Ecuador IVD Performance Studies: Article 24 Item 2.4 Requirements and Article 52 Instruction for Use

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

1. Key takeaways

Performance studies for in vitro diagnostic devices (DMDIV) in Ecuador are governed by item 2.4 of Article 24, which applies to all risk classes and covers 4 categories of studies: precision, repeatability and reproducibility; sensitivity; specificity; and other applicable performance or clinical studies — accuracy, detection limit, linearity, interference, robustness and comparable parameters, conducted in accordance with internationally recognised guidelines and technical standards.

The practical estimate used by GRZAN is that an IVD dossier in Ecuador succeeds or fails on the alignment between the performance claims and the study data: the instruction for use must state the performance characteristics and the limitations of the method, and every stated claim must be traceable to a study in the file. Companies that maintain a claim-to-data mapping table from the first draft of the dossier avoid the most frequent correction findings in this product family.

2. Legal basis

RegulationNumberDateProvision
Alternative Technical Health Regulation for the Sanitary Registration, Control and Surveillance of Medical Devices for Human UseResolución ARCSA-DE-2026-003-DASPSigned 2026-04-28Article 24, item 2.4 (IVD studies, 4 categories)
Same regulationSame numberSame dateArticle 52 (IVD instruction for use elements)
Same regulationSame numberSame dateArticle 14 (grouped certification by methodology and analyte group) and Article 32 (DMDIV limit of 50 entries)

The elements listed by Article 52 for the IVD instruction for use are: the trade name; the application and intended use; the components and the kit contents; the methodology; the principle of the method; the performance characteristics and the limitations of the method; the preparation of the reagents; the storage conditions and stability of the reagents; the specimen or sample; the operating procedure; the internal quality control; and the reference intervals, where applicable.

3. Expert insights

Insight one: IVD performance studies carry no class threshold — all risk classes are covered alike. Unlike the tiered logic applied to other study types, item 2.4 applies to DMDIV of every risk class. The reason is structural: the risk of an in vitro diagnostic lies in the clinical consequences of an incorrect test result, which exists at every class level.

Insight two: the fourth category is an open list, and the methodological basis must be internationally recognised. After listing accuracy, detection limit, linearity, interference and robustness, the text closes with other performance studies conducted in accordance with internationally recognised guidelines and technical standards. This gives companies an explicit basis for applying the methodologies of CLSI, ISO and comparable standard-setting bodies.

Insight three: the performance conclusions must correspond line by line with the instruction for use. Article 52 requires the IVD instruction for use to state the performance characteristics and the limitations of the method, and those statements must be directly supported by the study data of item 2.4. Claims that exceed what the data support are a common cause of both corrections and failures in Type 2 post-market sampling.

Insight four: the grouped certification conditions and the performance studies constrain each other. Article 14 requires that the IVD products covered by a single registration certificate belong to the same analytical methodology and the same group of analytes to be detected. Reagents based on different methodologies cannot be merged into one certificate even when their performance studies are complete.

Insight five: Article 32 sets a separate ceiling of 50 entries for DMDIV. The ordinary procedure of Article 10 allows 100 entries for devices and 50 for in vitro diagnostics; the simplified procedure of Article 32 mirrors the split, with 100 entries for DMI, DMNI and DMA and 50 for DMDIV. Batch planning for IVD product families must therefore be calculated on the 50-entry practical estimate.

The 50-entry ceiling interacts with the grouping rule of Article 14 in a way that deserves explicit arithmetic. Because a certificate may only contain reagents of the same analytical methodology and the same analyte group, a product family that spans two methodologies cannot simply be split into batches of 50 — it must first be split by methodology, and each resulting group must then stay at or below 50 entries. A family of, say, 120 immunoturbidimetric assays and 80 enzymatic assays becomes two projects, not three, only if each methodology group fits within the ceiling; otherwise the change procedure of the regulation becomes the expansion route once the initial certificates are granted. Building this arithmetic into the project plan before submission avoids the suspension that Article 10-type over-limit handling imposes on applications awaiting correction.

4. Practical recommendations

  1. Build a standard template around the 4 study categories: cover precision, repeatability and reproducibility, sensitivity and specificity, and select the fourth-category studies according to the characteristics of the product.
  2. Cite the international standard behind each methodology: reference the relevant CLSI or ISO documents to satisfy the internationally recognised requirement.
  3. Align every performance claim in the instruction for use with the study data: maintain a claim-to-data mapping table as a living document.
  4. Group applications by methodology and analyte group: file reagents based on different methodologies in separate applications.
  5. Batch IVD families at the 50-entry ceiling: extend the scope beyond it through the change procedure rather than exceeding the limit.

5. Verification checklist

One more alignment point closes the loop between the studies and the label. Article 49 requires the label to state the storage conditions and, where applicable, the expiry date, and Article 52 requires the instruction for use to state the storage conditions and stability of the reagents. The values printed on the label and in the instruction for use must therefore be the same values the stability and real-time data support; a kit whose studies were run at a narrower temperature range than the label claims has a claim-to-data gap that the mapping table described above is designed to catch before submission rather than after a correction finding.

  • □ Precision, repeatability and reproducibility studies have been completed.
  • □ Sensitivity studies have been completed.
  • □ Specificity studies have been completed.
  • □ The fourth-category studies have been completed according to the product characteristics, with the international standard basis cited.
  • □ Every performance claim in the instruction for use corresponds to the study data.
  • □ Applications have been grouped by methodology and analyte group.
  • □ Batch planning has been completed within the DMDIV limit of 50 entries.

Regulatory information verified as of October 29, 2026

GRZAN supports manufacturers, importers and authorised representatives preparing IVD performance study packages for Ecuador medical device registration, from Article 24 item 2.4 study templates and CLSI-aligned methodologies to claim-to-data mapping against Article 52. For tailored assistance with your Ecuador medical device registration, contact us at info@grzanglobal.com.

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail