Kazakhstan IVD Registration: Same-Track Procedures and Dedicated Requirements
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
In vitro diagnostic devices (IVDs) register on the same track as medical devices in Kazakhstan, applying the same Order No. ҚР ДСМ-16 and the same expert review procedures, but with dedicated requirements at three points: classification (under Order No. ҚР ДСМ-281/2020 and the IVD-specific rules of Decision No. 173 at Union level, with the treatment of these categories updated by Amendment No. 18 of February 18, 2025), samples and stability evidence (shelf life, shipping and storage condition verification for reagent products), and clinical evidence (the combined package of analytical and clinical performance data). The most common bottleneck in IVD projects is not the procedure itself but the preparation of sample and stability materials. Reagent kits cannot be summoned quickly: their production lots, shipping conditions and remaining shelf life all constrain the laboratory study calendar, and a delay in any one of them propagates directly into the expert review timeline. Because the expert center procures or receives samples under defined rules, sample logistics should be treated as a critical-path workstream with its own owner, rather than as an administrative afterthought to dossier writing.
2. Legal basis
| Regulation | Reference / number | Date | Provisions |
|---|---|---|---|
| Order of the Minister of Health approving the Rules for State Registration, Re-registration and Variations to Registration Dossiers of Medicines and Medical Devices | No. ҚР ДСМ-16 | February 9, 2021 (Ministry of Justice registration No. 22175) | Registration dossier requirements |
| Order of the Minister of Health approving the Classification Rules for Medical Devices | No. ҚР ДСМ-281/2020 | 2020 | IVD classification rules |
| Decision of the Eurasian Economic Commission Executive Board on the Classification of Medical Devices According to the Potential Risk of Application | Decision No. 173 (as amended by No. 65 and No. 18) | December 22, 2015; amendments May 23, 2023 and February 18, 2025 | Updated IVD classification criteria |
| Order of the Minister of Health approving the Rules for the Expert Assessment of the Safety, Quality and Efficacy of Medicines and Medical Devices | No. ҚР ДСМ-10 | January 27, 2021 | Expert review and laboratory studies |
| Order of the Minister of Health approving the Rules for the procurement of medicines and medical devices for the purpose of expert assessment | No. ҚР ДСМ-154/2020 | 2020 | Procurement of samples for expert review |
3. Expert insights
Insight one: IVD classification was specifically adjusted in 2025, so old conclusions must be re-verified. Amendment No. 18 of February 18, 2025 adopted by the Eurasian Economic Commission Executive Board adjusted the treatment of software and IVD products within Decision No. 173. Every IVD classification conclusion formed before 2025 should be re-checked against the new criteria, particularly for companion diagnostics, self-testing products and blood screening products, where the risk of a class shift is highest. A classification change after submission invalidates the evidence plan, so the re-check is best performed before commissioning any new testing.
Insight two: whether reagent and instrument are integrated or separated determines the dossier structure. Reagents and their paired instruments can be registered separately or registered together as a system. The choice drives the future frequency of variations: with separate registration, an instrument software upgrade and a reagent formulation change proceed without touching each other; with system registration, any change on one side may require an assessment of its impact on the other. The decision should therefore be made on projected change frequency and commercial structure, not on submission convenience alone, and once made it should be recorded in the project file with its rationale.
Insight three: stability evidence is the most underestimated part of an IVD project. The shelf life of a reagent, in-use stability, shipping stability and storage condition verification together form a distinct evidence set. The review checks whether the remaining shelf life of the samples satisfies the study period requirement, and it also checks the consistency between the stability conclusions and the label claims. The GRZAN experience is that stability studies must be scheduled before sample shipping, not after the samples have arrived, because the clock on remaining shelf life runs from manufacture and cannot be paused while the dossier waits.
Insight four: measuring devices may additionally fall within metrology control. IVD instruments with measurement functions, such as blood glucose meters and biochemical analyzers, must also be assessed against the type approval and state registration requirements under the law on ensuring the uniformity of measurements. This is an independent compliance track that runs in parallel with device registration and carries its own evidence and timeline, so it must be planned alongside rather than after the registration project. The evidence demanded by the two regimes also differs: registration looks at analytical and clinical performance, while metrology looks at measurement accuracy and traceability of reference values, so a single combined study rarely satisfies both and the two evidence sets should be built deliberately.
4. Practical recommendations
- After Amendment No. 18 of February 18, 2025 takes effect, re-verify the classification conclusions of all existing IVD products in the portfolio.
- At project initiation, decide between separate registration of reagents and instruments or system registration, and assess the resulting future variation frequency.
- Schedule stability studies before sample shipping, and confirm that the remaining shelf life of samples covers the full study period including any buffer for review questions.
- For measuring instruments, verify the metrology type approval and state registration requirements in parallel.
- File analytical performance and clinical performance evidence in separate volumes so that reviewers can retrieve each set without searching a mixed file.
5. Verification checklist
- □ IVD classification has been re-verified against the criteria after Amendment No. 18 of February 18, 2025
- □ The registration structure for reagents and instruments (separate or system) is decided
- □ Shelf life, in-use stability, shipping stability and storage condition verification are complete
- □ The remaining shelf life of samples covers the study period
- □ Metrology type approval requirements for measuring instruments have been checked
- □ Analytical and clinical performance evidence are filed in separate volumes
GRZAN provides registration support in Kazakhstan for manufacturers, importers and authorized representatives, covering IVD classification, stability evidence planning and expert review follow-up. For assistance, contact info@grzanglobal.com.
Regulatory information verified as of October 29, 2026.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us