Four Risk Classes: Classification from Class A to Class D

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

South Africa applies a four-tier risk classification aligned with IMDRF principles: Class A (low risk), Class B (low to moderate risk), Class C (moderate to high risk) and Class D (high risk), with risk directed at patients, users or public health. The classification rules are set out in SAHPGL-MD-04, the Guideline on the Classification of Medical Devices and IVDs, of which the current version is v5, dated 27 February 2025. Non-IVD devices are governed by 16 classification rules, and IVDs by 7 rules. Where multiple rules apply simultaneously or the outcome is unclear, the higher risk class applies; classification disputes are decided by SAHPRA. Classification is not an administrative formality in South Africa: it drives the establishment licence scope, the registration expectations under any future call-up notice, and the depth of technical documentation. Getting it wrong at the start propagates through every downstream submission, so companies should treat classification as a documented engineering decision with traceable reasoning rather than a one-line form entry.

2. Legal basis

InstrumentReference numberPublication dateProvisions
Regulations Relating to Medical Devices and In Vitro Diagnostic Medical Devices (IVDs)Government Gazette No. 40480, Notice No. 1515Published on 9 December 2016Regulation 11 (classification): subregulation (1) (definitions of Class A, B, C and D); subregulation (2) (registration under a call-up notice); subregulation (3) (determination according to the classification rules); subregulation (4) (higher class where unclear); subregulation (5) (individual assessment by design and intended purpose)
Guideline on the Classification of Medical Devices and IVDsSAHPGL-MD-04 v527 February 2025Full text; non-IVD classification rules (Rules 1–16); IVD classification rules (Rules 1–7); glossary and flow charts
Classification matrix formGLF-MD-14ACurrent version, verified as of 29 October 2026Full form (used to document classification decisions)
SAHPRA Communication to Industry on the Expression of Interest for the voluntary medical device registration feasibility studyCommunication to Industry17 May 2024Cites SAHPGL-MD-04 as the classification basis

3. Expert insights

Insight one: the decisive factor in classification is the intended purpose, not the product name. The guideline makes clear that classification is based on the intended purpose assigned by the manufacturer or distributor, the level of risk to patients and users (the probability and severity of harm), the degree of invasiveness and the duration of use and exposure. The same device applied to different body sites may fall into different classes; the intended purpose can be drawn from the instructions for use, labelling, the original manufacturer’s advertising materials and technical files. This explains why identically named products can receive different classification outcomes in different companies’ hands, and why marketing language needs to be written with regulatory consequences in mind.

Insight two: the 7 IVD rules and the 16 non-IVD rules are two independent systems. The IVD rules divide by analyte and use: Rule 1 covers high-risk pathogens (Class D); Rule 2 covers blood grouping (Class C and D); Rule 3 covers Class C; Rule 4 covers self-testing (Class C); Rule 5 covers control materials (Class B); Rule 6 covers reagents and instruments (Class A); and Rule 7 is the residual rule (Class B). Companies cannot extrapolate IVD classifications from non-IVD reasoning, and this is the single most common error GRZAN sees when reviewing submission pipelines for first-time South African entrants with combined diagnostic portfolios.

Insight three: Rules 13, 14, 15 and 16 are South African additional rules that may not align internationally. The guideline specifically flags that manufacturers and distributors should pay close attention to Rules 13 to 16 (additional rules), because their application may not be consistent across jurisdictions. Practical estimate: for products touched by these rules, it is advisable to confirm classification with SAHPRA through a procedural enquiry or in writing before submission, so that the recorded class is not challenged at review when timelines matter most.

Insight four: combinations, systems and procedure packs take the highest class. Where a device is intended for use with other devices, the classification rules must be applied to each device individually; for device groups, systems or procedure packs, the overall classification is the highest classification of any single device in the combination. The inclusion of a registered medicine in a procedure pack does not affect the classification. This rule directly determines how finely a product list must be built: a pack containing one Class C device is managed as Class C as a whole, and under-specifying the pack contents invites rework at review.

Insight five: software is treated as an active device, but accessories are classified separately. The guideline states that software meeting the device definition also qualifies as an active medical device (because its operation depends on an energy source); software intended to drive, control or influence the functioning of a device generally falls in the same class as that device; software that is an accessory to a device is classified separately; and software not necessary for the device to operate is treated as an accessory. AI and ML software is classified within this same framework (see also MD08-2025/2026), so algorithmic products should not assume a special pathway.

4. Practical recommendations

  1. Fix the intended purpose before applying the rules: extract it from the instructions for use and labelling, not from the product name.
  2. Use the 7 rules for IVDs and the 16 rules for non-IVDs: the two systems must never be mixed.
  3. Where in doubt, take the higher class: this is an explicit regulatory rule and also the safest choice.
  4. Manage combinations at the highest class: annotate this when building the product list.
  5. Confirm Rule 13 to 16 products early: consult SAHPRA in writing where necessary.
  6. Record the outcome in GLF-MD-14A: the classification matrix is the common baseline for later submissions and inspections.

5. Verification checklist

  • □ Verify the four-tier class definitions against Regulation 11(1)
  • □ Verify that the current text of SAHPGL-MD-04 v5 (2025-02-27) has been obtained
  • □ Verify whether the product is an IVD (7 rules) or a non-IVD (16 rules)
  • □ Verify that the intended purpose has been extracted and locked from the instructions for use and labelling
  • □ Verify that each applicable rule has been assessed and the higher class taken where unclear
  • □ Verify that the applicability of Rules 13 to 16 has been assessed separately
  • □ Verify that combinations, systems and procedure packs are classified at the highest class
  • □ Verify that the classification outcome has been recorded in the GLF-MD-14A matrix

Regulatory information verified as of 29 October 2026. The instruments, guidelines and effective dates described in this article are limited to the content published by the official sources listed under the legal basis above as at that date; where official documents are updated, the latest published version prevails.

  • South Africa local support: GRZAN assists manufacturers in the Chinese mainland and abroad with SAHPRA Medical Device Establishment Licence (MDEL) applications and product registration preparation, including Authorised Representative (AR) arrangements, ISO 13485 certification pathway planning and CAB selection, GMDN coding and classification decisions, technical documentation, and the setup of vigilance and adverse event reporting mechanisms.
  • Service capabilities: product attribute determination and risk classification (Class A, B, C, D), establishment licence applications and renewals, reliance pathway arguments, technical documentation and essential principles conformity arguments, labelling and instructions for use localisation in English, and post-market vigilance and annual compliance submissions.
  • Contact: info@grzanglobal.com; hito.lin@grzan.cn; website www.grzan.cn.

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail