SaMD and Software Medical Device Regulation in Australia

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

The classification rules for software-based medical devices (SaMD) in Australia are items 4.5–4.8 of Schedule 2, added by F2019L01660 and effective from 25 February 2021 (and earlier, from 25 August 2020, for new applications). The widely circulated claim of a SaMD reclassification on 1 November 2024 is in fact a transitional deadline: legacy products already on the market or under assessment had to be re-included in the ARTG under the new classification by that date, and products that missed it must cease supply and be cancelled. Confusing a commencement date with a transitional cut-off is the single most consequential date error in this domain, because it inverts the compliance timeline: the obligations began years earlier than the headline date suggests, and the headline date marks the end of the grace period, not the start of the duty.

2. Legal basis

  1. Therapeutic Goods (Medical Devices) Regulations 2002 (F2002B00237, compilation F2026C00883): the software classification rules are Schedule 2 Part 4, items 4.5–4.8; regulation 1.3 provides that software that is a medical device is an active medical device.
  2. Addition and commencement: the Therapeutic Goods Legislation Amendment (2019 Measures No. 1) Regulations 2019 (F2019L01660) added cl 4.5–4.8, with its Schedules 2 and 3 effective from 25 February 2021; the transitional provisions of the same instrument apply cl 4.5–4.8 to applications made from 25 August 2020, and to “transitional class devices” from 1 November 2024.
  3. Transitional deferral mechanism: the Therapeutic Goods Legislation Amendment (2024 Measures No. 2) Regulations 2024 (F2024L00686, effective 1 July 2024) added regulation 11.45(4A)–(4C), applying to devices for which an application for a conformity assessment certificate was made between 25 February 2021 and 1 November 2024 and which were not yet included in the Register before 1 November 2024; for them the new classification rules apply from the day after the certificate procedure concludes.
  4. Software-related Essential Principles: 13B (version numbers and build numbers, effective 25 February 2021) and 12.1 (including cybersecurity requirements, which from 25 February 2021 operate as a clarification of the existing requirement rather than a new requirement).
  5. Exemptions and exclusions: Schedule 4, item 2.15 exempts qualifying clinical decision support system (CDSS) software; there is also an exclusion category, comprising 15 classes, listed in Schedule 1 of the Therapeutic Goods (Excluded Goods) Determination 2018.
  6. TGA software exclusion guidance series: an overview page and individual guidances published on 16 March 2026, covering imaging storage and transmission, population analysis, healthcare facility management, clinical workflow, calculators, LIMS/LIS, general health and wellness, communication, health self-management, digital mental health tools, behaviour change and coaching, electronic health records, health alert systems, patient questionnaires, middleware, and other categories.

3. Expert insights

Insight 1: Date confusion is the most expensive error. Writing 1 November 2024 as the date the new rules took effect leads a company to miss the true commencement point (25 February 2021) and thereby misjudge whether it has been in compliance all along. The transitional deadline and the commencement date are two different things and must be stated separately in every document. The cost of the confusion compounds over time: a team that believes the rules start in late 2024 will plan reclassification work late, discover the certificate queue is long, and arrive at the deadline without a submission rather than with one.

Insight 2: Multi-function software must be determined function by function. The exclusion list applies only where every function of the software meets the exclusion conditions; if even one function falls within the device category, the whole product is managed as a device. When we assess AI products we therefore work through a function list item by item, rather than reaching a conclusion about the product as a whole. This granularity protects both directions of error: it prevents an exclusion claim that one medical function would defeat, and it prevents an unnecessary device registration for software whose only medical function is truly peripheral.

Insight 3: Exemption and exclusion have different consequences but are easily conflated. CDSS software travels through the item 2.15 exemption, remains a medical device, and retains notification, adverse event reporting, and advertising obligations; software on the exclusion list is not subject to TGA device regulation at all. The evidentiary burdens and the residual duties of the two routes differ. A product owner must therefore be able to name not only which route the product travels but why, because the supporting analysis is the first thing an auditor or inspector will ask to see.

4. Practical recommendations

  1. Maintain a three-column “function — purpose — determination” list for every software product, checking each entry against the section 41BD definition and the exclusion lists, and record the date of the determination and the version of the sources relied upon.
  2. For marketed software, verify whether reclassification under cl 4.5–4.8 has been completed; where a deferral under regulation 11.45(4A)–(4C) was relied upon, track the conclusion date of the certificate procedure and re-start the clock from the following day.
  3. Prepare cybersecurity evidence as a stand-alone volume mapped to Essential Principles item 12.1, and bring version and build number management under a controlled process mapped to item 13B, so that both files are retrievable on demand during assessment.
  4. For AI and adaptive algorithm products, prepare a change management plan in line with the TGA software guidance before release, so that version iteration does not unexpectedly trigger a fresh application (see article 30 for details).

5. Verification checklist

  • □ Software classification has been determined under cl 4.5–4.8, distinguishing the commencement date from the transitional deadline
  • □ For multi-function software, the exclusion lists have been checked function by function
  • □ Whether the item 2.15 exemption or the exclusion list applies has been confirmed
  • □ Cybersecurity evidence has been compiled into a volume mapped to Essential Principles item 12.1
  • □ Version and build number management is under a controlled process (item 13B)

Regulatory information verified as of 29 October 2026.

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Australian market. Our team manages TGA inclusion and ARTG listing applications, coordinates conformity assessment evidence, and supports post-market vigilance obligations. Business enquiries: info@grzan.cn

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail