Uzbekistan Medical Device Clinical Trial Requirements and Evidence Rules

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Presidential Decree No. UP-137 establishes a principled requirement: the national registration of medical devices is based on the positive results of clinical trials, and devices with a high probability of risk are subject to mandatory verification based on positive clinical trial results. The only exception concerns devices with a low probability of risk, whose list is approved by the Ministry of Health. At the dossier level, item 2.18 of Resolution No. 738 requires the submission of a clinical effectiveness and safety report, which is generally not required for Class I devices, but sterile, measuring, and invasive Class I devices are excepted. The combination of a principled decree and a detailed dossier rule means that manufacturers must assess clinical evidence obligations at two levels at once: the policy level set by the decree and the document level set by the registration regulation.

Read together, the decree and the regulation give manufacturers a workable sequence: classify the device, test the three exceptions, and only then decide whether new clinical work is needed at all. Devices holding recognized foreign approvals can usually anchor their evidence on the clinical package behind that approval, while devices outside the recognition pathway need an evidence plan that can satisfy item 2.18 without duplicating studies that already exist elsewhere.

2. Legal basis

RegulationReference numberDateProvisions
Presidential Decree on Additional Measures for Regulating the Circulation of Medicinal Products and Medical DevicesNo. UP-137August 19, 2025From January 1, 2026, the national registration of medical devices is based on the positive results of clinical trials; devices with a high probability of risk undergo mandatory verification based on positive clinical trial results; the list of devices with a low probability of risk is approved by the Ministry of Health
Cabinet of Ministers Resolution approving the Regulation on State Registration of Medicinal Products and Medical Devices and the Issuance of Registration CertificatesNo. 738Adopted November 24, 2025; effective February 26, 2026Item 2.18: clinical effectiveness and safety report; generally not required for Class I, except for sterile, measuring, and invasive devices
Cabinet of Ministers Resolution approving the Regulation on State Registration of Medicinal Products and Medical Devices and the Issuance of Registration CertificatesNo. 738Adopted November 24, 2025Item 2.16: protocol of the study (trial) on the evaluation of biological action; item 2.15: adverse event and recall reports and the plan for collecting and analyzing post-market safety and effectiveness data
Presidential Decree on Additional Measures for Regulating the Circulation of Medicinal Products and Medical DevicesNo. UP-137August 19, 2025Establishment of bioanalytical laboratories to study the bioequivalence of medicinal products and drugs

3. Expert insights

Insight one: being based on positive results is a different question from having to run a local trial. The decree establishes a requirement of basis: the registration decision must be supported by positive clinical results. It does not necessarily require every registration to involve a new clinical trial conducted in Uzbekistan. Existing foreign clinical data, the clinical evidence underlying a foreign approval, and the reference approval under the recognition pathway can all serve as carriers of the positive results, provided the evidence can be verified and linked to the exact configuration submitted.

Insight two: three exceptions narrow the clinical exemption for Class I devices. Item 2.18 sets three exceptions to the Class I exemption: sterile, measuring, and invasive devices. The categories most heavily exported by manufacturers in the Chinese mainland, including single-use sterile consumables, electronic blood pressure monitors, and various invasive catheters, fall squarely within these three exceptions. Before assuming that a low-risk device is exempt from clinical evidence, manufacturers must first pass the three-exception test.

Insight three: clinical evidence and post-market data form one continuum, not two separate segments. Item 2.15 requires a plan for collecting and analyzing post-market safety and effectiveness data. The submission of clinical evidence is therefore not an endpoint but the starting point of a continuum that connects to post-market surveillance. When designing the clinical evaluation, manufacturers should include the post-market data collection plan in the same exercise, so that the two documents link together without gaps.

Insight four: the newly established bioanalytical laboratories target medicinal products, but they signal the direction of regulatory capacity building. The decree provides for bioanalytical laboratories dedicated to bioequivalence studies. Although they primarily serve medicines, this investment shows that the health system is strengthening its technical review capability. The implication for the device side is that the depth of review of clinical and performance data will keep increasing over time.

4. Practical recommendations

  1. At project initiation, confirm whether the product falls within the list of devices with a low probability of risk, using the list currently approved by the Ministry of Health.
  2. For Class I devices, first self-assess against the three exceptions of sterile, measuring, and invasive, and only then decide whether to prepare clinical evidence.
  3. Where foreign clinical data already exist, organize them into a verifiable evidence package covering the trial protocol, reports, ethics conclusions, and statistical conclusions.
  4. Compile the clinical evaluation report and the post-market data collection plan together, so that the endpoint of one document connects to the starting point of the other.
  5. For high-risk devices, reserve flexible time in the project schedule for the possible triggering of a local clinical study.

5. Verification checklist

  • □ It is confirmed whether the product falls within the list of low-probability-risk devices approved by the Ministry of Health
  • □ Class I devices have been self-assessed against the three exceptions of sterile, measuring, and invasive
  • □ Existing foreign clinical data have been organized into a verifiable evidence package
  • □ The clinical effectiveness and safety report has been compiled in accordance with item 2.18
  • □ The plan for collecting and analyzing post-market safety and effectiveness data has been prepared in parallel
  • □ Flexible time has been reserved in the high-risk device schedule for a possible local clinical study

GRZAN provides registration support in Uzbekistan for manufacturers, importers, and authorized representatives, including clinical evidence strategy and dossier compilation. For assistance, contact info@grzanglobal.com.

Regulatory information verified as of October 29, 2026.

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail