Kazakhstan Medical Device Risk Classification: Classes 1, 2A, 2B and 3
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Medical device classification in Kazakhstan uses a four-class system: Class 1 (low risk), Class 2a (medium-low risk), Class 2b (medium-high risk) and Class 3 (high risk). The classification outcome functions as a master switch: it simultaneously determines the depth of expert review, the scope of technical testing and toxicological studies, whether a manufacturing site inspection is required, and the coverage of that inspection. Classification is determined under Order No. ҚР ДСМ-281/2020 of the Minister of Health for the national pathway, while the Eurasian Economic Commission Executive Board Decision No. 173 (December 22, 2015, as amended by No. 65 in 2023 and No. 18 in 2025) applies for the Union pathway. A wrong classification decision forces the entire evidence package to be rebuilt, which makes misclassification the most expensive category of rework in a Kazakhstan project. Because the class drives every downstream resource decision, from laboratory budgets to inspection logistics, the classification conclusion should be documented, reviewed and frozen before any testing is commissioned. Once the class is fixed, the remaining planning variables fall into place: the number and depth of laboratory studies, the format of the clinical evidence, and the inspection expectations can each be budgeted with a known risk profile instead of provisional ranges.
2. Legal basis
| Regulation | Reference / number | Date | Provisions |
|---|---|---|---|
| Order of the Minister of Health approving the Classification Rules for Medical Devices | No. ҚР ДСМ-281/2020 | 2020 | Full text of the classification rules; effective 10 calendar days after first official publication |
| Decision of the Eurasian Economic Commission Executive Board on the Classification of Medical Devices According to the Potential Risk of Application | Decision No. 173 (as amended by No. 65 and No. 18) | December 22, 2015; amendments May 23, 2023 and February 18, 2025 | Four-class classification rules at Union level |
| Order of the Minister of Health approving the Rules for the Expert Assessment of the Safety, Quality and Efficacy of Medicines and Medical Devices | No. ҚР ДСМ-10 | January 27, 2021 | Trigger conditions for review and inspection |
| Decision of the Council of the Eurasian Economic Commission on requirements for the implementation, maintenance and evaluation of quality management systems for medical devices | Decision No. 106 | November 10, 2017 | Link between inspection coverage and risk class |
3. Expert insights
Insight one: two sources of classification rules, and choosing the pathway means choosing the classification basis. On the Kazakh national pathway, classification is determined under Order No. ҚР ДСМ-281/2020; on the Union pathway, it is determined under Decision No. 173. Both systems use a four-class structure, but the detailed rules differ, particularly for software and for in vitro diagnostic devices, and the amendment No. 18 of February 18, 2025 was aimed precisely at adjusting the treatment of these two categories. The GRZAN approach is to fix the pathway first and the classification second, because applying Union criteria to a national dossier, or the reverse, produces a classification statement that does not survive review and must be rewritten at the cost of schedule and credibility.
Insight two: the class determines the strength of evidence, not merely the number of documents. Taking manufacturing site inspections as an example, under Decision No. 106 the evaluation of the quality management system is mandatory for Class 2a sterile, Class 2b and Class 3 devices, while Class 1 and non-sterile Class 2a devices generally fall outside mandatory inspection. This boundary moves the project timeline directly: an inspection can take up to 90 working days and the period is not counted within the review timeline. Budgeting for a Class 2b project without a full inspection window therefore produces plans that fail the first time the inspection calendar is confirmed.
Insight three: sterility is an independent classification variable. Devices with the same intended use may attract different inspection requirements depending on whether they are supplied sterile or non-sterile. When planning a product family, managing sterile status as an independent dimension prevents two different compliance cost structures from emerging inside one model family. This practical estimate also simplifies change control, because the consequences of a packaging or sterilization change become predictable across the family rather than case by case.
Insight four: classification must be frozen before submission, not adjusted on the way. The classification conclusion is written into the classification section of the registration dossier and is bound to the review pathway, the test program and the inspection scope. Changing the classification during review is equivalent to restarting project initiation. GRZAN recommends locking the conclusion with a written classification statement before formal submission, including the rule provisions relied on and a comparison against similar classified products. That statement then becomes the anchor for every later conversation with the reviewer: when a question arises about test scope or inspection depth, the answer is traced back to the class and the rule provision rather than renegotiated case by case, which keeps the project defensible and the schedule intact.
4. Practical recommendations
- At project initiation, issue a written classification statement citing the applicable rule provisions and listing comparable products with their assigned classes.
- Re-verify the classification of software and in vitro diagnostic products after Amendment No. 18 of February 18, 2025, since prior conclusions may no longer match the current criteria.
- Predict inspection needs along the two dimensions of class and sterile status, and reserve an inspection window of up to 90 working days in the project plan.
- Where a model family contains configurations spanning several classes, prepare the evidence package to the highest class represented in the family, and record the mapping of each configuration to its class so that later additions can be placed without reopening the analysis.
5. Verification checklist
- □ The pathway (national or Union) is chosen and the classification basis rule is selected accordingly
- □ The classification statement cites the determining provisions and comparable products
- □ Sterile status is marked as an independent dimension
- □ Whether a manufacturing site inspection is triggered has been predicted from the class
- □ Cross-class configurations within the model family are prepared to the highest class
- □ The software and IVD criteria after Amendment No. 18 of February 18, 2025 have been re-checked
GRZAN provides registration support in Kazakhstan for manufacturers, importers and authorized representatives, covering classification determination, evidence planning and review follow-up. For assistance, contact info@grzanglobal.com.
Regulatory information verified as of October 29, 2026.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us