Ecuador Single-Use Devices and the Ban on Fractioning: General Provisions Ten to Twelve

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

The new regulation adds two prohibitions in its general provisions. General provision ten (DECIMA) provides that devices whose label, instructions for use and/or user manual carries the single use only (un solo uso) symbol may not be reused in public or private healthcare institutions anywhere in the country. General provision eleven (DECIMA PRIMERA) prohibits the fractioning (fraccionamiento) of devices: commercialisation must follow the commercial presentation approved by ARCSA.

The two provisions work on different actors. The reuse ban binds healthcare institutions at the point of clinical use; the fractioning ban binds whoever puts the product on the market, and it converts the approved presentation from a commercial preference into a regulatory boundary.

Both provisions rely on definitions and declarations made earlier in the lifecycle. Article 3 defines a single-use device as one intended by the manufacturer for use in a single procedure on a single patient, after which it is discarded, and Article 23, item (f) requires the commercial presentation and/or model or reference to be declared at filing. What is fixed at filing becomes the limit after approval.

2. Legal basis

RegulationReference numberPublication dateArticles
Substitute Technical Health Regulation for the Health Registration, Control and Surveillance of Medical Devices for Human UseResolucion ARCSA-DE-2026-003-DASPSigned 2026-04-28General provision ten (prohibition of reuse)
SameSameSameGeneral provision eleven (prohibition of fractioning)
SameSameSameArticle 49, item (m), sub-item 3 (single use or non-reusable symbol)
SameSameSameArticle 3 (definition of a single-use device), Article 72 (sanctions)

3. Expert insights

Insight one: general provision ten binds healthcare institutions, not manufacturers. The text states that devices bearing the single use only (un solo uso) symbol on the label, instructions for use and/or manual may not be reused in public or private healthcare institutions nationwide. Once the symbol is applied, a mandatory obligation is triggered at the clinical end, and the market space for reprocessing and reusing the device is closed indirectly.

Insight two: general provision ten pairs with Article 49, item (m), sub-item 3. Article 49 requires the label to carry the disposable or non-reusable (Producto desechable o no reusable) symbol for devices that are sterile and not reusable. Read together, the two provisions lock a sterile, non-reusable device at both ends: the label declares the status, and the institution is barred from reuse.

Insight three: general provision eleven prohibits fractioning against the approved presentation. The text prohibits fractioning (fraccionamiento) of devices for human use and requires commercialisation in accordance with the commercial presentation approved by ARCSA. A product imported in a full carton therefore cannot be broken down and sold in units, unless the broken-down packaging has itself been approved.

Insight four: Article 23, item (f) makes the benchmark a declared item. The application must state the commercial presentation and/or the model or reference. Because the approved presentation is definite, the test for whether a sale amounts to fractioning is also definite, and it is not a matter of local commercial custom.

Insight five: Article 42, item (h) provides the lawful route to adjust the presentation. That article lists a change, addition or removal of the commercial presentation among the situations that do not require re-registration and are handled through the change procedure. A company that genuinely needs to sell in smaller units should obtain approval through that route rather than splitting stock in the warehouse.

Insight six: general provision twelve opens an expert review channel. ARCSA may convene its internal advisory committee (CAI) or an external expert committee (CEE) at any stage of registration, whether ordinary or simplified, or of a change application, and the report issued serves as the basis for the regulatory decision. For a technical dispute about presentation, sterility or reuse, that is the structured route to a reasoned outcome.

4. Practical recommendations

  1. Apply the single use only symbol deliberately: it triggers the reuse ban at the clinical end, so the marking must match the product positioning and the validated reprocessing status.
  2. Sell in the approved commercial presentation, and do not break bulk cartons into units for distribution.
  3. Where fractioning is genuinely required, obtain approval through the change procedure under Article 42, item (h) before selling.
  4. Write the no-fractioning obligation into distributor agreements, with a defined remedy for breach.
  5. In a technical dispute, assess whether to invoke general provision twelve and ask ARCSA to convene an expert committee.

5. Verification checklist

  • □ Whether the product carries the single use only symbol, and its consequences, have been confirmed
  • □ The sales format has been checked against the commercial presentation approved in the certificate
  • □ Distributor agreements state the no-fractioning obligation
  • □ Any requirement to sell in smaller units has been filed through the change procedure
  • □ The route for invoking the CAI or CEE expert review mechanism is understood
  • □ The end-use restriction on single-use devices has been explained to customers

Regulatory information verified as of October 29, 2026

GRZAN core service lines

  1. Global registration access: EU CE (MDR/IVDR), US FDA (510(k)/De Novo/PMA), UK UKCA, Australia TGA, Canada Health Canada, Brazil ANVISA and Russia RZN, together with pre-market registration pathway planning and submission execution across Latin America, Southeast Asia and the Middle East.
  2. Authorized representative services: EU EC REP, UK RP, Swiss CH REP, Australian sponsor and US Agent holding and representation services, with support for neutral license-holder structures.
  3. Quality system compliance: ISO 13485 establishment and maintenance, MDSAP audit preparation, GMP conformity coaching, and technical documentation (TD) drafting and remediation.
  4. Clinical and evaluation support: clinical evaluation plans (CEP) and reports (CER), equivalence argumentation, PMCF planning and execution, and IVD performance evaluation.
  5. Post-market compliance: vigilance and adverse event reporting, FSCA and recall execution, registration change and renewal management, and maintenance of national regulatory databases.

Why manufacturers choose GRZAN

  • Local regulatory teams: target-market work is performed by locally licensed regulatory professionals, and submission dossiers are written directly in the local official language without secondary translation.
  • Neutral license-holder structures: GRZAN offers holding arrangements that are not tied to any distribution channel, preventing registration certificates from being locked in by channel partners.
  • Regional leverage: one technical dossier can be reused across Andean Community and Mercosur member states, spreading regional compliance costs.
  • Full lifecycle support: from classification determination and pathway design through dossier preparation, license holding, customs clearance, post-market vigilance and renewal.

Contact

  • GRZAN provides registration support in Ecuador for manufacturers, importers and authorized representatives.
  • Email: info@grzanglobal.com
  • Service inquiries: Ecuador and Andean market registration plans, license-holder structure design, official fee and timeline estimation

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail