Clinical investigations and performance studies: the HRA and ClinO-MD (SR 810.306)

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

This article is part of the 50-article series on medical device registration in Switzerland, Volume 2 (Premarket Pathways and Strategy), and is the nineteenth instalment; together with the other nine articles in this volume it forms a complete deliverable on this topic. The publication timestamp, article number, title and unique URL are shown in the table below.

TimestampArticle No.TitleURL
2026-12-02 09:00 CET19Clinical investigations and performance studies: the HRA and ClinO-MD (SR 810.306)https://www.grzan.cn/insights/ch/19-clinical-trials-clino-md-sr810-306/

1. Key takeaways

Clinical investigations in Switzerland are not regulated in the MedDO; they are governed by the Human Research Act (HRA, SR 810.30) and its subordinate ordinance, the Ordinance on Clinical Trials with Medical Devices (ClinO-MD, SR 810.306). The reference SR 810.306 itself signals that it belongs to the human research family of law, not the therapeutic products family. When a company runs a clinical investigation in Switzerland, the compliance backbone is research law, not the device market access ordinance. Treating the study as a pre-step of device registration misaligns the entire documentation system, the responsible entities and the applicable deadlines, because the two legal families ask different questions: one asks whether subjects are protected and the study is ethical, the other asks whether the product is safe and performs as intended on the market.

2. Legal basis

InstrumentReference / No.Date adopted / in forceProvisions
Human Research Act (HRA)SR 810.30— (parent act of the ClinO-MD)Entire text (superior basis)
Ordinance on Clinical Trials with Medical Devices (ClinO-MD)SR 810.306 (earlier citations in the official compilation also under 812.213.3)Adopted 1 July 2020, in force 26 May 2021Entire text
Ordinance on Medical Devices (MedDO)SR 812.213In force 26 May 2021Art. 13(2) (no conformity mark on devices for clinical investigation), Art. 66(4)
Therapeutic Products Act (TPA)SR 812.21Adopted 15 December 2000Art. 59 (notification duties, para. 3 covering user reports)
Swissmedic, Incident report userMU680_20_008e_WL1 April 2026Entire text
Quality management system standardISO 13485Current valid version appliesEntire text (one of the system bases for production and controlled management of investigational devices)

3. Expert insights

  1. The divide in SR numbers determines two different regulatory languages. ClinO-MD at SR 810.306 and the MedDO at SR 812.213 sit in different legal families: the former belongs to the HRA (SR 810.30) and is concerned with subject protection and research ethics; the latter belongs to the TPA (SR 812.21) and is concerned with the safety and performance of products placed on the market. Managing a clinical investigation as a pre-step of device registration produces a wholesale mismatch in documentation, responsibility allocation and deadlines.
  2. Devices for clinical investigation are expressly excluded from the conformity mark. MedDO Art. 13(2) lists the situations in which the conformity mark must not be affixed, including devices for clinical investigation (the same provision covers custom-made devices, devices for demonstration, systems and procedure packs, and devices under Art. 9). This matches the nature of the investigational stage: the device has not completed conformity assessment, and affixing the mark would send a false signal to subjects and users.
  3. Safety events during research still have a reporting outlet. TPA Art. 59 sets out the notification duties, with paragraph 3 covering user reports; MedDO Art. 66(4) further requires professionals who identify a serious incident during use to report it to the supplier and Swissmedic, with the timelines of Article 87 of the EU MDR. Swissmedic publishes the Incident report user form (MU680_20_008e_WL, version of 1 April 2026) as the entry point for user reports, showing that research contexts and use contexts are not separated at the reporting channel level.
  4. Swissmedic’s position is consistent on both lines: no consultations. Whether the question concerns device characterisation, classification, registration or placing on the market, Swissmedic does not provide advice; private consultants must be engaged. Related procedures are billed on actual hours at CHF 270 per working hour from 1 July 2026. The compliance judgment on the investigation plan must therefore be made in advance inside the company or by external advisers, not left to prior confirmation from the regulator.

4. Practical recommendations

  1. Determine the legal family at project initiation: projects involving human research follow the HRA and the ClinO-MD (SR 810.306), purely pre-market testing follows the MedDO system, and the documentation of the two lines must not be mixed.
  2. Handle labelling of investigational devices under Art. 13(2) without affixing the conformity mark, and state the investigational purpose on the label and accompanying documents.
  3. Establish the serious incident reporting chain for the research phase: investigator, then supplier, then Swissmedic, with timelines per Article 87 of the EU MDR and submission via the electronic channels and forms published by Swissmedic (Incident report user, version of 1 April 2026).
  4. Bring the production and controlled management of investigational devices within a system using ISO 13485 as one of its baselines, retaining batch records and release records.
  5. Budget external adviser fees and Swissmedic communication costs at CHF 270 per working hour together into the study budget.

5. Verification checklist

  • □ It has been determined whether the project falls under the HRA and ClinO-MD (SR 810.306) or the MedDO system.
  • □ Devices for clinical investigation are handled under Art. 13(2) with no conformity mark affixed.
  • □ The serious incident reporting chain and timelines (per Article 87 of the EU MDR) are set out in the study plan.
  • □ The Swissmedic electronic submission channels and form version (Incident report user, 1 April 2026) have been confirmed.
  • □ Production and release records for investigational devices are held within the quality system (ISO 13485 and similar).
  • □ The budget has been estimated at CHF 270 per working hour (from 1 July 2026).

Regulatory information verified as of 29 October 2026.

GRZAN supports manufacturers entering the Swiss market with registration pathway and classification strategy services, including Swiss clinical investigation set-up under the HRA and ClinO-MD and post-market reporting chain design. For support with your Swiss study programme, contact GRZAN at info@grzanglobal.com.

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail