UDI Carrier Timeline under MDR Article 123(3)(g): Dates by Class

Updated: 2026-11-12 · GRZAN Regulatory Team · Verification date 2026-10-20

UDI carrier deadlines in the EU MDR arrive in phases that differ by device class and by carrier location. This article explains the two separate timelines under MDR Article 123(3)(g), why the 2027 date for Class I reusable devices matters, and how the carrier duty differs from the EUDAMED registration duty.

1. Key takeaways

Under MDR Article 123(3)(g), the application of UDI carriers on labels is phased in three stages by device class: Class III and implantable devices from 2021-05-26; Class IIb and Class IIa devices from 2023-05-26; and Class I devices from 2025-05-26. For reusable devices, for which the UDI carrier must be placed on the device itself, the corresponding dates are 2023-05-26, 2025-05-26 and 2027-05-26.

The label timeline has now fully expired for all classes, while the device-itself timeline still has one date ahead for Class I reusable devices. The two schedules must not be conflated, and the carrier duty must also be kept distinct from the duty to register UDI data in the EUDAMED database, which becomes mandatory from 2026-05-28 under Commission Decision (EU) 2025/2371.

For planning purposes, the two timelines should be entered into the product roadmap as separate milestone rows, because they are driven by different provisions and fail independently. A device can be fully compliant on the label timeline while still missing its device-itself milestone, and the reverse is not possible: the device-itself marking never substitutes for the carrier on the label and the higher packaging levels, both of which remain required under Article 27(4).

2. Legal basis

The table below lists the provisions and sources cited in this article.

Regulation / documentProvisionsDatesSource
Regulation (EU) 2017/745Article 123(3)(g) (phased application of UDI carriers); Article 27(4) (carrier on label and packaging)Adopted 2017-04-05; applicable 2021-05-26EUR-Lex
Regulation (EU) 2017/745Article 123(3) (other phased application matters)Same as aboveEUR-Lex
Irish HPRA official pageSummary of the UDI timeline (Class III 2021-05-26 / 2023-05-26; Class IIb, IIa 2023-05-26 / 2025-05-26; Class I 2025-05-26 / 2027-05-26)Verified 2026hpra.ie

The phase dates above are fixed in the text of Article 123(3)(g) and are echoed in national authority summaries such as the Irish HPRA page, which is useful as an operational cross-check when planning product roadmaps.

3. Expert insights

Insight one: two timelines must be distinguished — the carrier on the label and the carrier on the device itself.

The first timeline governs when the UDI carrier must appear on the label: Class III and implantable devices from 2021-05-26, Class IIb and IIa from 2023-05-26, and Class I from 2025-05-26. The second timeline governs when reusable devices must carry the UDI carrier directly on the device itself: Class III from 2023-05-26, Class IIb and IIa from 2025-05-26, and Class I from 2027-05-26. All label deadlines have expired; the device-itself timeline still holds one open milestone for Class I reusable devices.

Insight two: 26 May 2027 for Class I reusable devices is the next hard deadline.

For Class I reusable devices still in use, the UDI carrier must be placed on the device itself from 2027-05-26, and under the requirements of Annex VI the marking must be permanent and remain readable after each reprocessing cycle. This is a product design and process task, since direct part marking of metal components differs fundamentally from printed labels, and it cannot be completed by merely updating the label artwork. Process validation must therefore start well ahead of the deadline.

Insight three: this timeline is not the same as the legacy certificate transition period.

The dates of 2027-12-31 and 2028-12-31 in Article 120(3a) govern how long legacy devices may continue to be placed on the market; the date of 2027-05-26 in Article 123(3)(g) governs when the UDI carrier must appear on the device itself. The two have different legal sources and different consequences, and neither can substitute for or offset the other in planning.

Insight four: the UDI carrier and the UDI database registration are two independent obligations.

Even where carriers have been applied according to the timeline, the registration of UDI data in the database under Article 29 remains a separate duty. From 2026-05-28 the EUDAMED UDI and device registration module is mandatory, following Commission Decision (EU) 2025/2371, so the two obligations run in parallel and are checked in parallel during market surveillance.

The verification of readability after reprocessing deserves particular care in the validation plan, because marking methods degrade differently under cleaning agents, autoclave cycles and mechanical abrasion. The validation protocol should therefore sample the full specified cycle count, not an abbreviated run, and define objective acceptance criteria for readability before the process is released for production.

4. Practical recommendations

  1. Build a matrix of device class against reusability and determine the applicable carrier deadline for every SKU individually.
  2. Start permanent marking process validation for Class I reusable devices early, including readability verification after reprocessing.
  3. Split the UDI carrier duty and the UDI database registration duty into two separate check lines, each with its own pre-release checkpoint.
  4. Link marking validation with the sterilisation and reprocessing validation, confirming that the marking remains readable after the specified number of cycles.
  5. Enter 2027-05-26 into the schedule for product changes and process modifications now, so tooling and validation work lands before the deadline.

Because the two obligation lines run in parallel, the release checkpoint for a single batch should verify both at once: that the carrier is present and readable at every required level, and that the database record for the Basic UDI-DI and UDI-DI is complete and current. Splitting the checks into two named lines with different owners keeps responsibility clear when one of them fails.

5. Verification checklist

  • □ A class-by-reusability matrix has been built and applied to every SKU
  • □ Label UDI carriers are in place according to the class timeline (2021-05-26 / 2023-05-26 / 2025-05-26)
  • □ Reusable devices are scheduled against 2023-05-26 / 2025-05-26 / 2027-05-26
  • □ Permanent marking process validation for Class I reusable devices has started
  • □ Marking readability after reprocessing has been verified
  • □ UDI database registration status has been confirmed (mandatory module from 2026-05-28)

Regulatory information verified as of 2026-10-29.

GRZAN supports medical device manufacturers entering and maintaining the EU market under Regulation (EU) 2017/745. Our team maps UDI carrier deadlines under MDR Article 123(3)(g) to individual SKUs, validates permanent marking processes, and coordinates carrier duties with EUDAMED registration. For technical file and conformity support for the EU MDR market, contact info@grzanglobal.com.

Disclaimer

Prepared by GRZAN from publicly available regulation. It is provided for general information only and does not constitute legal advice. Legislation and official fees are subject to change by the competent authority.

About GRZAN

GRZAN (Global Regulatory Zone Access Navigator) — the global market-access navigator for medical devices and IVDs.

GRZAN is a specialist consultancy for medical device and IVD market access, providing end-to-end regulatory solutions across the product lifecycle and making international market entry simpler and faster.

Key figures. More than 25 wholly owned subsidiaries worldwide covering 50+ countries and regions; chosen by 100+ listed and leading Chinese device manufacturers, with 4,000+ successful cases; 12+ years in the sector and 200+ regulatory training sessions and seminars delivered.

What we do. Global market access: multi-country registration programmes, technical documentation, regulatory advice. Clinical evidence: clinical evaluation reports (CER), usability studies, overseas clinical investigations. Quality systems: ISO 13485, MDSAP, FDA QMSR and BGMP implementation and audit readiness. Authorised representative and licence-holder services across Southeast Asia, Latin America, the Middle East, Africa, Europe and North America.

Why GRZAN. A real operating entity, with the certificate held by the manufacturer: our direct overseas subsidiaries act as the compliant local holder, so the regulatory relationship stays clear and you keep control rather than being tied to a distributor. Direct access to the team: no time-zone gap, no language barrier, no cultural gap. Review points anticipated: we know what FDA 510(k), EU MDR, Southeast Asian, Latin American and Middle Eastern reviewers look for, which shortens timelines and reduces deficiency letters. Pathway planning: pre-decision analysis and risk assessment before you commit.

Every engagement rests on three commitments: efficiency through pre-submission review, certainty through a clear pathway, and stability through a licence-holder that stays in place.

Take your devices global with GRZAN

We work with manufacturers worldwide to open new markets with a clear regulatory pathway.

Contact us
WhatsApp
QR Code
WhatsApp QR Code
Scan to chat on WhatsApp
Phone
Call
Office Mon-Fri 9:00-18:00
0755-88659477
Mobile 24/7 Service
+86 15816864648
Email
Mail