Home > News Center > Brazil Medical Device Registration

Brazil Medical Device Registration

11 月 26, 2025 Registration Insights 0 views

Brazil Medical Device Registration (ANVISA – Agência Nacional de Vigilância Sanitária)

1. Regulatory Authority & Framework

Brazil regulates medical devices under:

  • RDC 751/2022 – General medical device regulation (replaces RDC 185/2001)
  • RDC 36/2015 – GMP requirements
  • RDC 40/2015 – Changes/variations
  • RDC 661/2022 – IVD rules
  • IN 95/2021 – Classification rules

Authority:
ANVISA – Brazilian Health Regulatory Agency

System platforms:

  • Datavisa – registration submission
  • SGSCo – GMP inspection
  • Solicita – documentation portal

Brazil uses Portuguese as the regulatory language.

2. Device Classification (Class I / II / III / IV)

Categorization aligns with EU/GHTF principles:

ClassRisk LevelExamples
Class ILowNon-sterile consumables
Class IILow–ModerateSuction pumps, catheters (non-implant)
Class IIIModerate–HighBlood glucose meters, HF surgical devices
Class IVHighImplants, robotic surgical systems

Classification based on:
intended purpose, invasiveness, duration, active status, IVD category.

3. Registration Routes

Brazil has two primary pathways:

3.1 Cadastro (Notification) – For Class I & Class II

Simplified pathway for low-risk devices:

  • No ANVISA technical review
  • Automatic approval upon dossier submission
  • No local clinical trials needed
  • Validity: Indefinite (as long as GMP is maintained)

Documentation:

  • Administrative documents
  • Technical dossier summary
  • Labeling
  • Proof of regulatory approval or safety data

Approval time: 1–3 months

3.2 Registro – For Class III & Class IV

Full ANVISA technical review:

Includes:

  • Full technical dossier
  • Performance & biocompatibility data
  • Clinical evaluation
  • Brazilian GMP (ANVISA GMP) mandatory
  • Essential principles checklist

Validity: 10 years
Approval time: 9–18+ months

This is the pathway needed for robots, HF devices, blood glucose meters, lenses, many active devices.

4. Mandatory Requirement: ANVISA GMP (Brazil GMP)

ANVISA is strict: GMP certification is mandatory for Class III & IV devices before Registro approval.

GMP Options:

  1. ANVISA On-site Inspection (most common; longest timeline)
  2. MDSAP certificate (accepted for some devices but not all)
  3. Brazil GMP via local manufacturing (if applicable)

Typical GMP timeline: 8–12 months (sometimes longer).

5. Local Authorized Representative (Brazilian Registration Holder – BRH)

Foreign manufacturers cannot register directly and must appoint a:

Brazil Registration Holder (BRH) / Local Company

Responsibilities:

  • Owns the registration (CRF)
  • Communicates with ANVISA
  • Manages vigilance/FSCA
  • Handles labeling and import permits

Important:
BRH owns the registration → changing BRH requires re-registration.

6. Dossier Requirements (Brazil Technical File)

Brazil does not use ASEAN CSDT; instead, it uses Brazil-specific structure based on RDC 751/2022.

Administrative Documentation

  • Application form (petition)
  • BRH authorization letter
  • Manufacturer QMS certification
  • ANVISA GMP (for Class III/IV)
  • FSC / Market Authorization

Technical Documentation

  • Device description
  • Essential Principles of Safety & Performance
  • Risk management (ISO 14971)
  • Verification & validation data
  • Biocompatibility / sterilization
  • Electrical safety & EMC (IEC 60601 series)
  • Software documentation (IEC 62304 + cybersecurity)
  • Clinical evidence / CER
  • Labeling in Portuguese (IFU may be bilingual)

Labeling must be fully compliant with Portuguese language requirements.

7. Fees & Timelines

7.1 Official Fees (Typical)

  • Class I/II Cadastro: BRL 500–800
  • Class III/IV Registro: BRL 1,000–3,000
  • ANVISA GMP inspection: BRL 50,000–70,000 (varies by region)

7.2 Timelines

StepTimeline
BRH preparation1–2 months
ANVISA GMP8–12+ months
Cadastro (I/II)1–3 months
Registro (III/IV)9–18+ months
Total for Class III/IV12–24 months

8. Labeling Requirements

All labeling must be in Portuguese, including:

  • Product name, model
  • Manufacturer & BRH details
  • Country of origin
  • ANVISA registration number (post-approval)
  • Storage/handling conditions
  • Expiry/manufacture date (if applicable)
  • IFU in Portuguese (can be electronic where allowed)
  • UDI system (gradually implemented)

9. Importation Requirements

Post-registration import steps:

  • Use of Licenciamento de Importação (LI) via SISCOMEX
  • Pre-shipment instructions
  • Inspection at customs
  • Compliance with ANVISA, INMETRO (for some electrical devices), and MAPA (for biologics) if applicable
  • Batch control for specific device types

10. Post-Market Surveillance

BRH must manage:

  • Vigilance & incident reporting
  • Corrective and preventive actions (CAPA/FSCA)
  • Annual performance monitoring
  • Labeling changes
  • Variations per RDC 40/2015 rules
  • Renewal (Registro valid for 10 years)

11. Consultation Notes for Your Device Portfolio

Device TypeExpected ClassBrazil Requirements
Soft contact lensesClass IIIRequires Registro + ANVISA GMP + clinical evaluation
Surgical robots / AR-MR systems (SyngAR™)Class IVLongest pathway; full V&V + cybersecurity + performance + GMP
High-frequency electrosurgical devicesClass III/IVElectrical safety / EMC / clinical evidence + GMP mandatory
Blood glucose meters & stripsClass IIIISO 15197 + performance testing; Registro + GMP
Single-use serum preparation deviceClass II or IIICadastro if Class II; Registro+GMP if Class III

Brazil is one of the strictest markets globally because of the mandatory GMP for higher-risk devices.

12. Advantages of Brazil Market

  • Largest medical device market in Latin America
  • Well-defined regulatory framework
  • Recognition of EU/US certifications simplifies technical review (not GMP)
  • Long validity (10-year Registro)
  • Strong private healthcare growth