GMDN Coding and Device Grouping Rules in Tanzania
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Tanzania has hard rules on how much can go into one application. Part I §3.3.1 of the 2024 Compendium of Guidelines: an application may only contain one of the following:
- a) A single medical device
- b) One medical device family
- c) One medical device system
- d) One medical device group
In principle, a separate application is required for each medical device, but grouping is permitted, provided the manufacturer has demonstrated that the devices submitted together meet the grouping criteria. The criteria are set out in Part IX, the Guidelines on Grouping of Devices and IVDs, which provides five decision flowcharts: FAMILY, SYSTEM, IVD TEST KIT, IVD CLUSTER and GROUP.
Getting grouping right at the start determines both the fee level and the flexibility of future portfolio changes. The rules below show why grouping decisions are best treated as strategic choices, documented against the official flowcharts, rather than administrative conveniences decided late in the process.
2. Legal basis
| Instrument | Reference / number | Publication date | Provisions |
|---|---|---|---|
| Compendium of Guidelines for Marketing Authorization of Medical Devices, Diagnostics and Laboratory Equipment | No Government Gazette number (TMDA guideline document) | June 2024 | Part I §3.3.1, §3.4.1; Part IX (pages 146–174, including the five decision flowcharts) |
| The Tanzania Food, Drugs and Cosmetics (Control of Medical Devices) Regulations, 2015 | GN. No. 315 | 31 July 2015 | Regulations 7(1), 7(3)(c) |
| Guidelines on Submission of Application for Change(s) to Approved Medical Devices and In-Vitro Diagnostics | TMDA/DMD/MDA/G/003 | June 2022 | §4(h), §4(i) |
3. Expert insights
Insight 1: Regulation 7(1) of GN 315 and the Compendium form a principle-plus-exception relationship.
Regulation 7(1): an application for registration of a device shall be made in respect of each single device, or one device group, or one device family, or one device system. Regulation 7(3)(c) further requires the application to state the identification of the device, including the identification of any device that is part of a system, test kit, device group, device family or group of device families. In other words, the four forms in the statute match the Compendium exactly: single/group/family/system. There is no fifth form, and no general catch-all that would allow loose combinations of unrelated products in a single file.
Insight 2: GMDN is one of the five pre-submission essentials and is bound to the Category field.
Part I §3.4.1 requires the following before submission: a) device class; b) intended use; c) GMDN code and term; d) conformity assessment certificate; e) declaration of conformity. The Category field in Part II requires the applicant to state the GMDN category of the device. GMDN is therefore not merely a data entry item: it is also an input to the grouping decision — sharing the same GMDN category is usually one of the preconditions for forming a family.
Insight 3: A wrong grouping decision becomes a non-changeable matter at the change stage.
TMDA/DMD/MDA/G/003 §4(h) lists the addition of models that do not conform to the family/operation package/group definitions among matters that cannot be processed as change applications; §4(i) lists likewise the addition of devices under a brand name different from the registered device or IVD. In both cases, only a new application is possible. In other words, if grouping is made too broad at the outset, adding models later will be blocked and pushed back to a new application; if grouping is made too narrow, additional application fees must be paid. Both errors are avoidable with an early flowchart exercise.
Insight 4: IVDs have two flowcharts independent of non-IVDs — test kit and cluster.
Part IX sets up two separate decision flowcharts for IVDs, IVD TEST KIT and IVD CLUSTER, alongside the three non-IVD flowcharts of FAMILY, SYSTEM and GROUP. This parallels the treatment of IVD classification in Part VI as a separate track — Tanzania manages IVDs institutionally as an independent category, and the grouping logic reflects that separation.
Insight 5: Groups, families and kits filed under notification require a component-by-component description.
Part IV §3.3: device groups, families or kits may be notified under a single application, provided that (1) each component is fully described; (2) the intended use of each component is stated; and (3) for kits containing devices that require registration (such as sutures), the TMDA registration number of that device must be provided. This is the counterpart of the grouping rules on the notification pathway.
4. Practical recommendations
- Run the relevant Part IX flowchart before submission and archive the decision process as the grouping rationale.
- Confirm GMDN codes and terms model by model, and build a model-to-GMDN mapping table to submit with the application.
- Group conservatively against the official definitions, to avoid later model additions being blocked by §4(h) into a new application.
- Treat cross-family or cross-brand additions as new applications (§4(h), §4(i)); do not gamble on the change route.
- For IVDs, use the test kit or cluster flowcharts, and do not borrow the non-IVD family/system rules.
5. Verification checklist
- □ Confirmed that each application contains only one of the four forms: single device/family/system/group.
- □ Relevant Part IX decision flowchart completed and archived.
- □ GMDN code and term confirmed before submission (§3.4.1(c)).
- □ Model-to-GMDN mapping table established.
- □ Aware that §4(h) (new models not conforming to grouping definitions) and §4(i) (different-brand additions) require new applications only.
- □ IVDs assessed with the test kit/cluster flowcharts, not with the non-IVD rules.
Regulatory information verified as of 29 October 2026. All provisions cited in this document were checked against the official published texts of TMDA (the June 2024 Compendium of Guidelines, Part I §3.3.1 and §3.4.1, Part II, Part IX; GN. No. 315, regulation 7; TMDA/DMD/MDA/G/003 §4(h) and §4(i)), and will be re-checked when the regulations are updated.
- Tanzania local resources: licensed Local Responsible Person (LRP) and distributor resources can be coordinated to cover online submission, sample delivery, quality audit cooperation and port clearance.
- Service capabilities: classification determination, registration and notification pathway selection, technical documentation (CSDT/STED) preparation and localization, GMDN and grouping strategy, clinical evidence organization, software and AI device submissions, change management, vigilance and PMS reporting, annual retention fees and renewal management.
- Contact: hito.lin@grzan.cn; website www.grzan.cn.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us