Uzbekistan Medical Device Registration: Software and Cybersecurity Documentation
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
Item 2.13 of Annex 3a of Resolution No. 738 sets out a distinct group of requirements for dedicated software: verification and validation results, software architecture and life cycle information, cybersecurity measures covering integrity controls, access controls, data protection, and tamper resistance, identification and marking, manufacturer test results, and a compliance tracking system. At the same time, several items of Annex 3a are explicitly stated not to apply to software, but cybersecurity and verification and validation are not among them: they are the core questions every software device must answer. Registration Order No. 3888 reinforces the software identity framework by requiring each device to carry a single GMDN code and a single UDI code and by bringing software versions into the device identity information.
2. Legal basis
| Regulation | Number | Date | Provision |
|---|---|---|---|
| Cabinet Resolution on approving the Regulation on the procedure of state registration of medicines and medical devices and issuance of registration certificates | No. 738 | Adopted November 24, 2025, effective February 26, 2026 | Item 2.13: information on dedicated software (where applicable), covering verification and validation results, software architecture and life cycle information, cybersecurity measures, identification and marking, manufacturer test results, and a compliance tracking system |
| Cabinet Resolution on approving the Regulation on the procedure of state registration of medicines and medical devices and issuance of registration certificates | No. 738 | Adopted November 24, 2025, effective February 26, 2026 | Items 2.3 and 2.4 (manufacturer information, development and production information) do not apply to software; items 2.19 and 2.20 (color proofs of labeling and packaging, instructions for use) do not apply to software |
| Cabinet Resolution on approving the Regulation on the procedure of state registration of medicines and medical devices and issuance of registration certificates | No. 738 | Adopted November 24, 2025, effective February 26, 2026 | Item 2.8 (sterility information) does not apply to software and in vitro diagnostic devices |
| Rules for the preparation of medical device registration files (registration order) | No. 3888 | Registered July 2026, effective October 8, 2026 | Each device must be assigned a single GMDN code and a single UDI code; software versions are included in the device identity information |
3. Expert insights
Insight 1: the center of gravity of a software dossier shifts from manufacturing to verification and cybersecurity. In Annex 3a, item 2.3 (the production-related parts of manufacturer information) and item 2.4 (development and production information, including technical drawings and the process flow) do not apply to software. This is not a lowering of the bar but a relocation of scrutiny to the risk dimensions specific to software: whether functions are realized as designed, addressed by verification and validation; whether the structure is traceable, addressed by the architecture and life cycle information; and whether the product resists attack and tampering, addressed by the cybersecurity measures. A software dossier built around a hardware-style manufacturing narrative will answer questions the reviewer is not asking and leave the real ones open.
Insight 2: cybersecurity is broken down into four concrete directions. Item 2.13 refines the cybersecurity measures into integrity controls, access controls, data protection, and tamper resistance. These four can be used directly as an evidence checklist: each direction corresponds to a set of measures and supporting verification evidence, and a missing direction is a missing item. For each direction, the dossier should state the control implemented, the threat it addresses, and the test or analysis that demonstrates it works, so the reviewer can move from requirement to proof without inference.
Insight 3: the software version enters the device identity, and change management therefore becomes stricter. Registration Order No. 3888 requires a single GMDN code and a single UDI code, and item 2.1 requires a description of the device and its composition. Once a software version is updated, the device identity changes, and an assessment is needed of whether the update triggers a change filing. This echoes the identical configuration requirement applied in the recognition pathway: the software version is part of the configuration. Version discipline in development, mirrored in the regulatory identity, prevents the situation where the marketed version no longer matches the registered one.
Insight 4: the compliance tracking system is a software-specific continuing obligation. Item 2.13 requires information on the compliance tracking system. This means a software device, after certification, must maintain a mechanism that traces the correspondence between software versions and their compliance status, rather than treating the registration as a one-time delivery. The system should record each released version, the evidence set supporting it, and any pending or approved change filings, so that every version in the field can be mapped to a registered state on demand.
4. Practical recommendations
- Build the software evidence list around the item 2.13 sub-elements: verification and validation, architecture, life cycle, cybersecurity, identification and marking, test results, and compliance tracking.
- Prepare cybersecurity measures and verification evidence separately for each of the four directions: integrity controls, access controls, data protection, and tamper resistance.
- Bring the software version number into the device master data and govern it together with the GMDN and UDI codes.
- Assess before every software release whether the update triggers a change filing, especially where it affects model identification or the intended use.
- Establish and document the compliance tracking system, stating the correspondence between software versions, change records, and compliance status.
5. Verification checklist
- □ All item 2.13 sub-elements are covered (verification and validation, architecture, life cycle, cybersecurity, identification and marking, test results, compliance tracking)
- □ Cybersecurity is evidenced separately for integrity controls, access controls, data protection, and tamper resistance
- □ The software version number is in the device master data and governed with the GMDN and UDI codes
- □ Software exemptions in Annex 3a have been checked item by item (items 2.3, 2.4, 2.8, 2.19, 2.20)
- □ A mechanism for assessing change filings before software updates is in place
- □ The compliance tracking system is documented and sustainably maintained
GRZAN supports manufacturers, importers, and authorized representatives with medical device registration in Uzbekistan, including software documentation under item 2.13, cybersecurity evidence, and change management for software updates. For assistance, contact info@grzanglobal.com.
Regulatory information verified as of October 29, 2026.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us