TSE/BSE Attestation and Biological Materials Safety in Ethiopian Dossiers
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
The Ethiopian non-IVD dossier sets two documentation requirements for biological source materials: Section 1.7 of Chapter 1 requires submission of a risk-free TSE/BSE attestation, and Section 3.4.8 of Chapter 3 requires a safety evaluation of biological source materials (human or animal). The two sit respectively in the administrative dossier and the evidence dossier: the former is a declarative document and the latter is a technical evaluation report, and neither can substitute for the other.
The administrative placement of the attestation has real consequences. Chapter 1 of EFDA/GDL/097 Version 002 consists entirely of eligibility documents (cover letter, agency agreement, fee payment proof, QMS, GMP, Certificate of Free Sale, TSE/BSE attestation, DoC), so a missing attestation returns the dossier at screening rather than generating a correction request during assessment. Meanwhile the technical evaluation under Section 3.4.8 addresses a broader question than TSE/BSE alone, covering viral safety, immunogenicity and sourcing control for both human and animal materials.
Manufacturers should therefore start with a materials inventory: does the product contain animal-derived or human-derived components, and if so which. Where the answer is no, a clear negative attestation should be filed rather than leaving the item silent. Where the answer is yes, three deliverables come into play: the attestation in Chapter 1, the source material safety report under Section 3.4.8, and, because of the animal contact, a biocompatibility evaluation under Section 3.4.6.
2. Legal basis
| Regulation | Reference number | Publication date | Relevant provisions |
|---|---|---|---|
| EFDA Guideline for Requirements of Registration of Non-IVD Medical Devices | EFDA/GDL/097 Version 002 | Effective June 30, 2026 | Section 1.7 (risk-free TSE/BSE attestation) |
| EFDA Guideline for Requirements of Registration of Non-IVD Medical Devices | EFDA/GDL/097 Version 002 | Effective June 30, 2026 | Section 3.4.8 (safety of biological source materials: human/animal) |
| EFDA Guideline for Requirements of Registration of Non-IVD Medical Devices | EFDA/GDL/097 Version 002 | Effective June 30, 2026 | Section 3.4.6 (biocompatibility and toxicological evaluation) |
| EFDA recognized standards list | As published by EFDA | Online since July 17, 2026; verified as of October 29, 2026 | Standards related to biocompatibility and biological source materials |
3. Expert insights
Insight 1: the TSE/BSE attestation is placed in the administrative dossier, which shows it is an eligibility document. Chapter 1 consists entirely of eligibility administrative documents (cover letter, agency agreement, fee payment proof, QMS, GMP, Certificate of Free Sale, TSE/BSE attestation, DoC). Its inclusion in Chapter 1 means that its absence leads to a return at the screening stage, not a request for correction during assessment. This distinction changes how the document should be managed: it belongs in the pre-submission gate alongside signatures and certificates, not in the technical workstream that can still be completed while the dossier is under review.
Insight 2: the scope of the TSE/BSE attestation is tied to animal-derived materials. The risk of transmissible spongiform encephalopathies (including bovine spongiform encephalopathy, BSE) is directly related to animal source materials. Devices containing animal-derived components (such as gelatin, collagen and animal-derived excipients) must submit the attestation; for products that contain no animal source materials, an explicit negative attestation should normally be made in the dossier rather than simply omitting the item. A documented negative declaration costs almost nothing to produce and eliminates the only interpretation under which a silent dossier item becomes a defect.
Insight 3: Section 3.4.8 covers both human-derived and animal-derived materials. The item is worded as safety of biological source materials (human/animal), and its scope is broader than TSE/BSE: beyond transmissible spongiform encephalopathies, it also involves viral safety, immunogenicity and sourcing control. Manufacturers of products with human-derived materials, such as collagen of human origin or human cell-derived components, should address donor screening, viral inactivation or removal validation and traceability of the supply chain, using the recognized standards list as the methodological reference for each of these elements.
Insight 4: the evaluation is independent of biocompatibility (Section 3.4.6). Section 3.4.6 evaluates the biological response of the body to contact with the device (cytotoxicity, sensitization, irritation and the like); Section 3.4.8 evaluates the safety of the material at its source. Devices containing animal-derived materials usually need to submit both. The two reports answer different questions and are prepared from different data: a device can pass biocompatibility testing on its final finished form yet still require source material documentation, and vice versa for the sourcing dossier of a material whose contact profile is benign.
Insight 5: the issuing party of the attestation must be accountable. Consistent with the DoC, the TSE/BSE attestation, as an administrative document, must be issued and signed by a party capable of bearing responsibility. An attestation issued by a raw material supplier must be accompanied by supply chain traceability information. In practice this means the manufacturer should collect supplier attestations plus the chain-of-custody records showing where the material was sourced, and then issue its own covering attestation signed by an authorized person, so the authority receives one accountable document supported by verifiable upstream records.
4. Practical recommendations
- Determine first whether the product contains animal or human source materials: and decide from that whether the two documents apply.
- Where the relevant materials are absent, issue an explicit negative attestation: do not simply omit the item.
- Prepare the TSE/BSE attestation and the Section 3.4.8 report separately: they cannot substitute for each other.
- Where animal-derived materials are present, prepare a biocompatibility evaluation in parallel.
- Supplier attestations must be accompanied by supply chain traceability information: and state the issuing party.
- Select the basis for viral safety and sourcing control from the recognized standards.
5. Verification checklist
- □ Whether the product contains human or animal source materials has been determined
- □ The risk-free TSE/BSE attestation (or a negative attestation) has been submitted in Chapter 1
- □ The safety evaluation of biological source materials has been submitted under Section 3.4.8
- □ Products with animal-derived materials have also submitted a biocompatibility evaluation
- □ Supplier attestations are accompanied by supply chain traceability information
- □ The methodological basis adopted has been stated according to the recognized standards
Regulatory information verified as of October 29, 2026. The TSE/BSE attestation and biological source materials evaluation items listed in this article are based on the current text of EFDA/GDL/097 Version 002 Sections 1.7 and 3.4.8; when the guideline is revised, they must be re-checked.
- Ethiopia local resource: GRZAN can assist manufacturers from the Chinese mainland and overseas with marketing authorization (MA), Certificate of Competency (CoC) and import licence applications to the Ethiopian Food and Drug Authority (EFDA), including local agent planning and agency agreement review, SRA reliance pathway justification, UMDN coding and classification determination, technical documentation preparation, and the setup of vigilance and post-market surveillance systems.
- Service capabilities: product attribute determination and risk classification (non-IVD Class I–IV / IVD Class A–D), marketing authorization application and re-registration, applicability determination for the low-risk notification procedure, Essential Principles conformity justification, labelling and IFU localization (English), and vigilance reporting and field safety corrective action support.
- Contact: hito.lin@grzan.cn; website www.grzan.cn.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us