Malaysia is the second largest medical device market in ASEAN, and a market that Chinese companies cannot bypass when expanding into Southeast Asia. On first contact with MDA registration, three terms cause the most difficulty: AR, CAB and MeDC@St. This article sets out the core rules of Malaysian MDA registration through 24 questions and answers, and separately examines the Singapore-Malaysia-Thailand trilateral regulatory reliance mechanism that came into effect in 2026.

VERIFICATION BASISAll conclusions have been checked against the Medical Device Act 2012 (Act 737) and official MDA and HSA documents, and may be used directly as a reference for project planning.

1. Regulatory Framework and Basic Concepts

Q1. Which authority regulates medical devices in Malaysia, and under what law?

Medical devices in Malaysia are regulated by the Medical Device Authority (MDA) under the Ministry of Health (MOH).

LevelLegal BasisContent
Parent lawMedical Device Act 2012 (Act 737)Registration, licensing and enforcement fundamentals
Subsidiary legislationMedical Device Regulations 2012, P.U.(A) 500/2012Classification rules, conformity assessment requirements and fee schedule
Guidance documentsMDA/GD/xxxx seriesTechnical requirements for specific implementation

All registration applications, licence applications and change notifications are submitted through the MDA electronic system MeDC@St (Medical Device Centralised Online System).

Q2. How are medical devices classified in Malaysia?

Malaysia adopts the ASEAN Medical Device Directive (AMDD) framework, dividing devices into Class A, B, C and D by ascending risk. Indonesia, Thailand, Vietnam and other ASEAN countries use the same four-tier logic.

Classification is based on: duration of use, degree of invasiveness, whether the device contains energy or is active, intended use, and the vulnerability of the human body. The detailed rules are in the First Schedule of the Medical Device Regulations 2012 and the MDA classification guidelines (MDA/GD/0009 for general devices, MDA/GD/0001 for IVD).

KEY POINTMalaysia has required mandatory online classification since May 2025. Applicants must submit a classification request through the MDA system and obtain an electronic classification letter, which is the basis for every subsequent submission step.Classification cannot be decided by judgement alone; an incorrect classification will lead to the application being rejected.

Q3. What process does each of the four classes go through?

In one line: Class A is exempt from CAB, while Classes B, C and D must go through CAB. This is the most distinctive feature of Malaysia compared with some other ASEAN countries.

ClassConformity Assessment RequirementTypical Products
Class A** Low riskExempt from CAB review; self-declaration plus direct MDA reviewMedical gauze, non-sterile gloves, tongue depressors, stethoscopes
Class B** Low to medium riskMust undergo CAB conformity assessment plus final MDA reviewInjection needles, nebulisers, electronic blood pressure monitors, infusion sets
Class C** Medium to high riskMust undergo CAB conformity assessment (higher technical requirements) plus final MDA reviewDialysis equipment, ventilators, general orthopaedic implants, monitoring equipment
Class D** High riskMust undergo full CAB review plus final MDA reviewCardiac pacemakers, heart valves, implantable defibrillators, stents

Under MDA/GD/0070, Class A devices are exempted from the CAB conformity assessment procedure (see the Medical Devices (Exemption) Order 2024). Classes B, C and D must be assessed by a CAB registered with the MDA.

Q4. What is AR, and why must a foreign manufacturer have one?

AR (Authorized Representative) is the statutory representative of a foreign manufacturer in Malaysia.

MANDATORY REQUIREMENTUnder Section 5 of Act 737, a medical device must be registered with the MDA before it may be imported, exported or placed on the market. A foreign manufacturer cannot submit a registration application directly to the MDA and must appoint an AR established in Malaysia to hold the registration certificate, submit applications, respond to official enquiries and discharge post-market obligations.

An AR must hold both of the following qualifications, without exception:

• an Establishment Licence issued by the MDA; and

• a GDPMD certificate (Good Distribution Practice for Medical Devices), certified by a body recognised by the MDA.

PRACTICAL WARNINGIf the GDPMD certificate of your AR lapses, your registration process will stall immediately. When selecting an AR, always verify the status and validity dates of its licences and certificates.

Q5. What is the difference between an AR and a distributor?

The two have entirely different legal roles and must not be conflated.

RoleLegal PositionResponsibilities
AR (Authorized Representative)The statutory responsible party under Act 737Holds the registration certificate, liaises with the MDA, maintains records, handles post-market vigilance and change notifications, and discharges all compliance obligations
DistributorAuthorised by the AR or the manufacturerSells and distributes in the Malaysian market; has no direct registration legal relationship with the MDA

Both require an Establishment Licence, but only the manufacturer or the AR may apply for product registration.

PRACTICAL RISKMany companies take the shortcut of having a distributor also act as the AR. The problem is that the registration certificate is then bound to the distributor. If the commercial partner changes, the Change of Ownership becomes extremely difficult and may affect continuity of market access. Using an independent AR that is not a distributor keeps control of the certificate in your own hands.

2. Registration Routes and Core Process

Q6. How many conformity assessment routes are available for Classes B/C/D?

There are two main routes. Whether the faster route is available depends on whether the product already holds approval from a recognised reference authority.

RouteApplicable ConditionAssessment Method
FullThe product holds no approval from any MDA-recognised reference agencyFull technical review by the CAB (technical documentation, QMS and on-site audit if requested). Closest to the EU notified body model.
Verification (abridged)The product holds approval from at least one MDA-recognised reference agencyThe CAB carries out a verification-style assessment based on existing reference market assessment results, significantly reducing workload and cycle time.

MDA-recognised Competent Authorities include:

No.AuthorityRecognition Basis
1European Union — EU Notified BodiesIncluded in the first version of the list (March 2025)
2Japan — MHLW / PMDAIncluded in the first version of the list
3Australia — TGAIncluded in the first version of the list
4Canada — Health CanadaIncluded in the first version of the list
5United States — FDAIncluded in the first version of the list
6United Kingdom — MHRAIncluded in the first version of the list
7Singapore — HSAAdded by MDA Circular Letter No. 2 Year 2025 (October 2025), implemented together with MDA/GD/0068 and the second version of MDA/GD/0070
8Thailand — Thai FDAAdded by MDA Circular Letter No. 2 Year 2025 (October 2025), implemented together with MDA/GD/0068 and the second version of MDA/GD/0070
VERSION TRAPThe Annex 1 list of recognised agencies in the first version of MDA/GD/0070 (March 2025) contains only the first six agencies (EU NB, MHLW, TGA, Health Canada, FDA, MHRA) and does not include Singapore and Thailand.Singapore and Thailand were formally added to the list of approvals acceptable under the verification procedure by MDA Circular Letter No. 2 Year 2025 (October 2025), implemented together with MDA/GD/0068 and the second version of MDA/GD/0070.Citing the older guideline leads to the incorrect conclusion that Singapore or Thai approval cannot use the verification route. Always refer to the latest version.

Under Annex 1 of MDA/GD/0070, acceptable approval types include FDA 510(k) or PMA, EU MDR/IVDR certificates, TGA ARTG registration and Canadian Health Canada licence.

IMPORTANT EXCLUSIONSThe verification route is not available to devices authorised through special access pathways, including the EU special authorisation, PMDA compassionate use or emergency pathways, TGA SAS or authorised prescriber pathways, Canadian SAP or interim orders, and FDA EUA, expanded access or HDE.Approvals obtained through emergency or compassionate use channels cannot be used to access the abridged route.

Q7. What conditions must be met to use the verification route?

Under Section 4.3 of MDA/GD/0070, the product must satisfy all of the following:

• holds approval from at least one MDA-recognised reference agency or notified body;

• the design and intended use of the device to be marketed in Malaysia are fully identical to the approved device;

• no global safety issues in the last 1 year, specifically:

– no death reports; and

– no reports of serious deterioration in health status; and

– no unclosed field safety corrective actions, including recalls, at the time of submission; and

– all incident reports and corrective and preventive actions have been closed; and

• the product has not been refused or had its approval withdrawn by any reference agency.

Note: where the approval of the reference agency has been revoked or has lapsed, the product must return to the full assessment route.

Q8. What are the two stages of registration?

Registration proceeds in two stages (Class A is exempt from Stage 1):

• Stage 1 — Conformity Assessment: the manufacturer collects and compiles conformity evidence and submits it to the CAB for assessment;

• Stage 2 — Application to MDA: the CAB assessment report, certificate and complete technical dossier are submitted through MeDC@St, and the MDA reviews the application and decides whether to issue a certificate.

The conformity assessment must cover four elements (Part II of the Third Schedule of the Regulations):

#ElementDetail
1Quality Management System (QMS)As above
2Post-market Surveillance (PMS)As above
3Technical DocumentationAs above
4Declaration of Conformity (DoC)As above

All evidence must be structured according to the CSDT (Common Submission Dossier Template), the template for which is set out in MDA/GD/0008.

Q9. How long is the registration certificate valid, and when can renewal begin?

Five years. After registration or re-registration, the device is valid for 5 years and is listed in the Malaysian Device Access Register (MDAR).

Re-registration:

• the system opens the renewal window 1 year before the certificate expires;

• renewal is not a fresh submission from scratch; the MDA reviews whether the existing registration record, production documentation and conformity assessment status remain valid;

• the MDA verifies that the AR or manufacturer licences and the device CAB certificates are still within their validity periods.

RISK WARNINGMissing the renewal window means the product may no longer be lawfully sold in Malaysia until the registration is restored.The expiry date should be entered into the compliance calendar, with the inventory started at least a year in advance, because an expired QMS certificate, a lapsed CAB certificate or an unapproved change notification will each become an obstacle to renewal.

Q10. What are the official review timeframes at each stage?

Under Section 6.4 of MDA/GD/0070, MDA review timeframes, assuming complete documents and fees paid, are:

StageTimeframe
Class A — MDA review14 to 20 working days
Class B / C / D — MDA review30 working days (may sometimes be exceeded)
CAB — verification routeApproximately 1 to 2 months in total
CAB — full routeApproximately 2 to 10 months in total

The CAB timeframe covers completion of assessment, preparation of the report and issuance of the certificate; it does not include the time required for the company to respond to queries.

Indicative end-to-end estimates:

• Class A: CAB review exempted;

• Class B/C/D full route: official target 2 to 10 months;

• Verification route: approximately 1 to 2 months.

IMPORTANT SUPPLEMENTThere is a definite deadline for responding to MDA queries. Additional documents must be provided within 90 working days of receipt of the official enquiry. Failure to submit on time results in rejection of the application, and the application fee is not refundable.

Q11. What are the official fees?

Under Section 6.5 of MDA/GD/0070, fees are set out in the Fifth Schedule of the Medical Device Regulations 2012, charged per application and corresponding to a 5-year validity period:

CategoryApplication FeeRegistration Fee
Class A100 RM—
Class B250 RM1,000 RM
Class C500 RM2,000 RM
Class D700 RM3,000 RM
Devices containing a medicinal substance700 RM5,000 RM
IMPORTANT 2026 UPDATEUnder the Medical Devices (Amendment) Regulations 2025 [P.U.(A) 330], the fee structure for Class A changed with effect from January 2026: the application fee was raised from RM 100 to RM 500, and a new registration fee of RM 750 was introduced.This marks the point at which Class A also moved onto a cost-recovery basis of charging. Before filing, always check the current fee schedule on the MDA website.

Note: the above are government fees only. Third-party costs include CAB assessment fees (varying by class and body, typically RM 15,000 to 50,000 and above), AR annual fees, GDPMD certification fees and Establishment Licence annual fees.

3. Establishment Licensing and Compliance Operations

Q12. What is an Establishment Licence, and who needs one?

Under Section 15(1) of Act 737, no establishment may import, export or place a registered medical device on the market before obtaining an Establishment Licence.

Establishments Requiring a LicenceBasis
ManufacturerImports, manufactures or places devices on the market
ImporterImports devices
DistributorDistributes devices
Authorised Representative (AR)Holds the registration certificate on behalf of a foreign manufacturer

Under MDA announcements, a “One Licence per Role” policy has applied since 1 July 2024. Licences are valid for 3 years, independent of the 5-year cycle of the product registration certificate.

On practical points: where documents are complete, the MDA commits to approving licence applications within 14 to 21 working days.

Q13. What is GDPMD, and why is it so critical?

GDPMD (Good Distribution Practice for Medical Devices) is Malaysia’s quality management requirement for the distribution of devices, audited and certified by an MDA-recognised certification body (CAB).

It is not optional but a prerequisite for licensing:

• importers, distributors and ARs must hold a GDPMD certificate in order to obtain an Establishment Licence;

• manufacturers require ISO 13485 and must also satisfy GDPMD requirements.

TWO CONSEQUENCES TO REMEMBERLapse of the GDPMD certificate halts registration. If the AR’s GDPMD expires, all pending applications will be blocked.At renewal, the GDPMD must be within its validity period, otherwise the licence cannot be renewed.

Compliance recommendation: manage the AR’s GDPMD expiry date, the Establishment Licence expiry date and the product registration expiry date together on a single monitoring table.

Q14. Do import permits need to be obtained now?

Not yet mandatory, but they should be prepared in advance. The MDA has established import control with Malaysian Customs through the ePermit / DagangNet system. Under this system, all medical device importations will require an import permit.

MilestoneTiming
Original planned mandatory implementationJanuary 2026
Postponed toJuly 2027
Voluntary declaration window opensJune 2026

Recommendation: include the import permit in the 2026 to 2027 preparation list, complete system registration and process rehearsal ahead of time, so as to avoid being unprepared when it becomes mandatory.

Q15. What ongoing compliance obligations apply after registration?

Registration is only the beginning. The holder (AR or manufacturer) must continuously discharge post-market obligations under Act 737:

• Distribution records: must be traceable and retained for at least 5 years;

• Post-market surveillance (PMS) and vigilance: a PMS system must be established and maintained, available for official review;

• Complaint handling: all complaints must be recorded, investigated and closed out;

• Mandatory Problem Reporting: see Q16;

• Field Corrective Actions (FCA) and recalls: to be executed and reported where necessary;

• Periodic safety update reports: as part of registration maintenance;

• Change Notification: any change affecting safety, performance or labelling must be notified.

Legal anchors: Sections 37 to 42 of Act 737 respectively cover distribution records, post-market surveillance and vigilance, complaint handling, mandatory problem reporting, field corrective actions and recalls.

Q16. What are the statutory deadlines for adverse event reporting?

Under Section 40 of Act 737 and the MDA mandatory problem reporting requirements, deadlines vary by the nature of the event:

Event TypeReporting Deadline
Serious threat to public healthWithin 48 hours
Resulted in death or serious deterioration in health statusWithin 10 calendar days
Device malfunction or loss of effectiveness, without death or serious deteriorationWithin 30 calendar days

Supplementary rules:

• the investigation (final) report must be submitted within 30 days of the initial report;

• regardless of whether the incident occurred in or outside Malaysia, it must be reported to the MDA where the device is registered in Malaysia. An exemption applies where a report has already been made to the regulator of the country of occurrence and the necessary corrective actions have been taken for all affected devices in the Malaysian market.

Reporting channel: online through the MDA MeDCReSt system; combination products must be submitted by email to mpr@mda.gov.my.

4. Labelling, Changes and Systems

Q17. What language should the label and IFU use?

ItemRequirement
General requirementEnglish is sufficient for labels
Home use devicesThe IFU must be provided in Bahasa Malaysia; English may be added as a supplementary language
Submission documents and official correspondenceEnglish and/or Bahasa Malaysia are both acceptable

Mandatory label content (summary):

• the Malaysian medical device registration number (the MDA encourages use of the QR code on the registration certificate);

• device name, model, batch or serial number, date of manufacture and/or expiry;

• manufacturer name, address and contact details (contact details for foreign manufacturers are optional);

• the AR’s name, address and contact details;

• product description and intended use;

• adverse side effects, limitations, warnings and precautions;

• post-market service, retirement and disposal information.

TWO RED LINESThe registration number must be printed directly on the label; it cannot be affixed by a later sticker.The names or logos of the MDA, the Ministry of Health or any authority must not appear on the label, otherwise it will be deemed to imply official endorsement.

On Halal certification: Act 737 does not require medical devices to obtain Halal certification, and the MDA does not treat Halal as a registration requirement. Halal certification (MS 2636) is managed separately by JAKIM and is a voluntary commercial plus. Products using animal-derived materials, such as porcine or bovine components, should nevertheless monitor developments.

Q18. What should be done if the design or label is changed after registration?

A Change Notification (CN) process applies. Malaysia classifies changes into three categories according to their impact:

CategoryNature of ChangeFiling Requirement
Category 1Major changes that may affect safety or performanceMust be re-registered before implementation
Category 2Moderate changes requiring MDA review, such as labelling, materials or designMDA acceptance must be obtained before implementation
Category 3Minor administrative changes, such as software interface adjustments or non-substantive labelling changesNotification only; may be implemented immediately

Review timeframe: approximately 30 working days for an individual Category 2 or 3 application; approximately 60 working days for combined or batch submissions.

Q19. What is MeDC@St, and why should it be familiarised with in advance?

MeDC@St (Medical Device Centralised Online System) is the MDA unified electronic submission system, covering classification applications, registration applications, licence applications and renewals, change notifications and payment of fees.

Why it matters:

• all submissions must go through it; there is no offline channel;

• the application fee must be paid within the window notified by the MDA, normally 30 days; overdue applications are voided;

• if an official input request is received after submission, each point must be answered in writing with the corresponding file names identified in the system;

• the platform itself is evolving: MeDC@St 3.0 and an automated re-registration route are in progress, involving the transition to a new change management framework.

Practical recommendation: the AR must hold a valid MeDC@St account and be capable of operating it. Completing an account and file format rehearsal before filing significantly reduces delays caused by systemic issues.

5. Singapore-Malaysia-Thailand Trilateral Regulatory Reliance

Q20. What exactly is the “New Malaysia-Thailand-Singapore mutual recognition”?

Its accurate name is regulatory reliance, not “one certificate for all markets” or an automatic mutual recognition certificate.

THE MOST COMMON MISCONCEPTIONObtaining an HSA or Thai FDA registration does not mean you automatically obtain Malaysian registration.The essence of reliance is that one regulator relies on the regulatory assessment already completed by another regulator in order to compress its own review stages and cycle time. But the application, CAB verification, local AR, fee payment and final issuance of the certificate all remain mandatory.

The actual position across the three countries is two-way reliance on one pair, and none on the other:

PairDirectionStatusEffective Date
Malaysia ↔ Singapore** (MDA ↔ HSA)Bidirectional✅ In effectLong-term mechanism from 1 March 2026
Malaysia ↔ Thailand** (MDA ↔ Thai FDA)Bidirectional✅ In effectFully implemented from 1 May 2026, no end date
Singapore ↔ Thailand** (HSA ↔ Thai FDA)None❌ Not establishedNot established as at October 2026; Thailand unilaterally recognises HSA approval results only

In other words, the hub of the trilateral arrangement is the Malaysian MDA. Singapore and Thailand have each established bidirectional reliance with Malaysia, but there is no direct arrangement between Singapore and Thailand. For Chinese companies this means that using Malaysia as a springboard to reach all three countries with a single registration is genuinely feasible (see Q23).

Q21. How does the Malaysia–Singapore reliance work in practice?

This is a bidirectional arrangement. On 22 August 2025 the two countries signed a Memorandum of Understanding at the 14th ASEAN Medical Device Committee (AMDC) meeting in Siem Reap. The pilot ran from 1 September 2025 to 28 February 2026, and from 1 March 2026 it became a long-term mechanism.

Direction 1: Singapore to Malaysia

• Class B/C/D devices already registered on the Singapore Medical Device Register (SMDR) may use the MDA verification route;

• CAB verification takes approximately 30 working days (compared with 60 for a full assessment), followed by MDA registration review of approximately 30 working days;

• the product must be fully identical to the Singapore registered version, with no change in design, specification, intended use, indications or raw materials.

Direction 2: Malaysia to Singapore

• Class B/C/D devices registered with the Malaysian MDA through full CAB conformity assessment and MDA review may use the HSA abridged evaluation route;

• review time is reduced by up to approximately 30 per cent compared with the HSA standard full route;

• when filing in Singapore, select “Regulatory reliance Program: Malaysia Medical Device Authority (MDA)” under Evaluation Route in the SHARE system;

• submit: a copy of the MDA registration certificate, the CAB conformity certificate, and the abridged route documents specified by GN-15.

Note: according to the official MDA announcement, HSA approval has been included in the list of approvals acceptable under the verification procedure with effect from 1 March 2026; the official press release was issued during the 29th IMDRF meeting in Singapore.

Q22. How does the Malaysia–Thailand reliance work in practice?

Also bidirectional. The two regulators signed a confidentiality agreement at the 29th Global Harmonization Task Force (GHWP) annual meeting in Bangkok in December 2025. The pilot ran from 1 February 2026 to 30 April 2026, and it has been fully implemented since 1 May 2026 with no end date.

Direction 1: Thailand to Malaysia

• Class B/C/D devices already approved by the Thai FDA may use the MDA verification route in the same way as Singapore;

• CAB verification takes approximately 30 working days and MDA registration review approximately 30 working days.

Direction 2: Malaysia to Thailand

This direction is of considerable value to companies already registered in Malaysia:

• devices approved by the MDA are exempted from the external expert review that Thailand otherwise mandates when submitting a complete CSDT;

• the review period is shortened from approximately 150 calendar days to approximately 120 calendar days, and the expert review fee is waived;

• external expert review has historically been the most time-consuming and least predictable stage of the Thai registration process, so exempting it represents a substantive acceleration.

The hard conditions applying on the Thai side must each be satisfied:

• the device must hold a formal MDA registration approval, excluding products approved through special notification routes or exempted from quality, safety and effectiveness assessment;

• the version submitted to Thailand must be fully identical to the MDA-approved version: trade name, model designation, product code, intended use, indications, labelling, accompanying documents and packaging must be identical;

• a complete CSDT dossier identical to the version submitted to the MDA must still be submitted. The documentation is not reduced; what is removed is the review stage.

Q23. How can a Chinese company use this mechanism to reach three countries at once?

Because the hub of the reliance mechanism is Malaysia, using Malaysia as the regional pivot is currently the most efficient three-country strategy. Two recommended routes:

RouteSequenceSuitable For
Route A: Singapore first, then MalaysiaSingapore HSA registration → use the SMDR record to access Malaysia through the MDA verification route → enter Thailand through the unidirectional reliance channel (Thailand relies on Singapore)Products that already have, or can readily obtain, Singapore registration; products already approved by the five major reference agencies (US, EU, Canada, Australia, Japan)
Route B: Malaysia first, then SingaporeMalaysia full assessment registration → use the MDAR record to access Singapore through the HSA abridged route (approximately 30 per cent faster) → enter Thailand using the Malaysian certificateProducts for which Malaysia is the core market, or where Singapore approval is less certain

Three practical reminders:

  1. Product consistency is a hard threshold. The precondition for the three-country linkage is that the product is the same version. Any design, labelling or indication change made for one country may break the applicability of the entire chain, so regional versions must be planned consistently.
  2. Class A devices are not covered. The reliance mechanisms cover only Class B/C/D devices (corresponding to Thai Class 2/3/4). Class A procedures are already simple, so the conventional simplified route in each country is sufficient.
  3. Reliance does not remove the local holder requirement. Under any route, Malaysia still requires a local AR, Singapore still requires a local Registrant, and Thailand still requires a local licence holder. Institutional licensing and compliance obligations in each jurisdiction are not relieved by reliance.

Strategic implication: for Chinese companies expanding into ASEAN, “one core registration plus two reliance levers” can significantly compress the overall time and compliance cost for the three countries, avoiding three independent full assessment processes. This is a direction explicitly supported by the Director-General of the MDA at the 29th GHWP annual meeting: regulatory reliance is becoming the mainstream model of ASEAN cooperation, and early movers capture the benefit of the mechanism.

6. Five Common Pitfalls

No.PitfallExplanation
1Incorrect classificationThe classification letter is the foundation of all submissions. If the actual risk level is higher than declared, the application will be rejected and credibility affected. Independent review of the classification is recommended; do not assume EU classification equals Malaysian classification, particularly for AI and software devices (SaMD), where Malaysia applies AMDD rules rather than EU MDR Article 10(3) and boundary cases may reach different conclusions.
2Conflating the AR and distributor rolesHaving a distributor also act as the AR means handing over control of the registration certificate. Changing partner means changing certificate, and the transfer process is cumbersome and affects continuity of market access.
3Underestimating the eligibility threshold for the verification routeThe abridged route requires prior approval by a reference agency, full identity of intended use with the approved version, and no unresolved safety issues in the last year. Products approved through emergency or compassionate channels (EUA, SAS, SAP, HDE) cannot use the verification route. Many companies fall into this trap.
4Ignoring the validity linkage of GDPMD and licencesGDPMD expiry halts registration; licence expiry without renewal closes the import channel. The three certificates — GDPMD, Establishment Licence and product registration — must be managed together.
5Implementing changes before notificationCategory 1 and 2 changes must have the filing completed or the acceptance obtained before implementation, failing which the change is non-compliant. Many companies change first and deal with it later, which surfaces at renewal and causes registration interruption.