Uzbekistan Medical Device Risk Classification: Four Safety Classes and Review Timelines
Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20
1. Key takeaways
From January 1, 2026, Uzbekistan classifies medical devices into four safety classes: Class I (low risk), Class IIa (medium risk), Class IIb (high risk) and Class III (highest risk), with no Class IV. The class is not a label; it directly determines three things: the statutory review period (60, 90, 120 and 120 working days respectively), whether a clinical effectiveness and safety report is required (Class I devices generally do not require one, except sterile, metrological and invasive Class I devices), and whether verification of manufacturing conditions and local testing are triggered. The classification judgment should be completed and documented before submission, so that the company is not placed in the passive position of accepting a reclassification by the Center for Drug Safety during review. Because the review clock excludes deficiency responses, legalization, sample transport and clinical work, the working-day figures must be converted into calendar project time with care. A disciplined classification file, built on intended use, invasiveness, duration of use and the presence of a measuring function, is the single most effective instrument for keeping the review on the class the company originally assigned, and it doubles as the evidence base when the agency raises questions about the assigned class. Because the four-class structure entered into force together with the new registration procedure, companies that registered under the earlier rules cannot carry their previous class assignments forward unchanged: the class must be re-derived from the intended use and the exception list of the current regulation, and the derivation must be visible in the dossier.
2. Legal basis
| Regulation | Number | Date | Provisions |
|---|---|---|---|
| Presidential Decree on Additional Measures for Regulating the Circulation of Medicines and Medical Devices | No. UP-137 | August 19, 2025 | From January 1, 2026, medical devices are divided into four safety classes, with differentiated requirements applied by risk probability |
| Cabinet of Ministers Resolution on Approving the Regulation on State Registration of Medicines and Medical Devices and the Issuance of Registration Certificates | No. 738 | Adopted November 24, 2025, effective February 26, 2026 | Four safety classes; statutory review periods: Class I 60, Class IIa 90, Class IIb 120, Class III 120 working days |
| Cabinet of Ministers Resolution on Approving the Regulation on State Registration of Medicines and Medical Devices and the Issuance of Registration Certificates | No. 738 | Adopted November 24, 2025 | Item 2.18: Class I devices generally do not require a clinical effectiveness report, except sterile, metrological and invasive devices |
3. Expert insights
Insight one: the mapping between class and statutory period sets the baseline cycle of the project. Under Resolution No. 738, Class I carries 60 working days, Class IIa 90 working days, and both Class IIb and Class III 120 working days. Many companies convert the figure of 120 working days directly into half a calendar year, and this is where scheduling errors begin. The statutory period is a review clock: it does not include the time consumed by deficiency responses, notarization and legalization, sample transport or clinical studies. A realistic plan therefore separates the statutory clock from off-clock activities and sequences them explicitly.
Insight two: Class I is not synonymous with exemption from clinical evidence. Item 2.18 establishes three exceptions for Class I: sterile devices, metrological devices and invasive devices still require a clinical effectiveness and safety report. These three categories overlap heavily with the export volumes of manufacturers in the region: single-use sterile consumables, electronic blood pressure monitors and various invasive catheters all fall into the exceptions. Before assuming a low-risk fast track, companies should run every Class I candidate through the three-exception test, because a missed exception surfaces late in the review as a deficiency that resets the clock.
Insight three: classification evidence must come from use, not from marketing. When judging the class, the intended use, the claimed functions, the role of software and the clinical use setting together constitute the classification evidence. A device described as supporting diagnosis and one described as used for diagnosis may fall into different classes. The approach used by GRZAN is to freeze a list of intended-use statements at the qualification stage, so that all subsequent documents and marketing materials quote the same list. This single control removes most of the ambiguity that reviewers otherwise resolve against the applicant.
Insight four: an under-classified product costs more to fix than an over-classified one. Under-classification means the project was scheduled for 60 working days and the evidence was prepared without clinical requirements. Once the center reclassifies the device to Class IIa or IIb, the period lengthens, clinical evidence must be supplemented, and additional samples may be required. Over-classification by one class only produces a fuller but still acceptable evidence set. Where the class is genuinely borderline, preparing to the higher class is the cheaper form of insurance.
4. Practical recommendations
- Issue a written Safety Class Determination Statement at the project initiation stage, recording the basis of the judgment (intended use, invasiveness, duration of use, presence of a measuring function).
- Re-verify every Class I determination against the list of three exceptions: sterile, metrological and invasive.
- Freeze a list of intended-use statements and require all technical documentation, instructions for use, labels and marketing materials to quote it.
- In the project schedule, list the statutory period and off-clock activities on separate lines, so that 120 working days is never equated with half a year.
- For products at a classification boundary, prepare evidence to the higher class and keep a safety margin in the plan.
5. Verification checklist
- □ A safety class has been assigned to every model with a written determination statement
- □ Every Class I determination has been checked against the three exceptions of sterile, metrological and invasive devices
- □ Statutory review periods mapped by class (60, 90, 120 and 120 working days)
- □ Deficiency response, legalization, sample and clinical time listed separately in the project schedule
- □ The intended-use statement list has been frozen and is quoted by all documents
- □ Borderline products have evidence prepared to the higher class
Regulatory information verified as of October 29, 2026.
GRZAN provides registration support in Uzbekistan for manufacturers, importers and authorized representatives. For assistance with risk classification and review scheduling, contact info@grzanglobal.com.
Disclaimer
Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.
About GRZAN
GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.
GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.
Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.
What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.
Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.
Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.
Take your devices global with GRZAN
We work with manufacturers worldwide to open new markets with a clear regulatory pathway.
Contact us