Class I Sterile, Measuring and Reusable Devices in Australia

Updated: 2026-11-11 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

“Class I” in Australia is not a single unified pathway: Class I devices that are neither sterile nor have a measuring function require only a declaration of conformity made by the manufacturer under Schedule 3 clause 6.6, whereas Class I devices that are sterile or have a measuring function must obtain conformity assessment documents. This distinction directly determines project timelines and cost, because the declaration route requires no Secretary evaluation while the conformity assessment route requires third-party or TGA assessment before the application can be lodged.

  • Schedule 1 Part 1 Division 1 item 1 covers Class I non-sterile, non-measuring devices: a declaration of conformity made by the manufacturer under Schedule 3 clause 6.6 of the Therapeutic Goods (Medical Devices) Regulations 2002.
  • Division 2 items 1–7 cover Class I sterile or measuring devices, which must obtain conformity assessment documents.
  • Sterile devices proceed under Schedule 3 Part 1 (full quality system, excluding clause 1.6) or Part 4 (production quality assurance).
  • Measuring devices that are not sterile proceed under Part 1 (excluding clause 1.6), Part 3 (verification procedures), Part 4 or Part 5 (product quality assurance).
  • Where a system or procedure pack (SOPP) is supplied sterile, quality management system and product assessment documents must additionally be obtained under the Schedule 3 Part 2 table, pursuant to section 7(2) of the Determination.
  • Reusable surgical instruments are Class I under clause 3.2(4), yet clause 13.4(3) item 20 still requires the reprocessing steps and the number of safe reuses.

2. Legal basis

Therapeutic Goods (Medical Devices—Information that Must Accompany Application for Inclusion) Determination 2018 (F2018L01410, compilation version 13, effective 1 December 2025): Schedule 1 Part 1 Division 1 item 1 (Class I non-sterile, non-measuring) — a declaration of conformity made by the manufacturer under Schedule 3 clause 6.6 of the Therapeutic Goods (Medical Devices) Regulations 2002; Division 2 items 1–7 (Class I sterile or measuring) — conformity assessment documents must be obtained. The Division 1 and Division 2 split is therefore the first thing to resolve for any Class I product, before any other evidence planning begins.

Procedural limits of the sterile pathway: Division 2 item 1 — sterile devices proceed under Schedule 3 Part 1 (full quality system, excluding clause 1.6) or Part 4 (production quality assurance); measuring devices that are not sterile proceed under Part 1 (excluding clause 1.6), Part 3 (verification procedures), Part 4 or Part 5 (product quality assurance). The Division thus fixes in advance which Parts of Schedule 3 can be relied on for each attribute, leaving no room to mix and match procedures outside those entries.

Where a system or procedure pack (SOPP) is supplied sterile, quality management system and product assessment documents must additionally be obtained under the Schedule 3 Part 2 table, pursuant to section 7(2) of the Determination. The obligation is additional precisely because the sterile presentation of the pack raises questions that the pack assembly alone does not answer.

Classification of reusable surgical instruments: Schedule 2 clause 3.2(4) of the Therapeutic Goods (Medical Devices) Regulations 2002 classifies them as Class I; the definitions appear in the Dictionary. The classification settles only the registration burden, and it leaves the instruction obligations untouched.

Reuse instruction requirement: Essential Principle Schedule 1 clause 13.4(3) table item 20 — the process allowing reuse (cleaning, disinfection, packaging and, where necessary, re-sterilisation) as well as the number of safe reuses must be stated. The clause applies by virtue of the Essential Principles themselves, independently of how the instrument was classified for registration purposes.

3. Expert insights

Insight 1: the cost differential within Class I can reach several multiples. Non-sterile, non-measuring devices proceed under the Part 6 declaration of conformity without Secretary evaluation, while sterile or measuring devices proceed under one of Parts 1, 3, 4 or 5, which require third-party or TGA assessment. Quotations and schedules must first confirm these two attributes, because a single added attribute changes not only the price but also the parties involved in the assessment and the time needed before the application can be lodged.

Insight 2: the measuring function is an independent upgrade trigger. A Class I device with a measuring function must obtain conformity assessment documents even if it is not sterile. Many companies quote on the basis of an ordinary Class I device and discover at the submission stage that system evidence is required, which forces a rushed search for assessment bodies at exactly the point where the schedule has no room left for them.

Insight 3: reusable surgical instruments are Class I, yet reuse instructions remain a hard requirement. Clause 3.2(4) classifies them as Class I, but clause 13.4(3) requires the instructions for use to state the reprocessing steps and the number of reuses. A low classification does not mean low instruction requirements, and the two questions must be answered separately rather than assuming one follows from the other.

4. Practical recommendations

  • Confirm three attributes before quoting: whether the device is sterile, whether it has a measuring function and whether it is reusable, and determine the conformity assessment procedure and the evidence list accordingly.
  • For SOPP projects, check section 7(2) and the two-column requirements of Schedule 3 Part 2 separately.
  • Complete reuse-number verification for reusable devices early, and map the verification report to the instruction clauses one by one.
  • For sterile devices, check the sterilisation and packaging standards at the same time (see article 29 for details).

5. Verification checklist

  • □ Whether the device is sterile, has a measuring function and is reusable has been confirmed
  • □ The applicable Part of Schedule 3 has been determined from the attributes
  • □ SOPP projects have been checked against section 7(2) and Schedule 3 Part 2
  • □ Reuse-number verification has been completed for reusable devices
  • □ The instructions for use state the reprocessing steps and the number of reuses

How GRZAN can help. GRZAN provides medical device and IVD registration, quality system, authorised representative and local testing support for the Australian market. Our team prepares TGA-compliant CTOs and evidence dossiers, and supports post-market vigilance obligations. Business enquiries: info@grzanglobal.com

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail