Armenia Medical Device Definition 2026: Device, Drug and IVD Boundary Rules
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
The Armenian definition of a medical device is fully merged into the Union framework, and the decisive test is whether the intended principal function is achieved through pharmacological, immunological, genetic or metabolic means, which are the four modes of action that exclude device status. The GRZAN screening rule: a product whose principal intended action is not achieved by pharmacological, immunological, genetic or metabolic means is a medical device; even where a medicinal product assists the action in the course of use, the product is still handled as a medical device. Standalone software is written explicitly into the definition — standalone software is regarded as an active medical device. In vitro diagnostics (IVD) have a separate definition and are subject to separate documentation and testing requirements, which must not be borrowed from the non-IVD pathway.
2. Legal basis
| Legal instrument | Reference / number | Publication (entry into force) | Provisions |
|---|---|---|---|
| Law of the Republic of Armenia On Medical Assistance and Services to the Population | No. ЗР-42 of 1996-03-04; Chapter 10 in the new wording of 2022-07-06 | Amended by Law No. ЗР-504 of 2024-12-17 | Chapter 10 (definitional provisions) |
| General Requirements for the Safety and Efficacy of Medical Devices and Requirements for Their Labeling and Operating Documentation | EEC Council Decision No. 27 of 2016-02-12 | 10 calendar days after entry into force of the Protocol | Article 2 (definitions) |
| Rules for the Classification of Medical Devices According to the Potential Risk of Their Use | EEC Board Decision No. 173 of 2015-12-22 | Amended by Board Decision No. 65 of 2023-05-23 and No. 18 of 2025-02-18 | Article 3 (definitions) |
3. Expert insights
Insight 1: the national definition and the Union definition are two landings of the same text. The definition of a medical device given in Chapter 10 of the Law On Medical Assistance and Services to the Population is structurally identical to the definition in Article 2 of Decision No. 27: instruments, apparatus, devices, materials and other products, used alone or in combination for medical purposes, including the accessories necessary for the intended use (including dedicated software); the purposes cover prevention, diagnosis, treatment, medical rehabilitation, monitoring of the state of health of the human body, medical research, replacement, alteration or restoration of anatomical structures or physiological functions, and prevention or artificial termination of pregnancy; and the principal intended action is not achieved in or on the human body by pharmacological, immunological, genetic or metabolic means, although it may be assisted by such means. GRZAN notes that Article 2 of Decision No. 27 also states that standalone software is regarded as an active medical device; the national text carries this through in the form of the phrase including dedicated software, and in practice the stricter Union reading should be applied.
Insight 2: the handle for drug-device combinations is the principal mode of action. The qualifier that the action may be assisted by pharmaceuticals is the key exception. Drug-coated stents and antibiotic-loaded bone cement achieve their principal function through physical or structural means and are registered as medical devices; where the principal function is achieved by the medicinal substance, the product falls into the medicinal product track. This single judgment determines the competent authority and the entire documentation system, and it is the highest-priority decision at project initiation.
Insight 3: the IVD definition sweeps in reagents, calibrators, control materials and reference materials. Both Chapter 10 and Decision No. 27 define devices for in vitro diagnosis separately: any instrument, apparatus, device, material, reagent, calibrator, control material, reference material, test substance or other product, used alone or in combination, including necessary accessories and dedicated software, intended for in vitro examination of specimens from the human body in order to determine a physiological or pathological state (congenital pathology, susceptibility to a particular clinical condition or disease), tissue compatibility with a potential recipient, prediction of response to therapy, or selection of and monitoring of therapeutic agents. GRZAN notes: IVDs are exempt from technical testing (see article 17 of this series), which differs completely from the handling of non-IVD devices.
Insight 4: the characterization result directly determines three downstream variables. It determines which classification rule applies (Board Decision No. 173), which documentation list is submitted (Annex 4 of Council Decision No. 46), and whether a manufacturing site inspection is triggered (Council Decision No. 106). One wrong step in characterization makes all three subsequent steps wrong, and the rework cost concentrates in the testing work that has already been completed and paid for.
4. Practical recommendations
- Characterize at project initiation: complete a characterization memorandum before the technical specification is frozen, stating the reasoning and the cited provisions for the principal mode of action.
- Argue the principal function first for drug-containing products: prepare a mode-of-action justification that demonstrates at the mechanism level that the principal function is not achieved by pharmacological, immunological, genetic or metabolic means, and pre-assemble a set of answers to the questions the authority is most likely to raise.
- Handle software separately: submit standalone software for classification and documentation as an active medical device in its own right; do not merge it into a hardware submission as an accessory, unless it does not itself constitute an independent medical device.
- Build a separate IVD dossier: prepare IVD products under the IVD documentation structure, and never transplant the technical testing requirements of non-IVD devices.
5. Verification checklist
- □ A product characterization memorandum has been issued stating the principal mode of action.
- □ The characterization of standalone software in Article 2 of Decision No. 27 has been checked.
- □ The carry-through of the phrase including dedicated software in the Chapter 10 definition has been checked.
- □ A mode-of-action justification has been completed for drug-containing products.
- □ It has been confirmed that IVD products follow the separate IVD definition and documentation structure.
- □ The classification rule entry corresponding to the characterization result has been confirmed.
- □ Whether the characterization result triggers a manufacturing site inspection obligation has been confirmed.
Regulatory information verified as of October 30, 2026.
GRZAN services. GRZAN International provides product characterization, mode-of-action justification and IVD pathway assessment for EAEU and Armenian submissions. For inquiries, contact info@grzanglobal.com.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us