RUO and IUO Medical Devices in South Africa: SAHPRA MD01-2026/27 Boundaries

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

On 25 June 2026 SAHPRA published the communication Regulatory Requirements for Research Use Only (RUO) and Investigational Use Only (IUO) Medical Devices (including IVDs) (MD01-2026/27). The core boundary is this: RUO/IUO products are not used in clinical diagnosis or treatment, and therefore circulate on a different track from marketed devices; the moment a manufacturer’s claims, labelling or manner of use point towards diagnostic or therapeutic purposes, the product ceases to be RUO/IUO and must be handled under the full device requirements.

2. Legal basis

InstrumentReference / NumberPublication dateProvision
SAHPRA communication on regulatory requirements for research use only (RUO) and investigational use only (IUO) medical devices (including IVDs)MD01-2026/2725 June 2026Full text (definition of RUO/IUO, labelling and claims restrictions)
Medicines and Related Substances ActAct No. 101 of 1965Enacted and published in 1965, subsequently amended; verified as at 29 October 2026Definition provision for medical device (intended use as the core determinant); Section 14; Section 22C(1)(b)
Regulations relating to Medical Devices and In Vitro Diagnostic Medical Devices (IVDs)Government Gazette No. 40480, Notice No. 1515Published on 9 December 2016Regulation 22 (labelling); Regulation 23/24 (instructions for use); Regulation 16 (clinical trials)
SAHPRA guidance on the classification of medical devices and IVDsSAHPGL-MD-04 v527 February 2025Section 3.2 (intended use determines classification)

3. Expert insights

Insight 1: The anchor for determining RUO/IUO status is intended use, which shares its source with the device definition. The device definition in Act No. 101 of 1965 is built on the manufacturer’s intended use, and SAHPGL-MD-04 v5 likewise makes classification depend on intended use. Consequently, a reagent labelled RUO whose promotional materials describe it as suitable for testing a particular disease has already had its intended use altered — it no longer qualifies as RUO, whatever its certificate or catalogue page says. Companies should therefore treat any statement of clinical purpose, in whatever medium it appears, as a regulatory act with consequences.

Insight 2: MD01-2026/27 was issued in 2026, showing this is a line newly drawn by SAHPRA. The communication is the first medical device communication SAHPRA published in 2026 (numbered MD01-2026/27). Its significance lies in bringing RUO/IUO products, previously in a grey zone, onto an explicit track — for companies this is both a constraint and a protection, since clear rules reduce the uncertainty that previously surrounded research products. For products already circulating as RUO, the communication now provides the reference point against which existing labelling and claims should be reviewed.

Insight 3: RUO/IUO and clinical trials are two different pathways. Products used in pre-market clinical studies follow the clinical trial pathway under Regulation 16; only products used in basic scientific research qualify as RUO. Confusing the two pathways leads either to research products being used without the required trial authorisation, or to unnecessary regulatory burdens being shouldered for genuinely non-clinical research, both of which are costly outcomes. A simple internal gate — asking whether the product will generate data used to support market access — separates the two populations cleanly.

Insight 4: The RUO risk is highest in the IVD field. The boundary between research and diagnosis is blurriest for testing reagents, and test results inherently carry diagnostic implications. Practical estimate: IVD companies should review labelling, instructions for use, websites and sales scripts for RUO products in a single unified pass, because a diagnostic claim appearing on any one channel is enough to invalidate the RUO characterisation across the board. Distributor training matters as much as head-office review, because sales conversations happen far from the person who drafted the label.

Insight 5: Moving from RUO to a marketed device is a complete compliance transition. Once a product leaves RUO status, it must be handled under the full device requirements: establishment licence, classification, labelling and instructions for use, and essential principles conformity, with product registration under future call-up notices still to come. Companies should manage the transition from RUO to commercialisation as an independent project milestone, not as a mere labelling edit, because every stage of the compliance chain must be re-run. Planning the transition early also protects continuity of supply for existing research customers while the commercial file is built.

4. Practical recommendations

  1. Review claims for RUO/IUO products channel by channel: labelling, instructions for use, website, sales scripts.
  2. Clearly separate RUO products from clinical trial products: each follows its own pathway.
  3. IVD companies should focus on diagnostic claims: this is where characterisation most often fails.
  4. Set the transition from RUO to commercialisation as an independent project milestone: re-run the full compliance process.
  5. Keep internal determination records of product use: available to explain at inspections.
  6. Track subsequent revisions of MD01-2026/27.

5. Verification checklist

  • □ Current text of MD01-2026/27 (2026-06-25) obtained
  • □ Determined item by item, on intended use, whether products qualify as RUO/IUO
  • □ Labelling, instructions for use, website and sales scripts reviewed with no diagnostic claims
  • □ RUO products distinguished from clinical trial products governed by Regulation 16
  • □ Diagnostic claims reviewed with priority for IVD products (practical estimate)
  • □ Transition from RUO to commercialisation set as an independent project milestone
  • □ Internal determination records of product use retained
  • □ Subsequent revisions of MD01-2026/27 added to the monitoring list

Regulatory information verified as at 29 October 2026. The regulations, guidance and effective dates described in this article are limited to the content published by the official sources listed under Legal basis as at that date; where official documents have been updated, the latest published version prevails.

  • South Africa local resource: GRZAN can assist manufacturers from the Chinese mainland and overseas in engaging the South African Health Products Regulatory Authority (SAHPRA) on medical device establishment licence (MDEL) applications and product registration preparation, including authorised representative (AR) scheme design, ISO 13485 certification route planning and CAB selection, GMDN coding and classification determination, technical documentation preparation, and the establishment of vigilance and adverse event reporting mechanisms.
  • Service capabilities: product attribute determination and risk classification (Class A/B/C/D), establishment licence application and renewal, reliance pathway justification, technical documentation and essential principles conformity argumentation, labelling and instructions-for-use localisation in English, and post-market vigilance and annual compliance information submission support.
  • Contact: hito.lin@grzan.cn; website www.grzan.cn.

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail