D.S. 003-2020-SA Classification Rules: Peru and the IMDRF 18 Rules

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

Regulatory information verified as of October 29, 2026.

1. Key takeaways

D.S. N° 003-2020-SA, published on 2020-02-06, replaced experience-based classification with the 18 classification rules of Anexo 1, arranged in 4 groups covering non-invasive, invasive, active and additional rules; the technical line is derived from the same source as the IMDRF framework and the EU MDR. The practical conclusion of that common origin is that classification conclusions from other jurisdictions may be cited as argumentation material but cannot be transplanted directly. Once a class is fixed, it simultaneously locks the dossier depth (Art. 124 for Class I, Art. 125 for Class II) and the review clock (22 / 44 / 66 working days), and the final determination authority rests with DIGEMID. In planning terms this means the classification exercise is not a paperwork step at the edge of the project: it is the variable that drives evidence budgets, clinical evaluation needs, timelines and the applicable conformity documents, so it should be completed and documented before any dossier writing begins.

2. Legal basis

RegulationNumberPublication dateProvisions and key points
Reglamento que establece las Reglas de Clasificación y los Principios Esenciales de Seguridad y Desempeño para los Dispositivos MédicosD.S. N° 003-2020-SA2020-02-06Anexo 1 contains 18 classification rules (non-invasive / invasive / active / additional rules); establishes the essential principles checklist
Reglamento para el Registro, Control y Vigilancia Sanitaria…D.S. N° 016-2011-SA2011-07-27Art. 124 (Class I dossier), Art. 125 (Class II dossier), Art. 136 (calendar-day ceilings 30 / 60 / 90)
Modificatoria del D.S. N° 016-2011-SAD.S. N° 016-2017-SA2017-06-07Unifies Title IV terminology to principios esenciales, paving the way for the essential principles checklist
Ley de los Productos Farmacéuticos, Dispositivos Médicos y Productos SanitariosLey N° 294592009-11-26Superior legal basis for device regulation
Nuevo TUPA del MINSAD.S. N° 011-2026-SA2026-07-21Review clock: Class I 22, Class II 44, Classes III/IV 66 working days, in vitro diagnostic devices 30 working days (practical estimate; article numbers not verified line by line)

Regulatory information verified as of October 29, 2026.

3. Expert insights

Insight 1: The 18 rules form a decision sequence, not a menu from which to pick.

The rules of Anexo 1 are organized by the way the device contacts the human body, its duration of use and its source of energy, and they must be applied in order, rule by rule, stopping at the first hit. This explains why the same product can receive different classes in the hands of different companies: the common error is jumping ahead, applying the rule that seems to fit while skipping the preceding rules, and thereby underestimating the risk class. GRZAN requires a rule-by-rule comparison record in its grading work, documenting the exclusion reasons for every rule found inapplicable, precisely to close this loophole.

Insight 2: The additional rules are a lateral upgrade mechanism that often pushes Class II to Class III.

The additional rules address cross-cutting attributes: the presence of medicinal substances, animal or human derived materials, or a killing or inactivating function. Their mechanism is additive: the first three groups of rules produce a base class, and the additional rules then determine whether the class rises. In practice this group is the most underestimated. Companies grade by the physical form of the device and ignore upgrade clauses triggered by formulation or material properties; the dossier is prepared for Class II, the review demands Class III evidence, and the filing is rejected outright. The rework is not limited to rewriting one form: the entire evidence plan, including the essential principles checklist and the review-clock assumption, must be rebuilt, which is why the additional-rule check belongs at the very start of grading rather than at the end.

Insight 3: Common origin does not mean equivalence, and the class binds downstream costs.

A common technical lineage means the reasoning can be cross-referenced: a classification analysis completed under the IMDRF or EU MDR rules has reference value in its analytical framework. But the Peruvian class is the product of a local administrative determination, and once fixed it binds three things at once: dossier composition (Art. 124 and Art. 125 as the documentary bases for Class I and Class II), the review clock (22 / 44 / 66 working days), and the applicability of the essential principles checklist (Classes II-IV). The correct order is therefore to self-assess and keep records, then cross-validate against external opinions, and finally accept the DIGEMID determination, rather than copying the class printed on a foreign certificate into the application form.

4. Practical recommendations

  1. Produce a rule-by-rule application record: compare each of the non-invasive, invasive, active and additional rules in the 4-group order, stating the reasons for applicability and non-applicability.
  2. Run a dedicated additional-rule check: for each SKU, verify separately whether it contains medicinal substances, animal or human derived materials, or a killing or inactivating function, and judge whether an upgrade is triggered.
  3. Link class, dossier and clock: once the grading conclusion is issued, confirm in the same step the dossier composition, the review clock, and whether the essential principles checklist is required.
  4. Use foreign classifications as arguments, not conclusions: when citing the IMDRF or EU MDR analytical framework, mark it explicitly as argumentation material; the class on the application is subject to the DIGEMID determination.
  5. Bind grading to use claims: whenever a use claim changes, run the 18 rules again instead of reusing the old conclusion.

5. Verification checklist

□ Completed a documented rule-by-rule comparison under the 18 rules of Anexo 1

□ Completed a dedicated check of additional-rule triggers (medicinal substances, animal or human derived materials, killing or inactivating functions)

□ Confirmed the dossier articles corresponding to the class (Class I Art. 124, Class II Art. 125)

□ Confirmed the review clock corresponding to the class (22 / 44 / 66 working days)

□ Confirmed that Classes II-IV must attach the essential principles checklist

□ Confirmed that the class on the application follows the DIGEMID determination, with foreign conclusions used only as argumentation

Regulatory information verified as of October 29, 2026.

GRZAN Regulatory Team

GRZAN provides medical device registration support in Peru for manufacturers, importers and authorized representatives. For inquiries, contact info@grzanglobal.com.

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail