Annual Medical Device Registration Work Report, NMPA, 24 February 2026). After an adjustment the secondary category may change, so the earlier conclusion is rechecked.

The exemption boundary belongs here too: Class II devices selected in centralised procurement but outside the list are not yet mandatory. The test is the batch list — Class III in the first or second batch, Class II among the 103 varieties, Class II outside it not mandatory even if selected, Class I not covered. Voluntary implementation after selection is the enterprise’s decision.

Upload precedes sale, and therefore precedes shipment. Article 15 sets the point as before the product is placed on the market, which means upload completes before the first shipment, not alongside it and certainly not afterwards.

TimingActionBottleneck
After certificate or filingFix UDI-DI and complete codingChoice of issuing agency (Article 10)
Before first productionLabel design and coding rules settledLabel template review
Before marketingUpload identifier and data (Article 15)Confirmation of completion
During productionGenerate UDI-PI by batchExecution system capability

The third row is the one most often compressed: once shipment is planned, upload becomes paperwork rather than a precondition, and the first consignment leaves before the database holds a record. A hard upload-confirmation release gate is the only reliable safeguard.

The issuing and upload chain for an overseas holder. Article 20, paragraph 2 and Article 14, paragraph 2 put the landing on the China agent, and in UDI the agent’s tasks weigh more than in ordinary filing.

StepOverseas holderChina agent
Issuing agency (Article 10)Decides and contractsSupplies field requirements and Chinese data
Upload (Article 15)Bears primary responsibilityPerforms the upload
MaintenanceSupplies change informationUpdates domestic records

The difficulty sits in the second row: field requirements follow domestic rules, and data prepared to home-country habits often cannot be entered directly. Sending the field requirements to headquarters at project start beats correcting at upload (practical estimate).

4. Practical recommendations

  • Determine scope before starting, checking each secondary category against all three batch lists and recording the conclusion and its date.
  • Split DI and PI into two sub-projects: certificate change triggering DI synchronisation, and validation of execution-system coding before label design.
  • Select the issuing agency against Article 10, with domestic field completeness as one criterion.
  • Make upload a hard release condition with a completion checkpoint in the shipment process.
  • Set a recheck trigger for catalogue adjustments; with four batches and 123 entries since 2020, review at least annually (practical estimate).
  • Send field requirements early, translating the domestic field list before the agency is contracted.
  • Verify label, database and product agree, sampling after production switching or label revision.

5. Verification checklist

  • □ Scope determined variety by variety at secondary category level
  • □ Determination cites all three announcements (2020 No. 106, 2021 No. 114, 2023 No. 22)
  • □ Class II products checked against the 15 categories, 103 variety list
  • □ Issuing agency meets Article 10
  • □ Coding rules for UDI-DI and UDI-PI written into system documents (Article 7)
  • □ Identifier and data uploaded before marketing (Article 15)
  • □ Release process carries an upload completion checkpoint
  • □ Certificate changes trigger a DI data recheck
  • □ Execution system can generate UDI-PI by batch, and has been validated
  • □ Label, database and product sampled and confirmed consistent
  • □ Selected-but-unlisted Class II products handled as not yet mandatory, with a note on file
  • □ Secondary category and scope conclusion rechecked after catalogue adjustments

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Chinese market. Our team manages adverse event monitoring and recall filings with NMPA, maintains UDI compliance, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail